Skip to content

Cook G01377Recalls & reported problems

Cook Incorporated · Trocar Needle · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002013773), received between Nov 1, 2017 and Nov 5, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 1, 2024Open, Classified

    Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility

    Z-1526-2024 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Device Operates Differently Than Expected1
  • Difficult to Advance1
  • Difficult to Remove1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 5, 2025Difficult to Remove; Difficult to Advance; Difficult to Remove; Difficult to Advance

    IN ADDITIONAL INFORMATION RECEIVED ON 13NOV2025, IT WAS CONFIRMED THAT THE DEVICE DID NOT MAKE PATIENT CONTACT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23478076 ↗
  • InjuryNov 1, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected

    THE INTERNATIONAL CUSTOMER REPORTED THAT, DURING A LITHOTRIPSY PROCEDURE, A TROCAR NEEDLE WAS EMPLOYED TO PROVIDE ROUTINE ACCESS TO THE PATIENT'S KIDNEY. A ZIPWIRE BRAND GUIDEWIRE (WHICH WAS NOT OF COOK MANUFACTURE) WAS THEN THREADED THROUGH THE TROCAR NEEDLE, WHICH WAS ALSO REPORTEDLY PER PROCEDURE. WHEN THE PHYSICIAN BEGAN TO PULL THE GUIDE WIRE OUT OF THE ASSEMBLY, THE TROCAR PEELED THE ZIP WIRE AND CUT SOME OF THE EDGES OF THE WIRE OFF. THESE PIECES DISENGAGED INTO THE PATIENT'S CAVITY, BUT WERE SUCCESSFULLY REMOVED IN THEIR ENTIRETY BY THE OPERATOR. THE ONLY UNEXPECTED OUTCOME FOR THE PATIENT WAS AN EXTENSION TO THE SURGERY TIME OF ABOUT 20 MINUTES. THE PRODUCT IS REPORTEDLY AVAILABLE FOR RETURN; HOWEVER, AS OF THE DATE OF THIS REPORT, NO DEVICE HAS YET BEEN RECEIVED FOR EVALUATION.

    Patient problems: No Code Available

    MAUDE report 6993079 ↗

Other Trocar devices

All Trocar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page