Skip to content

neoWave LS 23-LCT-2211Recalls & reported problems

Xenix Medical LLC · LUMBAR CAGE, 10W x 22L x 11H, 8 DEG, STRAIGHT · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818345024298), received between Jun 14, 2024 and Jun 14, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration or Expulsion of Device1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 14, 2024Migration or Expulsion of Device; Migration or Expulsion of Device

    IT WAS REPORTED THAT A SURGEON HAD A CASE IN WHICH THE LUMBAR CAGE IMPLANT HAD SHIFTED POST-OPERATIVELY. THE REP SAID THAT THE INTERBODY AT L2-3 MIGRATED POSTERIORLY. IMAGING REVEALED THE POSTERIOR MIGRATED IMPLANT AS WELL AS SIGNS OF POTENTIAL SCREW LOOSENING AT THE L2 SCREW. DURING THE REVISION SURGERY, IT WAS DISCOVERED THAT THE L2 PEDICLE SCREW CAPS IN THE PEDICLE SCREW TULIP USED FOR FIXATION HAD LOOSENED ALLOWING THE ROD TO SHIFT. NOTE - THE PEDICLE SCREW SYSTEM IS FROM ANOTHER MANUFACTURER (UNKNOWN) AND IS NOT A XENIX MEDICAL PRODUCT. DURING THE REVISION PROCEDURE, THE LUMBAR CAGE IMPLANT WAS SHIFTED BACK INTO POSITION AND THE SET SCREW WAS REPLACED AND THE CONSTRUCT WAS TIGHTENED.

    Patient problems: Pain

    MAUDE report 19543337 ↗

Other Intervertebral Fusion Device With Bone Graft, Lumbar devices

All Intervertebral Fusion Device With Bone Graft, Lumbar devices →

Sources and updates