Drive 15201BVRecalls & reported problems
Drive Devilbiss Healthcare · BedsRails Product Description: Bed Rail,No-Gap,Half,BrownVein,1/cs · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 2
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00822383143026), received between Dec 7, 2020 and Oct 10, 2024 — classified as 2 deaths, 3 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Adverse Event Without Identified Device or Use Problem1
- Unintended Movement1
- Use of Device Problem1
Patient problems
- Physical Entrapment2
- Bone Fracture(s)1
- Fall1
- Laceration(s)1
- Limb Fracture1
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryOct 10, 2024Use of Device Problem; Use of Device Problem
DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A BED BY THE PROVIDER WHO REPORTED THE BEDRAIL "POPPED OFF" AND THE END USER FELL OUT OF THE BED DURING PERSONAL CARE. THE END USER WAS ADMITTED TO THE HOSPITAL WITH FRACTURES TO THEIR FEMUR AND TIBIA BONES AND LACERATIONS TO BOTH LEGS THAT REQUIRED SKIN GRAFTS. BASED ON OUR INVESTIGATION, IT APPEARS THAT THE CUSTOMER USED A BED RAIL THAT WAS INCOMPATIBLE WITH THE BED. AS PART OF ITS COMPLAINT REVIEW AND EVALUATION PROCESS, DRIVE DEVILBISS HEALTHCARE WILL ALSO NOTIFY THE MANUFACTURER OF THE PRODUCT ABOUT THIS COMPLAINT.
Patient problems: Laceration(s); Limb Fracture
MAUDE report 20421759 ↗InjuryNov 29, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A BED HALF RAIL BY A PROVIDER WHO REPORTED THAT THE HALF RAIL CAME OFF THE BED WHEN CAREGIVERS WERE TURNING THE PATIENT, CAUSING THEM TO FALL OFF THE BED AND BREAK THEIR COLLAR BONE, REQUIRING HOSPITAL TREATMENT. DRIVE DEVILBISS HEALTHCARE IS CURRENTLY INVESTIGATING THE INCIDENT, INCLUDING CONFIRMING THE SKU OF THE HALF RAIL INVOLVED IN THE INCIDENT, SINCE THE SKU ORIGINALLY REPORTED IS DIFFERENT THAN THE HALF RAIL THAT WAS RETURNED TO DRIVE FOR EVALUATION. IN ADDITION, SINCE THE RETURNED RAIL WAS OBSERVED TO BE INTACT AND UNBROKEN, DRIVE IS GATHERING FURTHER INFORMATION REGARDING HOW THE HALF RAIL WAS ATTACHED TO THE BED. AN UPDATE WILL BE FILED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
Patient problems: Bone Fracture(s)
MAUDE report 18230790 ↗DeathMar 2, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
FROM DRIVE DEVILBISS: DRIVE DEVILBISS HEALTHCARE IS THE INITIAL IMPORTER OF THE DEVICE WHICH IS A BED RAIL. THE INCIDENT WAS REPORTED TO CPSC. WE ARE FILING THIS REPORT PRIOR TO COMPLETION OF OUR INVESTIGATION.. THE REPORT WILL BE AMENDED WHEN ADDITIONAL INFORMATION IS ACQUIRED. THE END-USER WAS IN HOSPICE CARE. HIS EQUIPMENT WAS CHANGED WITHIN THE LAST TWO (2) WEEKS OF HIS LIFE INCLUDING HIS BED RAIL, LOW AIR LOSS MATTRESS AND HOSPITAL BED. THE FACILITY WAS INDEPENDENT LIVING, ADVANCED CARE HOME. HE WAS PLACED ON A TWO(2) HOUR WATCH SINCE HE HAD BEEN TRYING TO EXIT THE BED TO GET A DRINK. A CARE WORKER FOUND HIM TRAPPED UNDER THE BED RAIL WITH HIS FEET RESTING ON THE GROUND. HE DIED OF ACCIDENTAL, MECHANICAL ASPHYXIATION. THE NURSE ON STAFF THAT EVENING IS NO LONGER AN EMPLOYEE OF THE HOSPICE AND IS THEREFORE UNAVAILABLE FOR COMMENT. THE HOSPITAL SUPERVISOR REPORTED THAT THE DEVICES WERE RETURNED TO THE SERVICE PROVIDER AFTER THE INCIDENT. SHE ADVISED THAT THERE WERE NO ISSUES OR CONCERNS WITH THE RAILS, MATTRESSES, AND HOSPITAL BED. THERE WAS NO INVESTIGATION OF THE PRODUCTS AT THE HOSPICE LEVEL, DRIVE HAS UNABLE TO INDEPENDENTLY VERIFY THE CONFIGURATION OF THE PRODUCTS SINCE THE
Patient problems: Physical Entrapment
MAUDE report 13649139 ↗DeathJun 3, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
(B)(4) IS THE INITIAL IMPORTER OF THE DEVICE WHICH IS A BED RAIL. THE INCIDENT WAS REPORTED TO (B)(6). WE ARE FILING THIS REPORT PRIOR TO COMPLETION OF OUR INVESTIGATION.THE REPORT WILL BE AMENDED WHEN ADDITIONAL INFORMATION IS ACQUIRED. THE END-USER WAS IN HOSPICE CARE. HIS EQUIPMENT WAS CHANGED WITHIN THE LAST TWO (2) WEEKS OF HIS LIFE INCLUDING HIS BED RAIL, LOW AIR LOSS MATTRESS AND HOSPITAL BED. THE FACILITY WAS INDEPENDENT LIVING, ADVANCED CARE HOME. HE WAS PLACED ON A TWO(2) HOUR WATCH SINCE HE HAD BEEN TRYING TO EXIT THE BED TO GET A DRINK. A CARE WORKER FOUND HIM TRAPPED UNDER THE BED RAIL WITH HIS FEET RESTING ON THE GROUND. HE DIED OF ACCIDENTAL, MECHANICAL ASPHYXIATION. THE NURSE ON STAFF THAT EVENING IS NO LONGER AN EMPLOYEE OF THE HOSPICE AND IS THEREFORE UNAVAILABLE FOR COMMENT. THE HOSPITAL SUPERVISOR REPORTED THAT THE DEVICES WERE RETURNED TO THE SERVICE PROVIDER AFTER THE INCIDENT. SHE ADVISED THAT THERE WERE NO ISSUES OR CONCERNS WITH THE RAILS, MATTRESSES, AND HOSPITAL BED. THERE WAS NO INVESTIGATION OF THE PRODUCTS AT THE HOSPICE LEVEL, DRIVE HAS UNABLE TO INDEPENDENTLY VERIFY THE CONFIGURATION OF THE PRODUCTS SINCE THEY WERE DISASSEMBLED AND RETURNED TO THE PR
Patient problems: Physical Entrapment
MAUDE report 11935305 ↗InjuryDec 7, 2020Unintended Movement; Unintended Movement
(B)(6) IS THE INITIAL IMPORTER OF THE DEVICE WHICH IS A BED. WE HAVE NOT RECEIVED THE DEVICE FOR EVALUATION. WE WILL FILE AN UPDATED REPORT WHEN WE EVALUATE THE DEVICE. THIS REPORT IS IN RESPONSE TO MEDWATCH FILING 101545-2020-16. END-USER WAS TRYING TO GET INTO BED. SHE GRABBED THE RIGHT SIDE RAIL ON THE BED. SHE REPORTED THAT THE RAIL BECAME SHAKY. SHE LOST HER BALANCE, FELL AND HIT HER HEAD. SHE COMPLAINED OF PAIN IN SACRAL AND COCCYX AREA. THE RAIL MAY HAVE DETACHED FROM THE BED; HOWEVER, THIS HAS NOT BEEN CONFIRMED. IT IS UNKNOWN WHETHER A PRODUCT DEFECT OR USER ERROR/LACK OF MAINTENANCE LED TO THE INCIDENT. X-RAYS CONFIRMED A FRACTURED PELVIS.
Patient problems: Fall; Pain
MAUDE report 10958401 ↗
FDA clearance and classification
The FDA regulates product code FNJ (bed, manual) under 21 CFR 880.5120 as a Class I device. 21 CFR 880.5120 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Drive 15201BV
Has the Drive 15201BV been recalled?
No FDA recall record in our data names this exact device identifier (DI 00822383143026), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Drive 15201BV?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Dec 7, 2020 and Oct 10, 2024. In those reports, the most common specific device problems are “Unintended Movement” (1) and “Use of Device Problem” (1); the most common patient problems are “Physical Entrapment” (2), “Bone Fracture(s)” (1) and “Fall” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Drive 15201BV?
2 reports are classified as deaths, 3 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Drive 15201BV FDA approved?
Devices of this type (product code FNJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Drive Devilbiss Healthcare.
Who makes the Drive 15201BV, and is it still sold?
The FDA's device database lists Drive Devilbiss Healthcare as the company responsible for it (brand Drive, model 15201BV). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BED, MANUAL devices
- MedSurg Manual Bed FL23PUmano Medical Inc · Bed, Manual4 reports
- Drive 15021BVDrive Devilbiss Healthcare · BED, MANUAL2 reports
- SleepSafe Manual Hi-Lo ComboSleepsafe Beds, LLC · Bed, Manual1 report
- Drive 16500BVDrive Devilbiss Healthcare · BED, MANUAL1 report
- Drive 1222PDrive Devilbiss Healthcare · BED, MANUAL1 death1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page