Skip to content

THERMAGE FLX TH-6Recalls & reported problems

Solta Medical · THERMAGE FLX HANDPIECE · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00816995021278), received between Apr 18, 2024 and Nov 26, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Excessive Heating” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Excessive Heating2
  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Cooling1
  • Leak/Splash1

Patient problems

  • Burn(s)2
  • Erythema2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Cryogenic Burn1
  • Eye Pain1
  • Headache1
  • Rash1
  • Scar Tissue1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A USER FACILITY REPORTED THAT A PATIENT WAS BURNED ON THE EYELIDS DURING A THERMAGE FLX TREATMENT. FURTHER EVENT DETAILS WERE REQUESTED; HOWEVER, THE CUSTOMER DECLINED TO SUPPLY ANY DETAILS.

    Patient problems: Burn(s)

    MAUDE report 20787298 ↗
  • InjuryAug 18, 2024Insufficient Cooling; Excessive Heating; Insufficient Cooling; Excessive Heating

    A USER FACILITY REPORTED THAT DURING A THERMAGE FLX TREATMENT SESSION TO THE FACE, THE MACHINE STOPPED RELEASING CRYOGEN. THE ENERGY LEVEL WAS REDUCED AND THE PATIENT COMPLAINED THAT THE SENSATION WAS VERY HOT. THE TREATMENT WAS THEN TERMINATED IMMEDIATELY. ONE DAY POST TREATMENT, THE PATIENT CONTACTED THE USER FACILITY TO REPORT THAT SHE HAD SUFFERED BURNS TO THE FOREHEAD FROM THE THERMAGE TREATMENT AND SHOWED SIGNS OF INFECTION AND THAT THERE WERE SOME SCARS ON THE FOREHEAD AND HER CHEEKS ARE SAGGIER. THE NATURE OF THE INJURY IS REPORTED TO BE SCARS WITH SECONDARY INTERVENTION INCLUDING A SCAR CREAM (SAVLON) TO PREVENT SCARS AND PARACETAMOL TO DEAL WITH THE HEADACHE CAUSED BY EMOTION OF THE SITUATION. FIVE DAYS POST TREATMENT, IT WAS REPORTED THAT THE PATIENT WAS "OK AND THE MARKS ARE BARELY SEEN.¿ THE FACILITY HAS RECEIVED NO FURTHER UPDATES FROM THE PATIENT. AVAILABLE PICTURES (AT THE TIME OF THE REPORT) WERE REVIEWED BY THE MEDICAL REVIEWER AND ERYTHEMA AND SMALL BURNED AREAS WERE VISIBLE ON PATIENT'S FOREHEAD. ADDITIONAL PICTURES WERE RECEIVED A FEW DAYS AFTERWARDS. TWO PHOTOS TAKEN BEFORE THE TREATMENT DO NOT SHOW RED SPOTS ARE BURN MARKS WHILE PHOTOS AFTERWARDS SHOW RED SPO

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Burn(s); Erythema; Eye Pain; Headache; Scar Tissue

    MAUDE report 19999061 ↗
  • InjuryApr 18, 2024Leak/Splash; Excessive Heating; Leak/Splash; Excessive Heating

    A USER FACILITY REPORTED A CRYOGEN LEAKAGE BURN TO THE RIGHT SIDE OF THE PATIENT'S FACE THAT OCCURRED DURING A THERMAGE FLX TREATMENT. THE NATURE OF THE INJURY WAS REPORTED AS REDNESS, A RASH ALL AROUND THE FACE, BUBBLING, AND SENSATIONS OF HEAT AND STINGING. SECONDARY INTERVENTION INCLUDED BCN ASIAN CENTELLA AND B5, AND A EGF MASK TO APPLY POST TREATMENT. THE PATIENT'S CURRENT STATUS IS UNCERTAIN, AND IT IS UNKNOWN IF THERE WILL BE ANY PERMANENT DAMAGE OR SCARRING. A CONSULTATION FOLLOW UP WAS MENTIONED. AN AVAILABLE PICTURE WAS REVIEWED BY THE MEDICAL REVIEWER. PINPOINT BROWN LESIONS ARE SCATTERED ON ONE SIDE OF THE PATIENT''S FACE. IT IS UNCLEAR IF THE LESIONS ARE CRUSTS OR POST INFLAMMATORY HYPERPIGMENTATION. NO OTHER TREATMENTS (BESIDES THE ONE REPORTED) WERE PERFORMED IN SAME AREA WHERE SYMPTOMS WERE REPORTED. THE PATIENT HAS NOT UNDERGONE ANY OTHER TREATMENTS IN THE SAME SYMPTOM AREA WITHIN THE PAST 90 DAYS. THE PATIENT HAS NOT HAD PRIOR AESTHETIC TREATMENTS ON THE AFFECTED AREA. THE INCIDENT OCCURRED AT 193 REPS. THE HIGHEST ENERGY LEVEL USED WAS 2.5 MJ. ADDITIONALLY, AN ERROR MESSAGE OF "TIP TOO WARM" OCCURRED DURING THE TREATMENT.

    Patient problems: Cryogenic Burn; Erythema; Rash; Skin Burning Sensation

    MAUDE report 19140068 ↗

Other Massager, Therapeutic, Electric devices

All Massager, Therapeutic, Electric devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page