Light Adjustable Lens (LAL) 60007-0195Recalls & reported problems
Rxsight, Inc. · LAL+, Light Adjustable Lens, Sterile, Packaged, 19.5 Diopter · Class 3 · Implant
All models of Light Adjustable Lens (LAL) (33)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This RXSIGHT Light Adjustable Lens+ is an implantable eye lens whose power is fine-tuned after surgery using a light delivery device. Most of the 29 reports describe blurred vision, double vision and other visual disturbances, and in many cases the lens being removed and replaced, with 29 injuries and no deaths. Reports fell from 22 in the prior 12 months to 5 in the most recent 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 29
- Deaths reported
- 0
- Injuries reported
- 29
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 00818806021576), received between Sep 6, 2024 and May 22, 2026 — classified as 29 injuries.
The most frequently reported specific device problem is “Optical Decentration” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem23
- Optical Decentration3
- Use of Device Problem2
- Optical Problem1
Patient problems
- Visual Disturbances14
- Blurred Vision12
- Visual Impairment5
- Capsular Bag Tear2
- Diplopia/Double Vision2
- Unspecified Eye / Vision Problem2
- Eye Pain1
Recent adverse event reports
The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.
InjuryMay 22, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+) WAS EXPLANTED FROM THE RIGHT EYE DESPITE HAVING GOOD VISION POST-IMPLANTATION. THE TREATING PHYSICIAN STATED THAT THE PATIENT WANTED A MULTIFOCAL IOL DUE TO THE POSSIBILITY OF OBTAINING GREATER RANGE OF VISION. THE PATIENT IS REPORTED TO BE DOING WELL POST-OPERATIVELY. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Unspecified Eye / Vision Problem
MAUDE report 25257251 ↗InjuryMay 14, 2026Optical Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+) WAS EXPLANTED FROM THE LEFT EYE DUE TO DIPLOPIA. A NEW LAL WAS IMPLANTED AS A REPLACEMENT. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Diplopia/Double Vision
MAUDE report 25174016 ↗InjuryApr 6, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A BILATERALLY IMPLANTED PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED FROM THE LEFT EYE DUE TO VISUAL DISTURBANCES AND A PRESSURE SENSATION AND A NEW LIGHT ADJUSTABLE LENS (LAL, SN (B)(6), +19.0D) IMPLANTED AS A REPLACEMENT.
Patient problems: Eye Pain; Visual Disturbances
MAUDE report 24794812 ↗InjuryMar 31, 2026Optical Decentration
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS IMPLANTED (B)(6) 2024. AFTER THREE ADJUSTMENT LIGHT TREATMENTS, THE PATIENT COMPLAINED OF BLURRY VISION AND REPORTED DISSATISFACTION WITH THE CHOSEN REFRACTIVE TARGET. AT A FOLLOW UP VISIT, THE LENS WAS OBSERVED TO BE SLIGHTLY DISPLACED. THE PATIENT THEN UNDERWENT A PRK PROCEDURE AND A FOURTH ADJUSTMENT BEFORE THE LENS WAS EXPLANTED. A NEW LIGHT ADJUSTABLE LENS (LAL, SN (B)(6), +21.0D) IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision
MAUDE report 24747875 ↗InjuryDec 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED FROM THE RIGHT EYE DUE TO MONOCULAR DIPLOPIA AND VISUAL DISTURBANCES.
Patient problems: Visual Disturbances
MAUDE report 23829819 ↗InjuryOct 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED DUE TO PATIENT DISSATISFACTION. A NEW LAL (SN (B)(6), +19.5D) WAS IMPLANTED AS A REPLACEMENT.
Patient problems: Unspecified Eye / Vision Problem
MAUDE report 23406156 ↗InjuryOct 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.50D) WAS EXPLANTED FROM THE LEFT EYE DUE TO VISUAL DISTURBANCES AND BLURRY VISION.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 23386492 ↗InjuryOct 24, 2025Use of Device Problem; Use of Device Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.50D) WAS IMPLANTED IN THE RIGHT EYE IN 2024. POSTOPERATIVELY, THE PATIENT UNDERWENT 2 LIGHT ADJUSTMENTS AND NO LOCK-IN TREATMENTS. THE SITE REPORTED AN AREA ON THE LENS CONSISTENT WITH POLYMERIZATION AND THE LAL+ WAS EXPLANTED (B)(6) 2025.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 23386491 ↗InjuryOct 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED FROM THE RIGHT EYE DUE TO PATIENT DISSATISFACTION WITH THE CHOSEN REFRACTIVE TARGET.
Patient problems: Visual Impairment
MAUDE report 23375252 ↗InjuryOct 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED FROM THE LEFT EYE AFTER ONE ADJUSTMENT LIGHT TREATMENT DUE TO PATIENT DISSATISFACTION WITH VISION.
Patient problems: Blurred Vision
MAUDE report 23358881 ↗InjuryAug 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED FROM THE LEFT EYE DUE TO PATIENT DISSATISFACTION WITH CHOSEN REFRACTIVE TARGET AND VISUAL DISTURBANCES.
Patient problems: Visual Disturbances
MAUDE report 22903331 ↗InjuryAug 7, 2025Use of Device Problem; Use of Device Problem
A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN: (B)(6) WAS IMPLANTED IN THE PATIENT'S RIGHT EYE ON (B)(6) 2024. THE PATIENT DID NOT RETURN TO THE CLINIC FOR 9 MONTHS AFTER ON (B)(6) 2024. AT THE TIME, THE PATIENT HAD ONLY ONE LIGHT ADJUSTMENT AND NO LOCK-IN TREATMENTS. UPON RETURNING TO CLINIC ON (B)(6) 2025, THE PATIENT COMPLAINED OF AN ABRUPT CHANGE IN VISION AND HALOS. ON (B)(6) 2025, APPROXIMATELY 12 MONTHS AFTER THE IMPLANT SURGERY, THE LAL+ WAS EXPLANTED. A NEW LIGHT ADJUSTABLE LENS WAS IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 22741983 ↗InjuryJul 15, 2025Optical Decentration; Optical Decentration
A SITE REPORTED TO RXSIGHT THAT A LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED DUE TO BLURRY VISION AND LENS DISLOCATION.
Patient problems: Blurred Vision
MAUDE report 22509978 ↗InjuryJul 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED DUE TO BLURRY VISION AND VISUAL DISTURBANCE. A NEW LAL (SN (B)(6), +19.5D) WAS IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 22509977 ↗InjuryJun 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED DUE TO BLURRED VISION AND A NEW LIGHT ADJUSTABLE LENS (SN (B)(6), +19.5D) IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision
MAUDE report 22302973 ↗InjuryMay 30, 2025Optical Decentration; Optical Decentration
A SITE REPORTED TO RXSIGHT THAT A CAPSULE TEAR OCCURRED DURING IMPLANTATION OF THE LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) AND THE LAL+ WAS PLACED IN THE SULCUS.
Patient problems: Capsular Bag Tear
MAUDE report 22119338 ↗InjuryMay 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT, DUE TO A PATIENT'S DISSATISFACTION WITH VISION THE LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED FROM THE LEFT EYE AND A REPLACEMENT LIGHT ADJUSTABLE LENS (LAL, SN (B)(6), +18.5D) WAS IMPLANTED.
Patient problems: Visual Impairment
MAUDE report 22063422 ↗InjuryMay 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, (B)(6), +19.5D) WAS EXPLANTED FROM THE LEFT EYE DUE TO UNHAPPINESS WITH UNCORRECTED INTERMEDIATE AND NEAR VISION AND A NEW LAL (B)(6), +20.0D IMPLANTED AS A REPLACEMENT.
Patient problems: Visual Disturbances
MAUDE report 21998043 ↗InjuryApr 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A BILATERALLY IMPLANTED PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED FROM THE RIGHT EYE DUE TO BLURRY VISION AND VISUAL DISTURBANCES AND A NEW LAL (SN (B)(6), +19.5D) IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 21884623 ↗InjuryMar 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT WITH A HISTORY OF LASIK, LARGE PUPILS, AND DRY EYE HAD THEIR LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) EXPLANTED FROM THE LEFT EYE DUE TO VISUAL DISTURBANCES.
Patient problems: Visual Disturbances
MAUDE report 21688202 ↗InjuryMar 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED DUE TO DIPLOPIA AND VISUAL DISTURBANCES. A NEW LAL (SN (B)(6), +20.5D) WAS IMPLANTED AS A REPLACEMENT. THE PATIENT WAS REPORTED TO HAVE CONTINUED TO EXPERIENCE DIPLOPIA 1-DAY POST EXPLANT.
Patient problems: Diplopia/Double Vision; Visual Impairment
MAUDE report 21632884 ↗InjuryMar 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6) +19.5D) WAS EXPLANTED DUE TO BLURRY VISION AND VISUAL DISTURBANCE. A NEW LAL+ (B)(6), +20D) WAS IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 21600553 ↗InjuryMar 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED FROM THE PATIENT'S RIGHT EYE DUE TO PATIENT DISSATISFACTION WITH VISION.
Patient problems: Visual Impairment
MAUDE report 21575601 ↗InjuryFeb 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED FROM THE PATIENT'S LEFT EYE DUE TO VISUAL DISTURBANCES. A LIGHT ADJUSTABLE LENS (LAL, SN (B)(6), +20.0D) WAS IMPLANTED AS A REPLACEMENT.
Patient problems: Visual Disturbances
MAUDE report 21402811 ↗InjuryFeb 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +19.5D) WAS EXPLANTED DUE TO VISUAL DISTURBANCES AND A NEW LAL (SN (B)(6), +20.0D) IMPLANTED AS A REPLACEMENT.
Patient problems: Visual Disturbances
MAUDE report 21305197 ↗
Other Light Adjustable Lens (Lal) And Light Delivery Device (Ldd) devices
- RxSight Light Adjustable Lens (RxLAL) 60005-3205Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)50 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3200Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)47 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3210Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)39 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3190Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)32 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3195Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)31 reports
- Light Adjustable Lens (LAL) 60007-0200Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)31 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3215Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)30 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3185Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)29 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page