SOL-CARE™ Luer Lock Safety Syringe with Exchangeable Needle 160072IMRecalls & reported problems

Sol-Millennium Medical, Inc. · Used to inject fluids into, or withdraw fluids from the body. In addition, the device is designed to aid in the prevention of needle stick injuries. 10ml 21g x 1 1/2" exchangeable needle · Class 2

All models of SOL-CARE Luer Lock Safety Syringe with Exchangeable Needle (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00818392015041), received between May 16, 2025 and Mar 24, 2026 — classified as 18 malfunctions.

The most frequently reported specific device problem is “Decoupling” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 10 reportsJun 2025: 4 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Decoupling5
  • Fluid/Blood Leak3
  • Retraction Problem3
  • Break2
  • Contamination2
  • Crack1
  • Defective Component1
  • Difficult to Fold, Unfold or Collapse1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • Needle Stick/Puncture2

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2026Break

    CUSTOMER REPORTED THE SYRINGE WAS CRACKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24673966 ↗
  • MalfunctionFeb 17, 2026Break

    CUSTOMER REPORTED A CRACK WAS FOUND IN THE SYRINGE BARREL DURING USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24362469 ↗
  • MalfunctionDec 2, 2025Contamination; Contamination

    CUSTOMER REPORTED COTTON FIBERS WERE FOUND INSIDE THE SYRINGE, MAKING IT UNUSABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23692121 ↗
  • MalfunctionDec 1, 2025Leak/Splash; Premature Activation; Leak/Splash; Premature Activation

    CUSTOMER REPORTED DURING BLOOD COLLECTION, THE NEEDLE SUDDENLY RETRACTED, CAUSING BLOOD LEAKAGE. CUSTOMER REPORTED THE NEEDLE DETACHED, CAUSING LEAKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23679891 ↗
  • MalfunctionJun 20, 2025Crack; Crack

    CUSTOMER REPORTED THAT AFTER USING THE EMPTY SYRINGE FOR BLOOD COLLECTION, IRREGULAR CRACKSWERE FOUND ON THE OUTER BARREL OF THE SYRINGE, INDICATING A NEED TO ENHANCE QUALITY CONTROL. CUSTOMER REPORTED THAT THE SYRINGE BARREL IS CRACKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22294681 ↗
  • MalfunctionJun 20, 2025Contamination; Defective Component; Contamination; Defective Component

    CUSTOMER REPORTED THAT THEY WHEN OPENING THE NEEDLE FOR USE, IT WAS FOUND THAT THERE WAS FOREIGN MATTER (PLASTIC BLOCK) INSIDE THE EMPTY NEEDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22294680 ↗
  • MalfunctionJun 11, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    CUSTOMER COMPLAINT NO. (B)(4). CUSTOMER REPORTED THAT WHILE USING THE SYRINGE TO DRAW MEDICATION, IT SUDDENLY BECAME IMPOSSIBLE TO PUSH FORWARD OR PULL BACK. THE SAME ISSUE OCCURRED CONSECUTIVELY WITH TWO 10CC SYRINGES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22182343 ↗
  • MalfunctionJun 11, 2025Fluid/Blood Leak; Fluid/Blood Leak

    CUSTOMER COMPLAINT NO. (B)(4). CUSTOMER REPORTED THAT WHEN DRAWING MEDICINE WITH AN EMPTY NEEDLE, THE MEDICINE FLOWSBACKWARD. CUSTOMER COMPLAINT NO: (B)(4). CUSTOMER REPORTED THAT AFTER DRAWING THE MEDICATION, LEAKAGE OCCURS FROM THE BLACK RUBBERSTOPPER DURING INJECTION. CUSTOMER COMPLAINT NO: (B)(4). CUSTOMER REPORTED THAT WHEN DRAWING THE MEDICATION, IT LEAKS FROM THE NEEDLE HUB. CUSTOMER COMPLAINT NO: (B)(4). CUSTOMER REPORTED THAT THE SYRINGE MALFUNCTIONED, WITH LEAKAGE OCCURRING FROM THE BACK OF THERUBBER STOPPER. CUSTOMER COMPLAINT NO: (B)(4). CUSTOMER REPORTED THAT THE BACK END OF THE SYRINGE IS LEAKING. CUSTOMER COMPLAINT NO: (B)(4). CUSTOMER REPORTED THAT THE MEDICATION LEAKS OUT FROM THE NEEDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22182335 ↗
  • MalfunctionMay 29, 2025Retraction Problem; Retraction Problem

    "CUSTOMER COMPLAINT NO: (B)(4). CUSTOMER REPORTED THAT THE NEEDLE RETRACTED AUTOMATICALLY, CAUSING A NEEDLESTICK INJURY TO A HOSPITAL NURSE. ADDITIONALLY, THE COMPLAINT FREQUENCY FOR THIS BATCH IS EXCESSIVELY HIGH, AND A FULL BATCH REPLACEMENT IS REQUIRED. CUSTOMER COMPLAINT NO: (B)(4). CUSTOMER REPORTED THAT WHEN FIRST OPENED AND USED, THE NEEDLE SUDDENLY RETRACTED, CAUSING THE NURSING STAFF TO ALMOST EXPERIENCE A NEEDLESTICK INJURY. "

    Patient problems: Needle Stick/Puncture

    MAUDE report 22106380 ↗
  • MalfunctionMay 22, 2025Decoupling; Decoupling

    CUSTOMER REPORTED THAT THE NEEDLE SUDDENLY DETACHED DURING THE MEDICATION EXTRACTION PROCESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22073541 ↗
  • MalfunctionMay 16, 2025Fluid/Blood Leak; Fluid/Blood Leak

    CUSTOMER REPORTED WHEN USING A 10ML SAFETY INJECTION NEEDLE TO DRAW MEDICATION, LEAKAGE OCCURRED AT THE CONNECTION BETWEEN THE NEEDLE AND THE SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035747 ↗
  • MalfunctionMay 16, 2025Decoupling; Decoupling

    CUSTOMER REPORTED THAT THE NEEDLE DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035734 ↗
  • MalfunctionMay 16, 2025Decoupling; Decoupling

    CUSTOMER REPORTED THAT THE NEEDLE DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035731 ↗
  • MalfunctionMay 16, 2025Retraction Problem; Retraction Problem

    CUSTOMER REPORTED THAT THE WHILE USING A 10ML SAFETY SYRINGE TO DRAW MEDICATION, THE NEEDLE HUB SUDDENLY RETRACTED AND DETACHED, CAUSING CONTAMINATION OF THE MEDICATION, MAKING IT UNUSABLE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035728 ↗
  • MalfunctionMay 16, 2025Fluid/Blood Leak; Fluid/Blood Leak

    CUSTOMER REPORTED THAT THE AFTER DRAWING THE MEDICATION, THE NEEDLE DETACHED DURING ADMINISTRATION, CAUSING MEDICATION SPILLAGE. THE MANUFACTURER MUST COMPENSATE (B)(6) FOR THE MEDICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035719 ↗
  • MalfunctionMay 16, 2025Decoupling; Decoupling

    CUSTOMER REPORTED THAT DURING THE MEDICATION EXTRACTION PROCESS, THE NEEDLE AUTOMATICALLY DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035672 ↗
  • MalfunctionMay 16, 2025Decoupling; Decoupling

    CUSTOMER REPORTED THAT THE FRONT END OF THE SYRINGE BECAME LOOSE, MAKING IT IMPOSSIBLE TO DRAW MEDICATION AND CAUSING A NEEDLESTICK INJURY TO THE NURSING STAFF

    Patient problems: Needle Stick/Puncture

    MAUDE report 22035666 ↗
  • MalfunctionMay 16, 2025Retraction Problem; Retraction Problem

    CUSTOMER REPORTED UPON OPENING THE PACKAGING, IT WAS FOUND THAT THE NEEDLE HAD ALREADY RETRACTED AND WAS UNUSABLE. THIS ISSUE HAS OCCURRED MULTIPLE TIMES AND REQUIRES ENHANCED QUALITY CONTROL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035646 ↗

Other SYRINGE, ANTISTICK devices

All SYRINGE, ANTISTICK devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page