Skip to content

FlareHawk ASY-00029Recalls & reported problems

Integrity Implants Inc. · Shell, Lordotic 25 mm, FlareHawk 9 · Class 2 · Implant

All models of FlareHawk (29)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00818613021264), received between Dec 16, 2020 and Jun 1, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1
  • Migration or Expulsion of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 1, 2024Migration or Expulsion of Device; Migration or Expulsion of Device

    ON 5/3/2024, IT WAS REPORTED TO ACCELUS THAT DURING A CASE, A FLAREHAWK 9 CAGE WAS IMPLANTED BUT MIGRATED TO ANTERIOR OF VERTEBRAL BODY (VB) DURING BONE GRAFT PACKING STAGE. THE CAGE HAD TO BE REMOVED AND REPLACED WITH ANOTHER RESULTING IN A 180 MIN (3 HOUR) DELAY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND NO OTHER PATIENT HARM OR IMPACT WAS REPORTED. THE EVENT HAPPENED INTRA-OP. IT WAS REPORTED THAT THE SURGICAL TECHNIQUE FOR THE PRODUCT WAS UTILIZED AND THE INFORMATION PROVIDED INDICATED THAT THE IMPLANT WAS DEPLOYED CORRECTLY. THE EVENT OCCURRED DURING THE BONE GRAFT PACKING STAGE WHERE THE IMPLANT CONSTRUCT WAS REPOSITIONED. THERE IS POTENTIAL FOR USER ERROR TO LEAD TO THIS EVENT IF THE IMPLANT WAS NOT SIZED PROPERLY BY THE USER OR IF DURING THE BONE GRAFT PACKING EXCESSIVE FORCE IS APPLIED. NO OTHER HARMS BESIDES THE 3 HOUR SURGICAL DELAY WERE REPORTED, HOWEVER THE IMPLANT CONSTRUCT MIGRATING OUT OF THE VERTEBRAL BODY COULD CAUSE A SERIOUS INJURY IF IT WERE TO RECUR. THE MDR SUBMISSION SHALL BE UPDATED PENDING COMPLETION OF THE COMPLAINT INVESTIGATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19441908 ↗
  • MalfunctionDec 16, 2020Migration; Migration

    UPON REASSESSMENT OF THE REPORTED EVENT, IT WAS DETERMINED THAT NO SERIOUS INJURY OCCURRED AS WAS INITIALLY REPORTED AS A RESULT OF THIS PRODUCT COMPLAINT. THERE WAS NO REPORTED SURGICAL DELAY, DEATH OR SERIOUS ILLNESS OR INJURY ASSOCIATED WITH THIS COMPLAINT. THE INITIAL REPORT WAS SUBMITTED IN ERROR AND SHOULD BE VOIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11023116 ↗

Other Intervertebral Fusion Device With Bone Graft, Lumbar devices

All Intervertebral Fusion Device With Bone Graft, Lumbar devices →