EverFlex™ Protégé™ EverFlex™ PRB35-06-040-080Recalls & reported problems
Ev3, Inc · STENT PRB35-06-040-080 PROTEGE EF V07 · Class 3 · Implant
All models of EverFlex Protégé EverFlex (27)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00821684059876), received between Mar 13, 2023 and Jun 6, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
3 other recalls cover devices cleared under the same FDA submission (P110023); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Difficult to Advance1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 6, 2023Use of Device Problem; Use of Device Problem
A PROTÉGÉ EF WAS IMPLANTED FOR THE TREATMENT OF A 30MM CALCIFIED LESION IN THE LEFT PROXIMAL REGION OF THE SUPERFICIAL FEMORAL ARTERY. THE ARTERY WAS 6MM IN DIAMETER AND MODERATELY CALCIFIED. THE LESION WAS PRE-DILATED WITH A 5MM DEVICE. A NON MEDTRONIC 6FR SHEATH AND GUIDEWIRE WERE USED. IT WAS REPORTED THE STENT WAS TAKEN FROM THE SHELF AND IMPLANTED IN THE PATIENT WITHOUT REALIZING THE DEVICE WAS EXPIRED. NO ISSUES WITH STENT, OTHER THAN IT WAS EXPIRED. THE PROCEDURE WAS COMPLETED IN A NORMAL FASHION WITH GOOD RESULTS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17069907 ↗MalfunctionMar 13, 2023Detachment of Device or Device Component; Difficult to Advance; Detachment of Device or Device Component; Difficult to Advance
A PROTEGE EVERFLEX SELF EXPANDING STENT WAS BEING USED TO TREAT A 20MM SOFT TISSUE LESION WITH 70% STENOSIS IN THE MID RIGHT TIBIAL/POPLITEAL TRUNK. THERE WAS SEVERE TORTUOSITY AND MODERATE CALCIFICATION. THE ARTERY DIAMETER WAS 6MM. THERE WERE ABNORMALITIES REPORTED IN RELATION TO ANATOMY, TRANS TIBIAL RETROGRADE FASHION ACCESSING THE ANTERIOR TIBIAL, THE ORIGIN OF THE AT WAS VERY TORTUROUS HAVING DIFFICULTIES TRACKING CATHETERS. A 6FR NON MEDTRONIC SHEATH AND A NON MEDTRONIC GUIDEWIRE WERE USED. NO EMBOLIC PROTECTION WAS USED. THE LESION WAS PRE DILATED WITH AN EVERCROSS. THE DEVICE WAS PREPPED AS PER THE IFU WITH NO ISSUES IDENTIFIED. RESISTANCE WAS ENCOUNTERED ADVANCING THE DEVICE, NO EXCESSIVE FORCE WAS USED. IT WAS REPORTED THAT THE DELIVERY SYSTEM WAS DIFFICULT TO REMOVE FOLLOWING STENT DEPLOYMENT. STENT CATHETER WAS REMOVED APPLIED SOME FORCE (BY PULLING IT OFF THE WIRE) IN A WAY THAT THE PIN PULL CAME APART REMAINING ON THE WIRE AND LOSING WIRE ACCESS BECAUSE OF THAT REASON. THE PHYSICIAN HAD TO REGAIN WIRE ACCESS IN ORDER TO POST DILATE THE TREATED AREA BUT WAS ABLE TO BE DEPLOYED BUT OFF THE TARGET LESION WITH LIKE 5-7MM DISTALLY TO THE TARGET LESION WHICH AFTER POSTDIL
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16529846 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P110023. One submission often covers many models, so check the product and lot details in each record.
- Z-0623-2024 · Nov 20, 2023 · CompletedProtege GPS Self-Expanding Peripheral Stent System, product number SERB65-09-60-80, Sterile.
- Z-1644-2022 · Jul 14, 2022 · Open, ClassifiedEverFlex Self-Expanding Peripheral Stent with Entrust Delivery System
- Z-2048-2016 · May 23, 2016 · TerminatedEV3 EverFiex" Self-expanding Peripheral Stent System, Sterilized using Ethylene Oxide, Rx Only, PRB35-08-040-080. The Protege EverFiex" Self-expanding Peripheral Stent System is a
Other STENT, SUPERFICIAL FEMORAL ARTERY devices
- EverFlex™ EVD35-06-150-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 death110 reports
- SUPERA S-60-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY108 reports
- SUPERA S-55-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY88 reports
- LifeStent® Solo Vascular Stent EX062003CLBard Peripheral Vascular, Inc. · STENT, SUPERFICIAL FEMORAL ARTERY80 reports
- BioMimics 3D Vascular Stent System 144700-12Veryan Medical Limited · Stent, Superficial Femoral Artery74 reports
- SUPERA S-50-120-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY55 reports
- EverFlex™ EVD35-07-060-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
- EverFlex™ EVD35-07-120-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page