Skip to content

URO-GUIDE 5498526Recalls & reported problems

Gyrus ACMI, LLC · 8.5FR (2.81mm) X 26 cm URO-GUIDE URETERAL STENT · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00821925002845), received between Sep 29, 2024 and Sep 29, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Difficult to Fold, Unfold or Collapse” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 4 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Fold, Unfold or Collapse4
  • Migration or Expulsion of Device4

Patient problems

  • Sepsis4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2024Difficult to Fold, Unfold or Collapse; Migration or Expulsion of Device; Difficult to Fold, Unfold or Collapse; Migration or Expulsion of Device

    IT WAS REPORTED THAT AFTER THE PLACEMENT OF A DOUBLE J URETERAL STENT, DISLOCATION OCCURRED IN THE URINARY TRACT CAUSING UROSEPSIS. THE DEVICE WAS UNABLE TO CREATE AN ADEQUATE SEAL CURL IN THE UPPER URINARY TRACT, THUS SELF-DISLOCATING FROM THE RENAL PELVIS INTO THE BLADDER. FOLLOW-UP INFORMATION RECEIVED THAT PATIENT VISITED THE EMERGENCY DEPARTMENT WITH UROSEPSIS. THE PROCEDURE WAS A URETERAL BRACE PLACEMENT OR URETERAL CATHETERIZATION PERFORMED. NO DELAY DURING THE PROCEDURE, BUT PATIENT HAVING TO REPEAT THE PROCEDURE 2 TIMES FOR DISLOCATION OF THE URETERAL STENT. CUSTOMER REPORTED THE PRODUCT HAS NO ABNORMALITY OR DEFECT, BUT DEVICE DESIGNED IN SUCH A WAY THAT THE PROXIMAL INTRARENAL CURL DOES NOT FULLY CURL AND THEREFORE DOES NOT PROVIDE SUPERIOR ANCHORAGE AND DISLOCATION WITH HIGH PROBABILITY. THE DEVICE DISLOCATION CAUSED PROLONGED HOSPITALIZATION AND TREATMENT WITH INTRAVENOUS ANTIBIOTICS TO RESOLVE THE PATIENT'S SEPSIS CUSTOMER REFERS THAT SIMILAR EVENT OCCURRED IN OTHER 3 CASES IN THE MONTH.

    Patient problems: Sepsis

    MAUDE report 20330503 ↗
  • InjurySep 29, 2024Difficult to Fold, Unfold or Collapse; Migration or Expulsion of Device; Difficult to Fold, Unfold or Collapse; Migration or Expulsion of Device

    IT WAS REPORTED THAT AFTER THE PLACEMENT OF A DOUBLE J URETERAL STENT, DISLOCATION OCCURRED IN THE URINARY TRACT CAUSING UROSEPSIS. THE DEVICE WAS UNABLE TO CREATE AN ADEQUATE SEAL CURL IN THE UPPER URINARY TRACT, THUS SELF-DISLOCATING FROM THE RENAL PELVIS INTO THE BLADDER. FOLLOW-UP INFORMATION RECEIVED THAT PATIENT VISITED THE EMERGENCY DEPARTMENT WITH UROSEPSIS. THE PROCEDURE WAS A URETERAL BRACE PLACEMENT OR URETERAL CATHETERIZATION PERFORMED. NO DELAY DURING THE PROCEDURE, BUT PATIENT HAVING TO REPEAT THE PROCEDURE 2 TIMES FOR DISLOCATION OF THE URETERAL STENT. CUSTOMER REPORTED THE PRODUCT HAS NO ABNORMALITY OR DEFECT, BUT DEVICE DESIGNED IN SUCH A WAY THAT THE PROXIMAL INTRARENAL CURL DOES NOT FULLY CURL AND THEREFORE DOES NOT PROVIDE SUPERIOR ANCHORAGE AND DISLOCATION WITH HIGH PROBABILITY. THE DEVICE DISLOCATION CAUSED PROLONGED HOSPITALIZATION AND TREATMENT WITH INTRAVENOUS ANTIBIOTICS TO RESOLVE THE PATIENT'S SEPSIS. CUSTOMER REPORTS 3 OTHER SUCH EVENTS OCCURRING IN THE MONTH.

    Patient problems: Sepsis

    MAUDE report 20330497 ↗
  • InjurySep 29, 2024Difficult to Fold, Unfold or Collapse; Migration or Expulsion of Device; Difficult to Fold, Unfold or Collapse; Migration or Expulsion of Device

    IT WAS REPORTED THAT AFTER THE PLACEMENT OF A DOUBLE J URETERAL STENT, A DISLOCATION OCCURRED IN THE URINARY TRACT CAUSING UROSEPSIS. THE DEVICE WAS UNABLE TO CREATE AN ADEQUATE SEAL CURL IN THE UPPER URINARY TRACT, THUS SELF-DISLOCATING FROM THE RENAL PELVIS INTO THE BLADDER. FOLLOW-UP INFORMATION RECEIVED THAT PATIENT VISITED THE EMERGENCY DEPARTMENT WITH UROSEPSIS. THE PROCEDURE WAS A URETERAL BRACE PLACEMENT OR URETERAL CATHETERIZATION PERFORMED. NO DELAY DURING THE PROCEDURE, BUT PATIENT HAVING TO REPEAT THE PROCEDURE 2 TIMES FOR DISLOCATION OF THE URETERAL STENT. CUSTOMER REPORTED THE PRODUCT HAS NO ABNORMALITY OR DEFECT, BUT DEVICE DESIGNED IN SUCH A WAY THAT THE PROXIMAL INTRARENAL CURL DOES NOT FULLY CURL AND THEREFORE DOES NOT PROVIDE SUPERIOR ANCHORAGE AND DISLOCATION WITH HIGH PROBABILITY. THE DEVICE DISLOCATION CAUSED PROLONGED HOSPITALIZATION AND TREATMENT WITH INTRAVENOUS ANTIBIOTICS TO RESOLVE THE PATIENT'S SEPSIS. CUSTOMER REFERS THAT SIMILAR EVENT OCCURRED IN OTHER 3 CASES IN THE MONTH.

    Patient problems: Sepsis

    MAUDE report 20330495 ↗
  • InjurySep 29, 2024Difficult to Fold, Unfold or Collapse; Migration or Expulsion of Device; Difficult to Fold, Unfold or Collapse; Migration or Expulsion of Device

    IT WAS REPORTED THAT AFTER A DIFFICULT URETHRAL STENT PLACEMENT ON (B)(6) 2024, THE PATIENT RETURNED TO THE HOSPITAL PRESENTING WITH UROSEPSIS ON (B)(6) 2024. AN ABDOMINAL COMPUTED TOMOGRAPHY (CT) SCAN WAS PERFORMED, AND IT WAS DISCOVERED THAT THE STENT HAD MIGRATED (FROM RENAL PELVIS INTO THE BLADDER) AND THEREFORE HAD TO BE REMOVED AND A SIMILAR DEVICE PLACED. THE DOCTOR REPORTS NO ABNORMALITIES WITH RESPECT TO THE STANDARD DESIGN OF THE DEVICE (PRODUCT APPARENTLY NOT COMPROMISED) HOWEVER, THE DOCTOR COMPLAINS OF DIFFICULTIES IN THE PLACEMENT OF THE STENT AND UNABLE TO CREATE AN ADEQUATE SEAL CURL IN THE UPPER URINARY TRACT, THUS LED TO MIGRATING OF THE DEVICE. THERE WERE NO DELAYS REPORTED DURING PROCEDURE.

    Patient problems: Sepsis

    MAUDE report 20330487 ↗

Other Stent, Ureteral devices

All Stent, Ureteral devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page