Plato 17-160-000Recalls & reported problems
Scientia Vascular LLC · Microcatheter straight tip · Class 2
All models of Plato (3)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00818075010455), received between Jan 5, 2026 and Feb 20, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K210601, K223560); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Feb 1, 2023Open, Classified
Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response.
Z-1249-2023 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 20, 2026Break
IT WAS REPORTED THAT DURING A NEUROVASCULAR PROCEDURE THE PHYSICIAN ACCESSED THE MMA WITH A PLATO 17 (PL17-160-000) THROUGH A PENUMBRA 5F BER CATHETER. THE PHYSICIAN DID A MICRO INJECTION AND WHEN THEY LOOKED DOWN THE HUB OF THE PLATO 17 HAD BROKEN OFF. THE BROKEN DEVICE WAS REMOVED AND ANOTHER PLATO 17 WAS USED TO FINISH THE CASE. THE PATIENT WAS NOT HARMED AS A RESULT OF THE MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24407816 ↗MalfunctionFeb 11, 2026Break
IT WAS REPORTED THAT AFTER A SPINAL EMBOIZATION PROCEDURE USING A PLATO 17 CATHETER TO INJECT PARTICLES THE CATHETER SNAPPED IN HALF. THE PHYSICIAN SAID ONE SECOND IT WAS TOGETHER AND THE NEXT IT WAS IN TWO PIECES. THEY ALSO NOTED THEY DID NOT PUT ANY EXTRA PRESSURE ON THE CATHETER THAN NORMAL. THERE WAS NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24319463 ↗MalfunctionJan 5, 2026Break
DURING A NEUROVASCULAR PROCEDURE, THE DOCTOR WAS USING A PL17-160-000 CATHETER. THE DOCTOR WAS ABLE TO ACCESS THE MMA WITH THE CATHETER AND WHILE IN THE PROCESS OF DOING AN ANGIOGRAM USING A 3CC MEDALLION SYRING AND THE CATHETER BROKE AT THE HUB. THE CATHETER WAS REPLACED WITH ANOTHER PRODUCT AND THE PROCEDURE WAS COMPLETED WITHOUT INCIDENT OR ANY HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23961152 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K210601, K223560. One submission often covers many models, so check the product and lot details in each record.
- Z-1235-2025 · Jan 16, 2025 · Open, ClassifiedPlato 17 Microcatheter REF PL17-160-090 To assist in delivery of therapeutic devices and embolization materials and infusion of diagnostic agents to the neuro and peripheral vascu
Other Catheter, Percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page