Protege RX SECX-9-30-135Recalls & reported problems
Covidien LP · Carotid Stent System · Class 3 · Implant
All models of Protege RX (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00821684038574), received between Mar 29, 2023 and Jul 14, 2023 — classified as 2 injuries.
2 other recalls cover devices cleared under the same FDA submission (P060001); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- Bradycardia1
- Low Blood Pressure/ Hypotension1
- Obstruction/Occlusion1
- Stenosis1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING THE INDEX PROCEDURE ONE PROTÉGÉ RX CAROTID STENT WAS IMPLANTED IN THE LEFT COMMON ILIAC ARTERY (L1). ON THE SAME DAY, POST PROCEDURE, THE PATIENT SUFFERED FROM A POST PROCEDURE HYPOTENSION AND BRADYCARDIA RESPONSE. THE EVENT WAS TREATED WITH MEDICATION. THE MEDICATION WAS LATER DISCONTINUED AFTER VITAL SIGNS IMPROVED. THE PATIENT WAS DISCHARGED. THE PATIENT RECOVERED.
Patient problems: Bradycardia; Low Blood Pressure/ Hypotension
MAUDE report 17323595 ↗InjuryMar 29, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING THE INDEX PROCEDURE ONE PROTÉGÉ STENT WAS USED TO TREAT THE LEFT COMMON ILIAC (L1). APPROXIMATELY 6 MONTHS POST PROCEDURE THE PATIENT SUFFERED FROM IN STENT RESTENOSIS OF THE LEFT INTERNAL CAROTID ARTERY. THE PATIENT WAS ASYMPTOMATIC. 70% OCCLUSION WAS NOTED TO THE LEFT CAROTID ARTERY ON THE FOLLOW UP DUPLEX ULTRASOUND. AN ANGIOGRAM WAS PERFORMED WHICH REVEALED 70% OCCLUSION TO THE MID STENT FROM THE ORIGINAL PROCEDURE. THE EVENT WAS TREATED WITH RE-STENTING. THE PATIENTS MEDICATION WAS ALSO ADJUSTED. THE PATIENT IS RECOVERING.
Patient problems: Obstruction/Occlusion; Stenosis
MAUDE report 16634371 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P060001. One submission often covers many models, so check the product and lot details in each record.
- Z-0878-2015 · Dec 4, 2014 · Terminatedev3 Protege Rx Tapered Carotid Stent System, SECX-10-7-40-135, Sterile EO, Rx only. Indicated for the treatment of patients at high risk for adverse events from carotid endartere
- Z-0879-2015 · Dec 4, 2014 · Terminatedev3, Protg Rx Tapered Carotid Stent System, SEPX-8-6-40-135, Sterile EO, Rx only. Indicated for the treatment of patients at high risk for adverse events from carotid endarterect
Other STENT, CAROTID devices
- Neuroguard IEP® System NG-0740-140-2Contego Medical, Inc. · Stent, Carotid2 recalls32 reports
- Neuroguard IEP® System NG-NV-6-40Contego Medical, Inc. · Stent, Carotid2 recalls9 reports
- Neuroguard IEP® System NG-NV-7-40Contego Medical, Inc. · Stent, Carotid2 recalls8 reports
- ENROUTE Transcarotid Stent SR-0940-CSSilk Road Medical, Inc. · Stent, Carotid1 recall5 deaths58 reports
- Neuroguard IEP® System NG-0640-140-2Contego Medical, Inc. · Stent, Carotid2 recalls7 reports
- ENROUTE Transcarotid Stent SR-1040-CSSilk Road Medical, Inc. · Stent, Carotid1 recall3 deaths56 reports
- CGuard Prime Carotid Stent System CND0840Inspire M.D Ltd · Stent, Carotid1 recall1 death41 reports
- CGuard Prime Carotid Stent System CND0940Inspire M.D Ltd · Stent, Carotid1 recall31 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page