Skip to content

Curaplex 2021-15028Recalls & reported problems

Bound Tree Medical, LLC · Curaplex Nasopharyngeal Airway, Latex-Free, PVC, 28 FR · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00817617025780), received between Jul 2, 2019 and Jul 2, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Inadequacy of Device Shape and/or Size1
  • Malposition of Device1
  • Suction Problem1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 2, 2019Inadequacy of Device Shape and/or Size; Suction Problem; Malposition of Device

    I MANAGE QUALITY ASSURANCE FOR THE FIRE DEPARTMENT BASED PARAMEDIC UNITS IN (B)(6). WE HAD A NASOPHARYNGEAL AIRWAY BACKWARDS INTO THE OROPHARYNGEAL SPACE. THE EMS CREW PLACED TWO NASOPHARYNGEAL AIRWAYS ON A CARDIAC ARREST PATIENT (ROUTINE), AND WHEN THEY LATER PLACED A KING AIRWAY, THEY WERE UNABLE TO GET A DEFINITIVE ETC02 READING. THEY THEN PULLED THE KING AND DISCOVERED THE NPA HAD DISPLACED BACKWARD AND WAS BLOCKING THE KING FROM SEALING PROPERLY. THUS IS NOT THE FIRST TIME THIS HAS HAPPENED. THE HOSPITAL CONTACTED THE FIRE DEPARTMENT PREVIOUSLY TO REPORT A SIMILAR INCIDENT ON (B)(6) 2018. (NOTE--THE CASE WAS 8 MONTHS PRIOR AND WE WERE UNAWARE UNTIL THE OTHER CASE WAS REVIEWED). THE ICU STAFF WAS UNABLE TO CLEAR THE PATIENTS GURGLING RESPIRATIONS THROUGH SUCTIONING, AND DISCOVERED A SIMILARLY DISPLACED NPA 10 DAYS AFTER ADMISSION. UPON INVESTIGATION, IT WAS NOTED THAT THE CLEAR CURAPLEX NPAS THAT WE WERE GETTING THROUGH BOUND TREE HAVE A MUCH SMALLER FLANGE THAN PREVIOUS SUPPLIERS. WE ARE PULLING THE CURAPLEX NPAS FROM SERVICE TO THE EXTENT THAT WE HAVE OTHER OPTIONS AVAILABLE. IT IS OUR OPINION THAT THE DISPLACED NPA COULD CONCEIVABLY ENTIRELY OBSTRUCT THE PATIENT'S AIRWAY. FD

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8759793 ↗

Other AIRWAY, NASOPHARYNGEAL devices

All AIRWAY, NASOPHARYNGEAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page