EverFlex™ EVD35-08-060-080Recalls & reported problems

Covidien LP · STENT EVD35-08-060-080 EF ENTRUST V01 · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00821684051283), received between Oct 10, 2018 and Feb 14, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Malposition of Device” (1 report).

3 other recalls cover devices cleared under the same FDA submission (P110023); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Malposition of Device1

Patient problems

  • Insufficient Information1
  • Reaction1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 14, 2024Malposition of Device; Malposition of Device

    A PHYSICIAN WAS IMPLANTING AND EVERFLEX ENTRUST FOR TREATMENT OF AN UNKNOWN LESION IN THE LEFT COMMON ILIAC ARTERY. THE STENT DEPLOYED IN AN UNINTENDED LESION SITE. IT WAS REPORTED THE STENT DEPLOYED BACKWARD AND JUMPED FORWARD. THE PHYSICIAN EXTENDED THE STENT DISTALLY WITH ANOTHER STENT TO COVER THE LESION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 18704296 ↗
  • InjuryOct 10, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A DISSECTED FEMORAL ARTERY DURING A ST-ELEVATION MYOCARDIAL INFARCTION (STEMI) WHICH REQUIRED THE IMPLANTATION OF A NON-MEDTRONIC CORONARY STENT AND 2 EVERFLEX ENTRUST STENTS IN THE ILIACS. IT WAS REPORTED THERE WAS NO ALLERGY OR SYMPTOMS PRESENT DURING THE PATIENT¿S STAY AT THE FACILITY WHERE THE STENTS WERE IMPLANTED. HOWEVER, MONTHS POST THE REPORTED PROCEDURE, THE PATIENT WAS EXPERIENCING SOME TYPE OF ALLERGIC REACTION. A SLIGHT RASH AND SKIN DISCOLORATION WERE REPORTED. THE PATIENT HAS BEEN PROVIDED WITH ALL OF THE STENT NAMES AND METALS ASSOCIATED WITH THE EVERFLEX STENTS. THE PATIENT HAS REPORTED ATTENDING A DERMATOLOGIST AND HAS REPORTED HAVING A BIOPSY CARRIED OUT ON THE RASH. THE RESULTS OF THE BIOPSY HAVE NOT YET BEEN REPORTED. IT IS NOT KNOWN IF THE PATIENT RECEIVED ANY TREATMENT FOR THE REPORTED RASH.

    Patient problems: Reaction

    MAUDE report 7951617 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P110023. One submission often covers many models, so check the product and lot details in each record.

  • Z-0623-2024 · Nov 20, 2023 · Completed
    Protege GPS Self-Expanding Peripheral Stent System, product number SERB65-09-60-80, Sterile.
  • Z-1644-2022 · Jul 14, 2022 · Open, Classified
    EverFlex Self-Expanding Peripheral Stent with Entrust Delivery System
  • Z-2048-2016 · May 23, 2016 · Terminated
    EV3 EverFiex" Self-expanding Peripheral Stent System, Sterilized using Ethylene Oxide, Rx Only, PRB35-08-040-080. The Protege EverFiex" Self-expanding Peripheral Stent System is a

Other STENT, SUPERFICIAL FEMORAL ARTERY devices

All STENT, SUPERFICIAL FEMORAL ARTERY devices →

Sources and updates