EverFlex™ Protégé™ EverFlex™ PRB35-06-100-120Recalls & reported problems
Ev3, Inc · STENT PRB35-06-100-120 PROTEGE EF V07 · Class 3 · Implant
All models of EverFlex Protégé EverFlex (27)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00821684059944), received between Feb 16, 2021 and Oct 7, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (1 report).
3 other recalls cover devices cleared under the same FDA submission (P110023); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 7, 2024Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT AN ATTEMPT WAS MADE TO USE A PROTEGE EF TO TREAT A LESION. EMBOLIC PROTECTION WAS NOT USED. THE VESSEL WAS NOT PRE-DILATED. IFU WAS NOT FOLLOWED AS IT WAS REPORTED THAT AN EXPIRED PRODUCT WAS USED IN THE PATIENT. PHYSICIAN REALIZED THIS AFTER CASE. DEVICE REMAINS IMPLANTED. THERE WAS NO INJURY/INTERVENTION ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20391187 ↗MalfunctionFeb 16, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT A PHYSICIAN WAS ATTEMPTING TO USE A PROTEGE EF SELF-EXPANDING STENT DEVICE TO TREAT A RIGHT, MID PROXIMAL SUPERFICIAL FEMORAL ARTERY LESION IN A PATIENT. THE LESION WAS CALCIFIED, WITH LITTLE CALCIFICATION AND TORTUOSITY, WITH PLAQUE AND 70% STENOSIS. THE TARGET VESSEL WAS A 6MM VESSEL. THE LESION WAS PREDILATED, THE DEVICE PASSED THROUGH A PREVIOUSLY DEPLOYED STENT, NO RESISTANCE ENCOUNTERED WHEN ADVANCING THE DEVICE, EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT PHYSICIAN BELIEVES THE STENT IN THE PACKAGE WAS A 6X120 INSTEAD OF THE 6X100 INDICATED ON THE PACKAGING; THE STENT WAS DEPLOYED AND PLACED WITH NO ISSUES. A NON-MEDTRONIC 6X100 WAS USED BALLOON. THE PHYSICIAN NOTICED THE STENT WAS LONGER THAN ANTICIPATED AND COMPARED IT TO A 6X100 BALLOON. STENT WAS DEPLOYED WITH NO RESISTANCE. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11332061 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P110023. One submission often covers many models, so check the product and lot details in each record.
- Z-0623-2024 · Nov 20, 2023 · CompletedProtege GPS Self-Expanding Peripheral Stent System, product number SERB65-09-60-80, Sterile.
- Z-1644-2022 · Jul 14, 2022 · Open, ClassifiedEverFlex Self-Expanding Peripheral Stent with Entrust Delivery System
- Z-2048-2016 · May 23, 2016 · TerminatedEV3 EverFiex" Self-expanding Peripheral Stent System, Sterilized using Ethylene Oxide, Rx Only, PRB35-08-040-080. The Protege EverFiex" Self-expanding Peripheral Stent System is a
Other STENT, SUPERFICIAL FEMORAL ARTERY devices
- EverFlex™ EVD35-06-150-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 death110 reports
- SUPERA S-60-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY108 reports
- SUPERA S-55-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY88 reports
- LifeStent® Solo Vascular Stent EX062003CLBard Peripheral Vascular, Inc. · STENT, SUPERFICIAL FEMORAL ARTERY80 reports
- BioMimics 3D Vascular Stent System 144700-12Veryan Medical Limited · Stent, Superficial Femoral Artery74 reports
- SUPERA S-50-120-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY55 reports
- EverFlex™ EVD35-07-060-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
- EverFlex™ EVD35-07-120-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page