Skip to content

Drive RTL10266Recalls & reported problems

Drive Devilbiss Healthcare · Nitro Rollator,Red Frame,1/ea · Class 1

Adverse event reports
8
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00822383288819), received between Jun 11, 2020 and Nov 13, 2023 — classified as 6 injuries, 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Device Unsafe to Use in Environment1
  • Insufficient Device Problem Information1
  • Material Fragmentation1
  • Material Twisted/Bent1
  • Unintended Movement1

Patient problems

  • Laceration(s)4
  • Bone Fracture(s)2
  • Concussion2
  • Fall2
  • Hemorrhage/Blood Loss/Bleeding2
  • Swelling/ Edema1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 13, 2023Mechanical Problem; Mechanical Problem

    FROM DISTRIBUTOR'S MDR: DRIVE DEVILLBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR THAT THE CURB ASSIST BROKE AND THE END USER FELL, CUT HER HEAD AND REQUIRED 12 STITCHES. THE END USER WAS ADMITTED TO THE HOSPITAL FOR 2 DAYS WITH A CONCUSSION AND 2 PINTS BLOOD LOSS. DEVILBISS IS CURRENTLY INVESTIGATING THE INCIDENT, INCLUDING ATTEMPTING TO RETRIEVE THE DEVICE FOR INVESTIGATION. AN UPDATE WILL BE FILED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

    Patient problems: Concussion; Hemorrhage/Blood Loss/Bleeding; Laceration(s)

    MAUDE report 18118845 ↗
  • InjuryNov 8, 2023Mechanical Problem; Mechanical Problem

    DRIVE DEVILLBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR THAT THE CURB ASSIST BROKE AND THE END USER FELL, CUT HER HEAD AND REQUIRED 12 STITCHES. THE END USER WAS ADMITTED TO THE HOSPITAL FOR 2 DAYS WITH A CONCUSSION AND 2 PINTS BLOOD LOSS. DEVILBISS IS CURRENTLY INVESTIGATING THE INCIDENT, INCLUDING ATTEMPTING TO RETRIEVE THE DEVICE FOR INVESTIGATION. AN UPDATE WILL BE FILED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

    Patient problems: Concussion; Hemorrhage/Blood Loss/Bleeding; Laceration(s)

    MAUDE report 18095652 ↗
  • InjuryAug 3, 2023Material Twisted/Bent; Material Twisted/Bent

    DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR BY AN END USER'S DAUGHTER, WHO STATED THAT THE "RIGHT HANDLE SUDDENLY BENT DOWN WHERE THE PLASTIC HANDLE MEETS THE METAL FRAME" AND REPORTED THAT HER MOTHER LOST HER BALANCE AND BROKE HER LEG. THE END USER WAS TAKEN TO THE HOSPITAL AND UNDERWENT A WEEK OF TREATMENT FOR HER INJURIES. DRIVE REQUESTED THE UNIT BE RETURNED FOR INVESTIGATION AND WILL FILE AN UPDATE IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

    Patient problems: Bone Fracture(s)

    MAUDE report 17452696 ↗
  • MalfunctionJun 16, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF A COMPLAINT REGARDING A ROLLATOR BY AN END USER, WHO STATED THAT "THERE IS A PART BY THE RIGHT WHEEL THAT CAUSED ME AN INJURY." THE END USER DID NOT SPECIFY WHETHER THERE WAS ANY MALFUNCTION OF THE ROLLATOR, WHAT PART WAS INVOLVED, OR WHETHER OR HOW THE PART WAS DEFECTIVE. THE END USER REPORTEDLY EXPERIENCED SWELLING AND WAS GIVEN ANTIBIOTICS AS A RESULT OF THE OTHERWISE UNSPECIFIED INJURY. DRIVE IS CURRENTLY INVESTIGATING THE INCIDENT, INCLUDING ATTEMPTING TO RETRIEVE THE PRODUCT AND INSPECT IT, AND WILL FILE AN UPDATE IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

    Patient problems: Swelling/ Edema

    MAUDE report 17147844 ↗
  • InjuryMar 2, 2022Device Unsafe to Use in Environment; Unintended Movement; Device Unsafe to Use in Environment; Unintended Movement

    DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR ON JANUARY 25TH. THE USER WAS USING THE ROLLATOR GOING DOWN THE STEPS. THE BRAKES WERE PRESSED BUT DID NOT FUNCTION CAUSING THE END-USER TO FALL. THE USER STATES THAT SHE WAS HOLDING ONTO THE ROLLATOR ON THE FIRST STEP WHILE GOING TO THE NEXT STEP WITH THE BRAKES ENGAGED BUT THE ROLLATOR MOVED, AND SHE FELL TO THE NEXT STEP. HER PHYSICAL THERAPIST TRIED TO BREAK HER FALL, BUT THE HANDLEBAR POKED HER IN THE SIDE, NEXT TO HER LEFT BREAST. HE THEN HELPED HER GET INTO THE HOUSE AND NO MEDICAL ATTENTION WAS REQUIRED. THE END USER SOUGHT MEDICAL ATTENTION ON JANUARY 28TH, THROUGH X-RAY IT WAS DETERMINED THAT SHE SUFFERED TWO BROKEN RIBS . NO ADDITIONAL INFORMATION IS AVAILABLE CURRENTLY. THE DEVICE IS INTENDED TO BE USED ON EVEN SURFACES NOT TRAVERSING STEPS. DRIVE IS CURRENTLY INVESTIGATING THE INCIDENT, INCLUDING ATTEMPTING TO RETRIEVE THE PRODUCT TO INSPECT IT.

    Patient problems: Bone Fracture(s); Fall

    MAUDE report 13656510 ↗
  • InjuryOct 6, 2021Break; Break

    (B)(6) IS THE INITIAL IMPORTER OF THE DEVICE WHICH IS A ROLLATOR. THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. WE WILL FILE A FOLLOW-UP SUBMISSION WHEN ADDITIONAL INFORMATION IS AVAILABLE. WHILE IN USE THE BOTTOM PART BAR THAT GOES ACROSS THE UNIT JUST CAME OFF CAUSING HER TO TOPPLE OVER AND ROLL DOWN THE HILL. SHE RECEIVED CUTS ON RIGHT ELBOWS AND LEFT LEG. SHE WENT TO HER FAMILY DOCTOR WHO PRESCRIBED ANTIBIOTICS WHEN HER ELBOW GOT INFECTED.

    Patient problems: Laceration(s)

    MAUDE report 12586111 ↗
  • Oct 13, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DRIVE DEVILBISS HEALTHCARE IS THE INITIAL REPORTER OF THE DEVICE WHICH IS A ROLLATOR. WE ARE FILING THIS REPORT TO CORRECT INACCURACIES IN HTE MEDWATCH FILING MW5096025. THERE WAS NO SERIOUS INJURY, INTERVENTION REQUIRED, OR A DEVICE MALFUNCTION. THE END-USER LIVES IN A SENIOR CITIZEN RESIDENCE. WHILE ADJUSTING THE BRAKE THE CONSUMER PINCHED HIS PINKIE. IT BLED AND HER REQUIRED THE NURSE TO ASSIST. TREATMENT WAS APPLICATION OF A BANDAID.

    Patient problems: Laceration(s)

    MAUDE report 10674958 ↗
  • InjuryJun 11, 2020Material Fragmentation; Material Fragmentation

    DRIVE DEVILBISS HEALTHCARE IS THE INITIAL IMPORTER OF THE DEVICE WHICH IS A ROLLATOR. WE ARE FILING THIS REPORT IN AN OVERABUNDANCE OF CAUTION SUE TO A MDR REGRESSION ANALYSIS. THE END-USER WAS ON THE STEPS OF A SWIMMING POOL WHILE SITTING ON THE DEVICE. THE RIGHT FRONT WHEEL REPORTEDLY FELL OFF CAUSING HIM TO HIT HIS HEAD. HE RECEIVED MINOR ABRASIONS AS WELL. THE DEVICE WAS RECEIVED FOR EVALUATION. IT SHOWED NORMAL USAGE. THE BRAKES ON EACH SIDE WERE CONFORMING. THE FRONT LEFT WHEEL WAS DETACHED FROM THE UNIT. THE FRONT AXLE BOLT WAS DELIVERED WITH NO DAMAGED AND SIGNS OF STRIPPED THREADS. THE THREADS INSIDE THE FORK, HOWEVER, SHOW MINIMAL DAMAGE TO THE THREADS . THE ROOT CAUSE MAY BE ATTRIBUTED TO THESE DAMAGE THREADS FROM ASSEMBLY, CAUSING THE AXLE BOLT TO DETACH FROM THE UNIT DURING VIBRATIONS OVER TIME. THE UNIT IS DEFECTIVE HISTORICAL DATA FOR COMPLAINTS OF THE SAME NATURE WAS REVIEWED AND FOUND VERY LIMITED REPORTING WHICH SHOWS NO TREND.

    Patient problems: Fall

    MAUDE report 10146400 ↗

Other WALKER, MECHANICAL devices

All WALKER, MECHANICAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.