VBEAM Prima Laser System 9914-00-9080Recalls & reported problems

Candela Corporation · Pulsed Laser · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00817495022963), received between Nov 4, 2024 and Nov 4, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Electrical /Electronic Property Problem1
  • Insufficient Device Problem Information1
  • Noise, Audible1

Patient problems

  • Tinnitus1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 4, 2024Electrical /Electronic Property Problem; Detachment of Device or Device Component; Insufficient Device Problem Information; Noise, Audible; Electrical /Electronic Property Problem; Detachment of Device or Device Component; Insufficient Device Problem Information; Noise, Audible

    CANDELA CUSTOMER, A DISTRIBUTOR IN NETHERLANDS, REPORTED THAT THERE WAS AN "EXPLOSION" WITHIN A VBEAM PRIMA WHICH OCCURRED DURING PREVENTIVE MAINTENANCE OF THE DEVICE. IT WAS REPORTED THAT THERE WERE NO INJURIES RELATED TO THE EVENT; PCB BOARD WAS BOWED, AND NO FRAGMENTS BREACHED THE SYSTEM. CANDELA TEAM MET WITH CUSTOMER TO FURTHER DISCUSS THE REPORTED EVENT ON 10OCT2024. AT THIS TIME, IT WAS REPORTED THAT THE FIELD SERVICE ENGINEER (FSE) SERVICING THE DEVICE AT THE TIME OF THE MALFUNCTION HAD RINGING IN HIS EARS (TINNITUS) FOLLOWING THE EVENT. HE HAD NOT SOUGHT MEDICAL ATTENTION. PER CANDELA REGULATORY REPORTABILITY DECISION FOR LOUD NOISE OCCURRENCE, THE LASER OPERATES BETWEEN 50 DBA AND 70 DBA DURING NORMAL USE. IF A FAILURE WOULD OCCUR, IT WOULD BE AN IMPULSE ACOUSTICAL NOISE ABOVE 80DBA, BUT BELOW 140 DBA. THE USER OR PATIENT COULD POTENTIALLY BE SUBJECTED TO THE FOLLOWING CONDITION(S): TEMPORARY RINGING OR BUZZING IN EAR, MOST CONDITIONS LASTING LESS THAN 16 TO 48 HOURS. FURTHER FOLLOW UP FROM CUSTOMER ON (B)(6) 2024 REPORTED THAT THE FSE CONTINUED TO HAVE TINNITUS AND HAS SOUGHT MEDICAL ATTENTION. THE FSE FOLLOW-UP NOTES THAT HE STARTED THE ANNUAL PREVENTIVE MAINTENANCE OF

    Patient problems: Tinnitus

    MAUDE report 20601782 ↗