Skip to content

DRIVEMEDICAL 15301CVRRecalls & reported problems

Drive Devilbiss Healthcare · Raw MaterialsTurning Product Description: Cover for 15301 · Class 1

All models of DRIVEMEDICAL (111)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00822383534411), received between Jan 14, 2026 and Jan 14, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Unintended Deflation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Deflation1

Patient problems

  • Pressure Sore/Ulcer1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 14, 2026Unintended Deflation

    DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A BARIATRIC MATTRESS. THE END USER REPORTED THAT THE MATTRESS HAS FAILED ON THREE OCCASIONS BY "BOTTOMING OUT," WHICH RESULTED IN THE DEVELOPMENT OF BED SORES REQUIRING HOME HEALTH WOUND CARE. ATTEMPTS TO GATHER ADDITIONAL INFORMATION HAVE BEEN UNSUCCESSFUL. AS PART OF ITS COMPLAINT REVIEW AND EVALUATION PROCESS, DRIVE DEVILBISS HEALTHCARE WILL ALSO NOTIFY THE MANUFACTURER OF THE PRODUCT ABOUT THIS COMPLAINT.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 24071244 ↗

Other COVER, MATTRESS (MEDICAL PURPOSES) devices

All COVER, MATTRESS (MEDICAL PURPOSES) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page