Skip to content

neoWave LS 23-LCT-2509Recalls & reported problems

Xenix Medical LLC · LUMBAR CAGE, 10W x 25L x 9H, 8 DEG, STRAIGHT · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818345024380), received between Nov 16, 2020 and Nov 16, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Migration1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 16, 2020Migration; Migration

    INITIAL SURGERY WAS PERFORMED ON (B)(6) 2020 FOR PLIF SINGLE LEVEL FUSION AT L5-S1 WITHOUT FINAL POSTERIOR COMPRESSION OF SUPPLEMENTAL FIXATION. POSTERIOR MOVEMENT OF INTERBODY IMPLANTS DISCOVERED ABOUT (6) WEEKS AFTER ORIGINAL SURGERY IMPLANTATION. REVISION SURGERY PERFORMED TO TAMP THE INTERBODY IMPLANTS BACK INTO THE DISC SPACE ANTERIORLY AND THEN COMPRESS ON THE SCREWS TO INCREASE THE LORDOSIS AND INCREASE ENDPLATE CONTACT WITH IMPLANT. SURGICAL REVISION PROCEDURE WAS COMPLETED WITHOUT ISSUE AND PATIENT DISCHARGED IN NORMAL COURSE.

    Patient problems: Pain

    MAUDE report 10848596 ↗

Other Intervertebral Fusion Device With Bone Graft, Lumbar devices

All Intervertebral Fusion Device With Bone Graft, Lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page