Skip to content

ECLIPSE 2L ECL2L6x20Recalls & reported problems

Balt USA LLC · ECLIPSE 2L balloon catheters are catheters with dual lumen arranged in parallel configuration. The guidewire lumen or the working lumen is DMSO compatible and can be used for infusion/delivery of diagnostic and therapeutic devices. The inflation lumen is used for the inflation and deflation of the balloon. The inflation and deflation of the balloon can be achieved independently of the presence of the guidewire. ECLIPSE 2L balloon catheter is equipped with a non-detachable low inflation pressure super compliant balloon welded to the distal end of the catheter. · Class 2

All models of ECLIPSE 2L (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00818053023002), received between Apr 30, 2021 and Aug 26, 2022 — classified as 1 death, 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Deflation Problem1
  • Entrapment of Device1

Patient problems

  • Foreign Body Embolism1
  • No Clinical Signs, Symptoms or Conditions1
  • Rupture1
  • Ruptured Aneurysm1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • DeathAug 26, 2022Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT: "DURING AN ONYX CASE, THE BALLOON FRACTURED AND ENDED UP OCCLUDING THE RIGHT ICA." ON (B)(6) 2022, ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: "WERE ANY ADDITIONAL PROCEDURES PERFORMED BY THE PHYSICIAN? THROMBECTOMY AFTER THE BALLOON FRACTURED." ON (B)(6) 2022 - ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: "COULD YOU PLEASE TELL US WHAT WAS THE PLANNED PROCEDURE? ONYX INJECTION THROUGH E2L FOR A FISTULA. COULD YOU PLEASE GIVE US INFORMATION ON PATIENT'S STATUS? WITHDREW CARE. CAN YOU PLEASE CONFIRM THAT THERE IS NO PATIENT INJURY? NO. WAS THE PATIENT'S ANATOMY TORTUOUS? YES."

    Patient problems: Foreign Body Embolism; Ruptured Aneurysm

    MAUDE report 15300831 ↗
  • InjuryJun 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT: "THE BALLOON, AFTER BEING CORRECTLY PURGED AND INFLATED ONCE IN THE AREA TO BE TREATED, FAILED TO BE INFLATED FOR THE SECOND TIME. THE DISTAL PART OF THE BALLOON GET INFLATED, BUT NOT THE PROXIMAL HALF OF IT. AS THE OPERATOR KEEP PUSHING IN ORDER TO INFLATE THE PROXIMAL PART, THE DISTAL PART GET OVER INFLATED INSTEAD OF THE PROXIMAL, THAT KEEPS DEFLATED. VASSEL BREAK DUE TO THE BALLOON OVERINFLATION. FORTUNATELY THE PATIENT SURVIVED AND WILL PROBABLY RECOVER." ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: "THEY PUT THE BALLOON ON THE TRASH, THEY SAID THAT THE BALLOON WAS DEFORMED AS THEY MANIPULATE IT AFTER THE CASE (ONCE OUT OF THE BODY OF THE PATIENT). HONESTLY, I CHECKED THE BALLOON JUST AFTER THEY PULL IT AWAY AND IT INFLATE AND DEFLATE CORRECTLY. SOME MINUTES AFTER, THE PHYSICIANS MANIPULATE IT AND AFTER SEVERAL TIME THEY PUT IT ON THE TRASH." "THEY INJECTED GLUE AFTER THE RUPTURE OF THE VASSEL TO STOP THE BLEEDING. THE BALLOON DID NOT RUPTURE, THE PROXIMAL PART SIMPLY DID NOT INFLATE. THE PROXIMAL PART WAS OVER-INFLATED BUT THE BALLOON DID NOT RUPTURE." "JUST ONE UNIT OF THAT BATCH NUMBER." "THE HEMORRHAGE WAS QUICKLY STOPPED BY THE INR AND THE BLOOD

    Patient problems: Rupture

    MAUDE report 14734034 ↗
  • MalfunctionApr 30, 2021Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT: "THE PHYSICIAN WAS POST DILATING A SURPASS AND HE COULD NOT GET THE ECLIPSE TO DEFLATE. THE TECH CALLED ME AND I ASKED THEM TO PULL TENSION ON THE BALLON AND TRY TO DEFLATE. THE BALLOON CAME DOWN AND IS AVAILABLE TO RETURN. HE FINISHED THE CASE WITH A TRANSFORM BALLOON. "

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11755436 ↗

Other Catheter, Percutaneous devices

All Catheter, Percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page