Skip to content

Protege RX Tapered SECX-10-7-30-135Recalls & reported problems

Covidien LP · Carotid Stent System · Class 3 · Implant

All models of Protege RX Tapered (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684038710), received between Mar 29, 2017 and Mar 29, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

2 other recalls cover devices cleared under the same FDA submission (P060001); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Deformation1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 29, 2017Material Deformation; Material Deformation

    PHYSICIAN ATTEMPTED TO USE A PROTEGE RX SELF-EXPANDING STENT TO TREAT PATIENT IN THE RIGHT, PROXIMAL COMMON CAROTID ARTERY. LESION TYPE WAS CALCIFIED WITH SEVERE TORTUOSITY, SEVERE CALCIFICATION AND 90% STENOSIS. ARTERY DIAMETER WAS 6 MM AND LESION LENGTH WAS 25 MM. A 4 MM BALLOON WAS USED TO PRE-DILATE. IT IS REPORTED THAT POST-STENT DEPLOYMENT, REMOVAL DIFFICULTIES WERE EXPERIENCED WITH THE DELIVERY SYSTEM DUE TO HEAVILY CALCIFIED INTERNAL CAROTID AND TORTUOSITY. THE DELIVERY CATHETER CAUGHT ON THE STENT CAUSING STENT DEFORMATION. A 0.035", 4 X 40 ADMIRAL BALLOON WAS USED TO DILATE THE STENT AND NEXT TO THE DELIVERY SYSTEM. BALLOON AND DEVICE WERE THEN REMOVED. NO PATIENT HISTORY REPORTED

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6441919 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P060001. One submission often covers many models, so check the product and lot details in each record.

  • Z-0878-2015 · Dec 4, 2014 · Terminated
    ev3 Protege Rx Tapered Carotid Stent System, SECX-10-7-40-135, Sterile EO, Rx only. Indicated for the treatment of patients at high risk for adverse events from carotid endartere
  • Z-0879-2015 · Dec 4, 2014 · Terminated
    ev3, Protg Rx Tapered Carotid Stent System, SEPX-8-6-40-135, Sterile EO, Rx only. Indicated for the treatment of patients at high risk for adverse events from carotid endarterect

Other STENT, CAROTID devices

All STENT, CAROTID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page