Optima Coil System OPTI0413CSFMXRecalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

All models of Optima Coil System (379)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00818053025129), received between Dec 23, 2021 and Mar 11, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

3 other recalls cover devices cleared under the same FDA submission (K200030); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 8, 2025Class IIOpen, Classified

    Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.

    Z-1988-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

  2. Jan 3, 2024Class IIOpen, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Premature Separation2
  • Fracture1
  • Migration1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Thromboembolism1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 11, 2026Migration; Premature Separation

    IT WAS REPORTED THAT: "THE PHYSICIAN WAS TREATING A LARGE OPTHALMIC ANEURYSM WITH A PRIMARY COILING TECHNIQUE. WE WERE ROUGHLY 4 COILS INTO TREATMENT WHEN WE TRIED TO USE A 4MMX13CM MAX COIL. WE DID MAKE A FEW ADJUSTMENTS OF THE COIL TO TRY TO GET IT TO SIT PROPERLY. UPON THE THIRD ADJUSTMENT, THE COIL KICKED THE MICRO OUT OF THE ANEURYSM. THE PHYSICIAN PULLED COIL BACK AND TRIED TO ADVANCE MICRO BACK INTO ANEURYSM. WHILE DOING THAT---THE COIL PRE-MATURELY DETATCHED. WE HAD TO USE SEVERAL TECHNIQUES TO RETRIEVE COIL AND COIL WAS ULTIMATELY RETRIEVED WITH A STENTRIEVER. THERE WAS MODERATE VESSEL TOURTOSITY BUT THE COIL SHOULD ABSOLUTELY BE MORE DURABLE THAT IT WAS. THE PRE-MATURE DETATCHED COIL DID END UP TRAVELING INTO THE MCA DURING RETRIEVEL ATTEMPTS--BECAUSE OF THIS, THE PATIENT ENDED UP HAVING CLOT AT THE M1 BIFURCATION THAT WAS REMOVED WITH A RED 43 ASPIRATION CATHETER." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.

    Patient problems: Thromboembolism

    MAUDE report 24575688 ↗
  • MalfunctionSep 16, 2022Fracture; Stretched; Fracture; Stretched

    IT WAS REPORTED THAT: "15 X 14 ICA ANEURYSM. TOOK 16 COMPLEX 18 AND OPTIMAX COILS. THE 17TH COIL WAS A 4X13 MAX BUT WAS HAVING TROUBLE GOING IN, MANIPULATED MC MANY TIMES, COIL WAS HALF IN ANEURYSM, PULLED GENTLY TO REMOVE AND STARTED TO UNRAVEL. AFTER 2 HOURS WE WERE ABLE TO SNARE WITH NO HARM TO PATIENT." 25AUG2022, ADDITIONAL INFORMATION: "CAN YOU CLARIFY WHAT HAPPENED IN THE INCIDENT? WAS THE IMPLANT STILL CONNECTED TO THE DELIVERY PUSHER? YES. WAS THE IMPLANT DETACHED USING THE CONTROLLER WHEN IT WAS HALF IN THE ANEURYSM? OR DID THE COIL PREMATURELY DETACH? THEY EVENTUALLY PULLED COIL ENOUGH FROM BACK END THAT IT BROKE OFF FROM PUSHER WIRE." 12SEP2022, ADDITIONAL INFORMATION: DISCUSSION WITH THE SALES REPRESENTATIVE PRESENT DURING THE CASE INDICATED THAT THE PHYSICIAN NOTICED THAT THE IMPLANT BECAME STRETCHED AND BROKE FROM THE DELIVERY PUSHER AFTER IT WAS NOTED THAT THERE WAS NO MORE MOVEMENT WITH THE IMPLANT WHILE STILL RETRACTING THE DELIVERY PUSHER WHEN ATTEMPTING TO MANIPULATE THE COIL TO BETTER SEAT THE IMPLANT. THE IMPLANT COIL WAS THEN SNARED OUT IN APPROXIMATELY 6 DIFFERENT PIECES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15438691 ↗
  • MalfunctionDec 23, 2021Premature Separation; Premature Separation

    IT WAS REPORTED THAT: "PREMATURE DETACHMENT OF THE COIL DURING COIL POSITIONING INTO THE EMBOLIZATION SITE. MICROCATHETER USED TO DELIVER THE COIL: EXCELSIOR 17XT STRAIGHT. COIL RETRIEVED WITH A LASSO SYSTEM." FOLLOW UP 10DEC2021 - ADDITIONAL INFORMATION RECEIVED FROM ISSUER: " WAS THERE TORTUROUS ANATOMY INVOLVED? NO." FOLLOW UP 13DEC2021: - ADDITIONAL INFORMATION RECEIVED FROM ISSUER: "WHERE DID THE IMPLANT PREMATURELY DETACH? DID IT OCCUR INSIDE OF THE MICROCATHETER? ABSOLUTELY!"

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13078526 ↗

FDA clearance and classification

Cleared through Special 510(k) K200030 on Feb 1, 2020 (decision: Substantially Equivalent). Submitted by Balt USA, LLC as “Optima Coil System”. The FDA review took 25 days. FDA record ↗

The FDA regulates product code KRD (device, vascular, for promoting embolization) under 21 CFR 870.3300 as a Class II device. 21 CFR 870.3300 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K200030. One submission often covers many models, so check the product and lot details in each record.

  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.
  • Z-0660-2022 · Jan 4, 2022 · Open, Classified
    The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil del

Common questions about the Optima Coil System OPTI0413CSFMX

Has the Optima Coil System OPTI0413CSFMX been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1988-2025 (Class II), began on May 8, 2025: Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Optima Coil System OPTI0413CSFMX?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Dec 23, 2021 and Mar 11, 2026. In those reports, the most common specific device problems are “Premature Separation” (2), “Fracture” (1) and “Migration” (1); the most common patient problem is “Thromboembolism” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Optima Coil System OPTI0413CSFMX?

None of the reports are classified as deaths, 1 as an injury and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Optima Coil System OPTI0413CSFMX FDA approved?

It was cleared through Special 510(k) K200030 on Feb 1, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Optima Coil System OPTI0413CSFMX MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Optima Coil System OPTI0413CSFMX, and is it still sold?

The FDA's device database lists Balt USA LLC as the company responsible for it (brand Optima Coil System, model OPTI0413CSFMX). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.