Skip to content

Vive MOB1019PNKRecalls & reported problems

Vive Health LLC · Knee Walker (Pink) · Class 1

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00818323024937), received between Jan 27, 2025 and Jan 27, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Mechanical Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem2

Patient problems

  • Hemorrhage/Blood Loss/Bleeding2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 27, 2025Mechanical Problem; Mechanical Problem

    THE CUSTOMER RECENTLY HAD FOOT SURGERY, AND WHILE USING THE ITEM, SHE FELL AND HIT THE FOOT THAT UNDERWENT SURGERY. SHE REPORTS THAT THIS HAPPENED BECAUSE THE ITEM IS TOO UNSTABLE AND DANGEROUS.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21241187 ↗
  • InjuryJan 27, 2025Mechanical Problem; Mechanical Problem

    THE COMPANY ALREADY REPLACED THE FAULTY UNIT WITH ONE IN STOCK, SHE WAS ABOUT TO USE THE KNEE WALKER UNTIL THE BREAKS FAIL AND SHE FELT DOWN, WHEN HIS HUSBAND TRY TO PICK HER UP AND SEE WHAT HAPPENED HE END ALSO FALLING DOW. THEY DECIDE NOT GOING INTO THE HR BUT SHE HAS BRUISES.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21241183 ↗

Other Walker, Mechanical devices

All Walker, Mechanical devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.