Balt USA LLC · Magic Infusion Catheter is intended to operate for regional infusion of contrast materials into selected vessels in the neurovasculature. Magic Infusion Catheters are intended to facilitate access through distant, tortuous vasculature. Progressive suppleness ranging from a highly flexible tip to a semi-rigid proximal section allows the catheter to be advanced by the physician. It is not intended for use in the coronary vasculature. · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00818053022586), received between Jun 26, 2019 and Sep 11, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture2
  • Break1
  • Fluid/Blood Leak1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • Thromboembolism1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 11, 2026Material Rupture

    It was reported that:\\\"The physician was treating what appeared to be a mycotic right SCA aneurysm that formed quickly. This patient had just been treated for a ruptured basilar tip aneurysm in the previous days. She had other aneurysms that were also forming. Patient had some deficits before this procedure. The physician accessed right SCA with hybrid .007J and 1.2F Magic. Once Magic catheter reached target spot. Selective runs with contrast verfified position. Catheter prepped with D5 and a 4:1 ratio of glue was mixed and injected using a 5cc syringe. Glue didn't seem to be moving and doctor switched to a 1cc syringe of D5 and the Magic catheter ruptured proximally sending glue into the other branches of the PCA. The patient experience some severe posterior strokes but is in stable condition. The catheter is available for return and inspection.\"

    Patient problems: Thromboembolism

    MAUDE report 50738002 ↗
  • MalfunctionJun 30, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT: "A TUMOR EMBO WHERE A 1.2F MAGIC RUPTURED WHILE INJECTING NBCA. USING A 1CC SYRINGE TO INJECT THE GLUE."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10213603 ↗
  • MalfunctionJun 26, 2019Break; Fluid/Blood Leak; Break; Fluid/Blood Leak

    IT WAS REPORTED THAT: "I WAS NOT PRESENT AT THE CASE, BUT ON MY VISIT TO THE HOSPITAL ON (B)(6), DR. (B)(6) INFORMED ME THAT IN THE CASE ON (B)(6) THAT ONE OF THE TWO MAGIC CATHETERS USED RUPTURED NEAR THE TIP. THEY DID NOT SAVE THE CATHETER FOR INSPECTION. DR. (B)(6) COMMENTED THAT THE RUPTURE WAS SO CLOSE TO THE TIP, THAT IT DID NOT ADVERSELY EFFECT THE EMBOLIZATION OR PATIENT SAFETY. DR. (B)(6) SHARED THE LOT NUMBERS WITH ME, BUT WAS UNCERTAIN WHICH OF THE TWO CATHETER STICKERS BELONGED TO THE CATHETER THAT RUPTURED. DR. (B)(6) ADDED THAT THE PHYSICIAN WHO INJECTED THE GLUE USED A 3CC MEDALLION AND THAT THEY DID NOT HAVE A LOT OF EXPERIENCE USING GLUE."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8735214 ↗

FDA clearance and classification

Cleared through 510(k) K023351 on Feb 20, 2003 (decision: Substantially Equivalent). Submitted by Advanced Interventional Technologies, LLC as “Magic Infusion Catheter”. The FDA review took 136 days. FDA record ↗

The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Magic Infusion Catheter MAGIC1.2F

Has the Magic Infusion Catheter MAGIC1.2F been recalled?

No FDA recall record in our data names this exact device identifier (DI 00818053022586), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Magic Infusion Catheter MAGIC1.2F?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jun 26, 2019 and Sep 11, 2026. In those reports, the most common specific device problems are “Material Rupture” (2), “Break” (1) and “Fluid/Blood Leak” (1); the most common patient problem is “Thromboembolism” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Magic Infusion Catheter MAGIC1.2F?

None of the reports are classified as deaths, 1 as an injury and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Magic Infusion Catheter MAGIC1.2F FDA approved?

It was cleared through 510(k) K023351 on Feb 20, 2003. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Magic Infusion Catheter MAGIC1.2F, and is it still sold?

The FDA's device database lists Balt USA LLC as the company responsible for it (brand Magic Infusion Catheter, model MAGIC1.2F). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, Continuous Flush devices

All Catheter, Continuous Flush devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page