Skip to content

EdgeTaper SETF225Recalls & reported problems

US Endodontics, LLC · Sterile Endodontic Rotary Files · Class 1

All models of EdgeTaper (25)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00819514029892), received between Jun 30, 2025 and Jun 30, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 2, 2025Open, Classified

    Due to devices being irradiated above the established specification which my result in package integrity issues.

    Z-1114-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1

Patient problems

  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 30, 2025Break; Break

    THE FILE FRACTURED DURING AN ENDODONTIC TREATMENT (FRACTURE OCCURRED IN THE PATIENT'S MOUTH). THE AFFECTED PATIENT WAS NOT INJURED, AND NO MEDICAL CARE WAS REQUIRED. THE INCIDENT OCCURRED 20 MINUTES AFTER THE START OF THE TREATMENT. THE DENTIST STATED THAT THE INCIDENT HAPPENED DURING THE FIRST USE OF THE PRODUCT AND THAT THEY HAVE EXPERIENCE USING THIS PRODUCT, AS WELL AS A SIMILAR PRODUCT FROM RARE EUO. THE PRACTITIONER STATED THAT THEY DID NOT STERILIZE THE PRODUCT BEFORE USE. THE CLEANING PROTOCOL CONSISTED OF DISINFECTION AND LUBRICATION. REGARDING THE SEX OF THE AFFECTED PATIENT, THE DENTIST INDICATED BOTH MALE AND FEMALE. A REQUEST FOR CLARIFICATION HAS BEEN SENT TO DETERMINE IF MULTIPLE PATIENTS WERE INVOLVED. THE BROKEN FRAGMENTS COULD NOT BE REMOVED. THE SETTINGS USED WERE: POWER - 350 RPM, SPEED + 30 - 150, HANDPIECE USED FKG. THE CANNULA WAS BENT BEFORE INJECTION. THE FILES WERE DISCARDED (NO SAMPLE AVAILABLE FOR INVESTIGATION).QUANTITY OF PARTS IMPACTED BY THE DEFECT: 5.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 22359738 ↗

Other FILE, PULP CANAL, ENDODONTIC devices

All FILE, PULP CANAL, ENDODONTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page