Magic Infusion Catheter MAGIC1.2FMRecalls & reported problems
Balt USA LLC · Magic Infusion Catheter is intended to operate for regional infusion of contrast materials into selected vessels in the neurovasculature. Magic Infusion Catheters are intended to facilitate access through distant, tortuous vasculature. Progressive suppleness ranging from a highly flexible tip to a semi-rigid proximal section allows the catheter to be advanced by the physician. It is not intended for use in the coronary vasculature. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00818053022593), received between Jun 18, 2019 and Aug 23, 2019 — classified as 1 death, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break2
- Fluid/Blood Leak2
- Off-Label Use1
Patient problems
- No Known Impact Or Consequence To Patient2
- Death1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
DeathAug 23, 2019Off-Label Use; Off-Label Use
MEDWATCH REPORT WAS RECEIVED VIA US MAIL ON AUGUST 8, 2019. DURING EMBOLIZATION OF AN ARTERIOVENOUS MALFORMATION (AVM} WITH ONYX, STAFF NOTED THE ONYX WAS LEAKING OUT OF THE MID-PORTION OF THE MAGIC 1, 2FM CATHETER DURING THE 3RD INJECTION. THE ONYX SHOULD EXIT THE CATHETER AT THE TIP, NOT FROM THE MID-PORTION. THE PHYSICIAN HAD TO EMPLOY A SOLITAIRE DEVICE AND USE ASPIRATION TO RETRIEVE THE ONYX. PER THE OP NOTE, APPROXIMATELY 60% WAS RECAPTURED. A 2ND PHYSICIAN WAS CALLED IN TO ASSIST. MDS WORKED APPROXIMATELY 2.5 HOURS TO CORRECT THE ISSUE. ATTEMPTS WERE EVENTUALLY STOPPED TO PREVENT FURTHER HARM TO THE PATIENT. AS A RESULT OF THE EVENT, THE PATIENT HAS A LARGE RIGHT FRONTAL STROKE. SINCE THE EVENT, HE HAS DEVELOPED AN INTRACRANIAL HEMORRHAGE REQUIRING A CRANIECTOMY FOR DECOMPRESSION & HEMATOMA EVACUATION. HE REMAINS IN ICU WITH A POOR NEUROLOGICAL EXAM AND POOR PROGNOSIS. WHAT WAS THE ORIGINAL INTENDED PROCEDURE? : PER STAFF REPORT, THE ONXY IS INJECTED VIA THE MAGIC L,2FM CATHETER. IT SHOULD EXIT THE CATHETER AT THE TIP END, NOT FROM THE MID-PORTION. THE CATHETER WAS RETAINED. WHAT PROBLEM DID THE USER HAVE (CHECK ALL THAT APPLY} :DEVICE FAILED (E.G. BROKE, COULDN'T GET IT TO
Patient problems: Death
MAUDE report 8928471 ↗MalfunctionJun 26, 2019Break; Fluid/Blood Leak; Break; Fluid/Blood Leak
IT WAS REPORTED THAT: "I DON'T KNOW ANY DETAILS ON THE CASE. THE CATHETER TIP BROKE IN 3 PIECES. PLEASE REPLACE CATHETER TO HOSPITAL."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8735215 ↗MalfunctionJun 18, 2019Break; Fluid/Blood Leak; Break; Fluid/Blood Leak
IT WAS REPORTED THAT: "TWO INCIDENCES INVOLVED. FIRST CASE INVOLVED A CATHETER TIP DETACHING DURING NBCA INJECTION. AT TIME OF THIS REPORT, I DID NOT GET TO SPEAK TO DOCTOR DIRECTLY. INFO IS SECOND HAND FROM TECHNOLOGIST. SECOND CASE INVOLVED A CATHETER THAT TECHNOLOGIST CLAIMED WAS 'SQUIRTING OUT FROM THE SIDE OF THE CATHETER DURING PREP.' PRODUCT WAS NOT USED IN PATIENT."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8710127 ↗
Other Catheter, Continuous Flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page