Skip to content

HScribe HSCRIBE-6AC-XXXXBRecalls & reported problems

Mortara Instrument, Inc. · Holter Analysis System · Class 2

All models of HScribe (8)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K004017); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Oct 11, 2024Class IIOpen, Classified

    There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.

    Z-0320-2025 · Root cause: Software design · Recall details · FDA record ↗

FDA clearance and classification

Cleared through 510(k) K004017 on Jun 11, 2001 (decision: Substantially Equivalent). Submitted by Mortara Instrument, Inc. as “H-Scribe”. The FDA review took 166 days. FDA record ↗

The FDA regulates product code MLO (electrocardiograph, ambulatory, with analysis algorithm) under 21 CFR 870.2800 as a Class II device. 21 CFR 870.2800 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K004017. One submission often covers many models, so check the product and lot details in each record.

  • Z-0323-2025 · Oct 11, 2024 · Open, Classified
    Hillrom Welch Allyn Vision Express Holter Analysis System; Electrocardiograph

Common questions about the HScribe HSCRIBE-6AC-XXXXB

Has the HScribe HSCRIBE-6AC-XXXXB been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0320-2025 (Class II), began on Oct 11, 2024: There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the HScribe HSCRIBE-6AC-XXXXB FDA approved?

It was cleared through 510(k) K004017 on Jun 11, 2001. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the HScribe HSCRIBE-6AC-XXXXB, and is it still sold?

The FDA's device database lists Mortara Instrument, Inc. as the company responsible for it (brand HScribe, model HSCRIBE-6AC-XXXXB). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Electrocardiograph, Ambulatory, With Analysis Algorithm devices

All Electrocardiograph, Ambulatory, With Analysis Algorithm devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page