Skip to content

Arcus Staple System ARC-1310KRecalls & reported problems

Medartis Inc. · 13mm X 10mm Arcus Staple Kit · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00817701020196), received between Jan 15, 2020 and Jan 15, 2020 — classified as 1 malfunction.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • No Apparent Adverse Event1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 15, 2020No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT DURING A SURGERY WITH THE ARCUS STAPLE SYSTEM, THE DRILL WAS TIGHT IN THE DRILL GUIDE AND THE DRILL GUIDE BENT. DRILLING WAS ABLE TO BE COMPLETED. A STAPLE WAS PLACED AND THE SURGEON ATTEMPTED TO USE THE STAPLE INSERTER AS A TAMP. THE SURGEON COULD NOT RE-ENGAGE THE INSERTER WITH THE STAPLE. THIS OCCURED WITH TWO DIFFERENT STAPLES AND INSERTERS AND THE STAPLE COULD NOT BE FULLY SEATED AGAINST THE BONE. THE SURGEON REMOVED THE ARCUS STAPLE AND COMPLETED THE SURGERY WITH A MAX VPC SCREW. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9592632 ↗

Other Staple, Fixation, Bone devices

All Staple, Fixation, Bone devices →

Sources and updates