Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00821684060056), received between Apr 29, 2020 and Nov 10, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

3 other recalls cover devices cleared under the same FDA submission (P110023); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove1
  • Fracture1
  • Migration or Expulsion of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2023Fracture; Difficult to Remove; Fracture; Difficult to Remove

    IT WAS REPORTED THAT AN ATTEMPT WAS MADE TO USE A 7X40 PROTÉGÉ EF TO TREAT A CALCIFIED LESION IN THE LEFT PROXIMAL SUPERFICIAL FEMORAL ARTERY. DEGREE OF TORTUOSITY WAS MODERATE. DEGREE OF CALCIFICATION MODERATE. A 6FR 45 CM SHEATHE WAS USED. EMBOLIC PROTECTION WAS NOT USED. DEVICE WAS PREPPED PER IFU. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY-DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED. STENT DEFORMATION OCCURRED IN VIVO DURING POSITIONING/DEPLOYMENT. THERE WAS DIFFICULTY REMOVING THE STENT FOLLOWING STENT DEPLOYMENT. THE DELIVERY CATHETER DID NOT CATCH ON STENT UPON WITHDRAWAL. IT WAS REPORTED THAT A 7X60 PROTÉGÉ WAS PLACED WITH NO ISSUES. POST DILATED WITH NO ISSUES. THE 7X40 PROTÉGÉ STENT WAS PLACED, DELIVERY SYSTEM WOULD NOT COME OUT EASILY. END OF STENT NOT VISUALIZED. PORTION OF STENT FOUND IN ACCESS SHEATH AND REMOVED. THE ENTIRE STENT WAS FULLY REMOVED FROM THE PATIENT. PROCEDURE FINALIZED WITH BALLOON. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18110305 ↗
  • MalfunctionApr 29, 2020Migration or Expulsion of Device; Migration or Expulsion of Device

    PHYSICIAN INTENDED TO USE A PROTÉGÉ EVERFLEX SELF-EXPANDING STENT WITH A SPIDER FX FOR PATIENT TREATMENT. IFU WAS FOLLOWED. IT IS REPORTED THAT STENT JUMPING OCCURRED IN THE PATIENT'S VASCULATURE. THE DEVICE WAS SAFELY REMOVED. THE PROTÉGÉ STENT WAS REPLACED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10008416 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P110023. One submission often covers many models, so check the product and lot details in each record.

  • Z-0623-2024 · Nov 20, 2023 · Completed
    Protege GPS Self-Expanding Peripheral Stent System, product number SERB65-09-60-80, Sterile.
  • Z-1644-2022 · Jul 14, 2022 · Open, Classified
    EverFlex Self-Expanding Peripheral Stent with Entrust Delivery System
  • Z-2048-2016 · May 23, 2016 · Terminated
    EV3 EverFiex" Self-expanding Peripheral Stent System, Sterilized using Ethylene Oxide, Rx Only, PRB35-08-040-080. The Protege EverFiex" Self-expanding Peripheral Stent System is a

Other STENT, SUPERFICIAL FEMORAL ARTERY devices