ECLIPSE 2L ECL2L6x7SNRecalls & reported problems
Balt USA LLC · ECLIPSE 2L balloon catheters are catheters with dual lumen arranged in parallel configuration. The guidewire lumen or the working lumen is DMSO compatible and can be used for infusion/delivery of diagnostic and therapeutic devices. The inflation lumen is used for the inflation and deflation of the balloon. The inflation and deflation of the balloon can be achieved independently of the presence of the guidewire. ECLIPSE 2L balloon catheter is equipped with a non-detachable low inflation pressure super compliant balloon welded to the distal end of the catheter. · Class 2
All models of ECLIPSE 2L (5)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00818053022937), received between Oct 28, 2020 and Mar 7, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Entrapment of Device2
- Gas/Air Leak1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 7, 2025Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT: "THE PHYSICIAN USED AN ECLIPSE 2L OCCLUSION BALLOON CATH TODAY THAT WAS DEFECTIVE. HE WAS INJECTING ONYX THROUGH THE CATHETER AND THE TIP BROKE OFF. CAN WE GET A REPLACEMENT FOR THIS? THE STAFF DID NOT SAVE THE CATHETER, BUT I DO HAVE THE PACKAGING. REF# (B)(4). LOT# F240800474". INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED. ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER - 21FEB2025: "- COULD YOU PLEASE TELL US IF THE INCIDENT OCCURRED WHEN THE ECLIPSE2L WAS INSIDE OR OUTSIDE THE PATIENT? YES. - COULD YOU PLEASE TELL US HOW THE ECLIPSE WAS PREPARED? DFU. - COULD YOU PLEASE TELL US IF YOU SEE ANY DAMAGES DURING THE PREPARATION OF THE DEVICE? NO - COULD YOU PLEASE TELL US IF YOU INJECT SOMETHING INSIDE THE CATHETER OTHER THAN ONYX? YES. - COULD YOU PLEASE INFORM US ABOUT THE PATIENT CONDITION TO DATE? FINE. - COULD YOU PLEASE TELL US HOW THE PHYSICIAN ENDED-UP THE CASE? AS USUAL. - COULD YOU PLEASE TELL US IF YOU SUCCEED TO PURGE THE BALLOON? / REMOVE THE AIR FROM THE BALLOON? YES".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21548712 ↗MalfunctionDec 14, 2022Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT: "ECLISPE 2L BALLOON STUCK IN CAST OF ONYX IN THE PATIENT. PHYSICIAN PULLED MULTIPLE TIMES AND THE BALLOON WOULDN'T RELEASE. THE CATHETER SNAPPED IN HALF AND MULTIPLE PIECES OF THE CATHETER HAD TO BE SNARED FROM THE PATIENT. THE DEBRIS WAS REMOVED AND THE BALLOON PORTION AND DISTAL CATHETER REMAINED IN THE PATIENT. PATIENT IS OKAY." 15NOV2022, ADDITIONAL INFORMATION OBTAINED BY SALES REP - " PATIENT IS HEALTHY AND HAD NO COMPLICATIONS AS A RESULT OF THIS PROCEDURE. PRODUCT HAS NOT YET BEEN RETURNED TO BALT FOR ANALYSIS DUE TO HOSPITAL'S RISK MANAGEMENT DEPARTMENT STILL HOLDING THE DEVICE IN QUARANTINE. WE ARE STILL TRYING TO OBTAIN IT FOR RETURN. TYPE OF CASE: BILATERAL MMA EMBOLIZATION WITH ONYX EMBOLIZED LEFT MMA USING 6X9 ECLIPSE WITH NO ISSUE EMBOLIZED RIGHT MMA (ANTERIOR DIVISION) USING 6X9 ECLIPSE WITHOUT ISSUE EMBOLIZED RIGHT MMA (POSTERIOR DIVISION) USING 6X7 ECLIPSE WITH SAME TECHNIQUE WHICH THEN HAD THE PROBLEM RETRACTING THE BALLOON AFTER DEFLATION AND ONYX INJECTION BALLOON WAS NEVER INFLATED FOR MORE THAN 1-1.5 MINUTES DURING INJECTION OF ONYX USED SOFIA 5F INTERMEDIATE CATHETER FOR SUPPORT AND ASSISTANCE IN RETRACTION IF NECESSARY. POSTERIOR DIVIS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15977961 ↗MalfunctionDec 15, 2020Gas/Air Leak; Physical Resistance/Sticking; Gas/Air Leak; Physical Resistance/Sticking
IT WAS REPORTED: "PHYSICIAN WAS USING THE ECLIPSE 2L TO DELIVER ONYX IN AN AVM. HE PREPPED THE BALLOON EXACTLY AS DFU DESCRIBES AND IT APPEARED TO BE IN GOOD CONDITION. THE BALLOON INFLATED IN THE ~2.2MM VESSEL AND IT APPEARED TO WELL OPPOSE THE VESSEL WALL. WHILE DELIVERING THE ONYX, THE BALLOON APPEARED TO BE DEFLATING SLIGHTLY WHICH CAUSED THE MD TO NEED TO RE-INFLATE. THE SAME THING HAPPENED 2 OR 3 MORE TIMES. THE STOPCOCK WAS APPROPRIATELY CLOSED AFTER EACH INFLATION. UPON COMPLETION OF THE CASE AND REMOVAL OF THE BALLOON, THE VESSEL WAS PERFORATED. THE MD FELT THAT ONYX REFLUXED AROUND THE BALLOON AND CAUSED RESISTANCE TO THE WITHDRAWAL."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11011936 ↗MalfunctionOct 28, 2020Break; Break
IT WAS REPORTED: "MMA EMBO USING ECLIPSE 6X7SN & ONYX. BALLOON WAS PUSHED FAIRLY DISTAL INTO SMALLER VESSEL AND WAS NEVER INFLATED. THE DISTAL END OF CATHETER IN TURN BECAME STUCK & BROKE OFF IN THE ONYX CAST. THE SURGEON WAS ABLE TO SNARE THE MAJORITY OF THE BROKEN CATHETER. THE PRODUCT WAS NOT ABLE TO BE SAVED TO EVALUATE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10752429 ↗
FDA clearance and classification
Cleared through 510(k) K183045 on Apr 30, 2019 (decision: Substantially Equivalent). Submitted by Balt USA, LLC as “Eclipse 2L”. The FDA review took 179 days. FDA record ↗
The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ECLIPSE 2L ECL2L6x7SN
Has the ECLIPSE 2L ECL2L6x7SN been recalled?
No FDA recall record in our data names this exact device identifier (DI 00818053022937), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ECLIPSE 2L ECL2L6x7SN?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Oct 28, 2020 and Mar 7, 2025. In those reports, the most common specific device problems are “Break” (3), “Entrapment of Device” (2) and “Gas/Air Leak” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ECLIPSE 2L ECL2L6x7SN?
None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ECLIPSE 2L ECL2L6x7SN FDA approved?
It was cleared through 510(k) K183045 on Apr 30, 2019. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the ECLIPSE 2L ECL2L6x7SN, and is it still sold?
The FDA's device database lists Balt USA LLC as the company responsible for it (brand ECLIPSE 2L, model ECL2L6x7SN). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, Intravascular Occluding, Temporary devices
- Pruitt F3 Carotid Shunt 2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary1 death158 reports
- GORE Molding Occlusion Balloon Catheter MOB37W. L. Gore & Associates, Inc. · Catheter, Intravascular Occluding, Temporary6 deaths152 reports
- Bridge Occlusion Balloon 590-001Spectranetics Corporation · Catheter, Intravascular Occluding, Temporary4 deaths57 reports
- Pruitt F3 Carotid Shunt e2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary53 reports
- Pruitt F3 Carotid Shunt 2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary46 reports
- Equalizer™ M001171100Boston Scientific Corporation · Catheter, intravascular occluding, temporary18 reports
- Pruitt F3 Carotid Shunt e2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary17 reports
- TRANSFORM M003SSC07070Stryker Corporation · Catheter, intravascular occluding, temporary17 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page