Skip to content

EverFlex™ EVD35-07-100-150Recalls & reported problems

Covidien LP · STENT EVD35-07-100-150 EF ENTRUST V01 · Class 3 · Implant

All models of EverFlex (41)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684051696), received between Apr 30, 2025 and Apr 30, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Premature Activation” (1 report).

3 other recalls cover devices cleared under the same FDA submission (P110023); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Activation1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 30, 2025Premature Activation; Premature Activation

    PHYSICIAN WAS ATTEMPTING TO USE A EVERFLEX ENTRUST 7X100X150 IN THE SUPERFICIAL FEMORAL ARTERY, PROXIMAL MID DISTAL PERIPHERAL ARTERY(S). CALCIFIED AND PLAQUE LESION, MINIMAL TORTUOSITY AND MODERATE CALCIFICATION. 70-90 % LESION STENOSIS, LESION LENGTH 200MM AND A RTERY/VEIN DIAMETER 5-6MM. A 5FR SHEATH WITH A 035 BENSON GW WAS USED. IFU WAS FOLLOWED. EMBOLIC PROTECTION WAS NOT USED, THE VESSEL WAS PRE-DILATED WITH BARD DILATION DEVICE. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT AND NO RESISTANCE WAS ENCOUN TERED WHEN ADVANCING THE DEVICE. THE THUMBSCREW/LOCK-PIN WAS CHECKED FOR SECUREMENT PRIOR TO PROCEDURE. IT WAS REPORTED THAT THERE WAS DEPLOYMENT ISSUES, THE STENT PRE-DEPLOYMENT (ON WAY TO TARGET SITE) IN THE PATIENT. PARTIAL DEPLOYMENT, 75% OF THE STENT DEPLOYED AND THE DISTAL PART OF THE STENT GOT STUCK IN THE INNER SHEATH AND WOULD NOT RELEASE THE STENT. PHYSICIAN ENDED UP USING FORCE AND PULLED THE STENT CATHETER OUT THROUGH THE 5FR SHEATH. THE TIP OF THE STENT SHEARED OFF. ATTEMPTS WERE MADE TO REMOVE THE STENT BUT REMAINS IN THE PATIENT. NO DEFORMATION NOTED TO THE DEPLOYED STENT AND THE DELIVERY SYSTEM WAS SAFELY REMOVED. NO OTHER ACTIONS TAKEN AS A RESU

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21934446 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P110023. One submission often covers many models, so check the product and lot details in each record.

  • Z-0623-2024 · Nov 20, 2023 · Completed
    Protege GPS Self-Expanding Peripheral Stent System, product number SERB65-09-60-80, Sterile.
  • Z-1644-2022 · Jul 14, 2022 · Open, Classified
    EverFlex Self-Expanding Peripheral Stent with Entrust Delivery System
  • Z-2048-2016 · May 23, 2016 · Terminated
    EV3 EverFiex" Self-expanding Peripheral Stent System, Sterilized using Ethylene Oxide, Rx Only, PRB35-08-040-080. The Protege EverFiex" Self-expanding Peripheral Stent System is a

Other STENT, SUPERFICIAL FEMORAL ARTERY devices