Skip to content

SpiderFX™ SPD2-US-030-190Recalls & reported problems

Ev3, Inc · EP SPD2-US-030-190 SPIDER FX V04 · Class 2

All models of SpiderFX (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00821684068298), received between Jul 18, 2019 and Jun 26, 2020 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Mechanical Problem1

Patient problems

  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 26, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    THE PHYSICIAN ATTEMPTED TO USE A SPIDER FX TO TREAT THE LEFT DISTAL VEIN GRAFT TO DIAGONAL. TYPE OF TARGET LESION, DEGREE OF TORTUOSITY, DEGREE OF CALCIFICATION AND % LESION STENOSIS ARE UNKNOWN. IT IS UNKNOWN IF THE VESSEL WAS PRE-DILATED OR POST DILATED, OR IF THE IFU WAS FOLLOWED DURING PREPARATION, PROCEDURE, POST-PROCEDURE. IT WAS REPORTED THAT THE FILTER WIRE BASKET BROKE OFF AND REMAINED IN THE PATIENT. THE DEVICE WAS UNABLE TO BE SNARED AND/OR RETRIEVED TO BE REMOVED FROM BODY AND A STENT WAS PLACED TO PIN FILTER BASKET AGAINST VESSEL WALL AND MAINTAIN LUMINAL INTEGRITY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10199354 ↗
  • MalfunctionJul 18, 2019Mechanical Problem; Mechanical Problem

    THE PHYSICIAN WAS INTENDING TO USE A SPIDERFX FOR TREATMENT. IT WAS REPORTED THAT THE PHYSICIAN PROLAPSED THE DEVICE AND BENT THE WIRE TIP DISTAL TO THE SPIDER BASKET. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADVERSE EVENT REPORTED FOR THE PATIENT. WHEN THE DEVICE WAS RETURNED TO THE MANUFACTURING FACILITY, IT WAS NOTED THAT THE DEVICE WAS DAMAGED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8801635 ↗

Other Temporary Carotid Catheter for Embolic Capture devices

All Temporary Carotid Catheter for Embolic Capture devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page