Optima Coil System OPTI0208CSS10 (REF N/A)Recalls & reported problems
Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant
All models of Optima Coil System (379)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00818053025815), received between Jul 12, 2023 and May 14, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Inaccurate Information” (1 report).
3 FDA recall records name this device's identifier, 3 still open.
1 other recall cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 8, 2025Class IIOpen, Classified
Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
Z-1988-2025 · Root cause: Vendor change control · Recall details · FDA record ↗
- Jan 3, 2024Class IIOpen, Classified
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
- Sep 7, 2022Class IIOpen, Classified
Product pouch label does not match up with carton label
Z-0103-2023 · Root cause: Error in labeling · Recall details · FDA record ↗
Most reported problems
Device problems
- Inaccurate Information1
- Premature Separation1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 14, 2026Premature Separation
IT WAS REPORTED THAT: "THE PHYSICIAN WAS TREATING AN AVM ON A 3.5-YEAR-OLD PATIENT. HE TREATED THE FEEDING VESSEL WITH 3 OPTIBLOCK MINI COILS AND HAD A GREAT RESULT. HE PERFORMS RESEARCH AND TAKES A COIL AND MOVES IT AROUND IN THE AREA OF THE AVM SO HE CAN TAKE THE CELLS AND DO TESTING FOR RESEARCH. UNFORTUNATELY, WHILE HE WAS ALLOWING THE COIL TO MOVE INSIDE THE AVM, THE COIL DETACHED. PHYSICIAN UNDERSTANDS IT MAY BE DUE TO THE STRESS THAT WAS PUT ON THE COIL. HE WAS ABLE TO SNARE THE COIL AND DID NOT CAUSE ANY NEGATIVE HARM TO PATIENT AND WAS ABLE TO RESUME CASE SUCCESSFULLY. USED A HEADWAY DUO 167 MICRO." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25174428 ↗MalfunctionMar 15, 2024Stretched; Stretched
IT WAS REPORTED THAT: "<PATIENT INFORMATION> SEX: 75 YEARS OLD FEMALE DISEASE NAME: 7MM OF AN IN IC-PC PROCEDURE: PLANNED PROCEDURE MICRO CATHETER: GREACH/TOKAI MEDICAL ASSIST STENT: UNKNWON Y CONNECTOR: UNKNOWN USED COILS: AVENIR/JLL <DESCRIPTION> THE PHYSICIAN EXPERIENCED 2 UNRAVEL EVENTS. [?]OPTIMAX: 7MM-20CM(REPORTED AS YA2023-1095) [?]OPTIMA 2MM-8CM(YA2023-1097) YA2023-1097: AFTER YA2023-1095, 5 OPTIMAX COILS AND 5 AVENIR COILS WERE PLACED WITH TRANS CELL TECHNIQUE. THE PHYSICIAN CHOSEN 2-8MM OPTIMA. DURING INSERTION, 4 CM PROXIMAL SIDE OF THE IMPLANT COIL COULD NOT BE PLACED TARGET LESION. MULTIPLE ATTEMPTS OF REPOSITIONING RESULT IN UNRAVEL. THE PHYSICIAN TRIED TO RETRIEVE UNRAVELED COIL BUT THE MASS OF THE COIL BECAME TRAPPED IN THE VESSEL AROUND C4 AND DIFFICULT TO RETRIEVE. THE PHYSICIAN DID EVERYTHING TO RECOVER THE MASS OF COILS, WHICH FELL FROM THE C4 POSITION TO THE PETRAS AREA, BUT IT WAS UNABLE TO RETRIEVE. THE PHYSICIAN HAD TO PLACE LVIS STENT(ONLY 5.5MM WAS AVAILABLE DESPITE THE VESSEL WAS 4.5MM) TO PRESS THE MASS OF COIL INTO THE VESSEL WALL. THE PROCEDURE WAS COMPLETED WITH PTA USING SHORYU (BALLOON)." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18912518 ↗MalfunctionJul 12, 2023Inaccurate Information; Inaccurate Information
IT WAS REPORTED THAT: "THIS IS TO INFORM YOU THAT WE FOUND A PRODUCT WHOSE INFORMATION ON THE POUCH LABEL DIFFERS FROM THE LABEL ON THE BOX DURING INCOMING INSPECTION. AS OF NOW THE AMOUNT OF PRODUCT WITH DISCREPANCY IS 1. BOX: OPTI0208CSS10 F220601068 POUCH: OPTI0308CSS10 F220601089".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17313322 ↗
FDA clearance and classification
Cleared through 510(k) K172390 on Feb 18, 2018 (decision: Substantially Equivalent). Submitted by Blockade Medical, LLC (D.B.A. Balt Usa) as “Optima Coil System”. The FDA review took 194 days. FDA record ↗
The FDA regulates product code HCG (device, neurovascular embolization) under 21 CFR 882.5950 as a Class II device. 21 CFR 882.5950 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.
- Z-1680-2022 · Jul 26, 2022 · Open, ClassifiedOptima Coil System Model OPTI0520CSF10. For endovascular embolization.
Common questions about the Optima Coil System OPTI0208CSS10
Has the Optima Coil System OPTI0208CSS10 been recalled?
Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-1988-2025 (Class II), began on May 8, 2025: Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Optima Coil System OPTI0208CSS10?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jul 12, 2023 and May 14, 2026. In those reports, the most common specific device problems are “Inaccurate Information” (1), “Premature Separation” (1) and “Stretched” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Optima Coil System OPTI0208CSS10?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Optima Coil System OPTI0208CSS10 FDA approved?
It was cleared through 510(k) K172390 on Feb 18, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Optima Coil System OPTI0208CSS10 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Optima Coil System OPTI0208CSS10, and is it still sold?
The FDA's device database lists Balt USA LLC as the company responsible for it (brand Optima Coil System, model OPTI0208CSS10). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Device, Neurovascular Embolization devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
- POD RBYPODJ60Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION131 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page