Coda G03832Recalls & reported problems

Cook Incorporated · Coda, LP Balloon Catheter · Class 2

All models of Coda (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002038325), received between Jul 16, 2018 and Aug 12, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture5
  • Difficult to Advance3
  • Inflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2025Material Rupture; Material Rupture

    NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22786255 ↗
  • MalfunctionMar 12, 2025Difficult to Advance; Difficult to Advance

    ADDITIONAL INFORMATION WAS PROVIDED ON 10MAR2025. IMAGING AND PROCEDURAL NOTES FOR THE EVENT ARE NOT AVAILABLE. THE COOK CODA BALLOON WAS THE ONLY DEVICE THAT EXPERIENCED DIFFICULT ADVANCEMENT OR DIFFICULT REMOVAL IN THE PROCEDURE. ALTHOUGH THERE WAS RESISTANCE, THE DELIVERY WAS ACHIEVED, AND THE PROCEDURE WAS COMPLETED WITH A TOUCH-UP USING A CODA BALLOON. WHEN ASKED "WAS THE DEVICE DIFFICULT TO ADVANCE THROUGH THE SHEATH OR THROUGH THE PATIENT'S VESSELS, THE RESPONSE WAS "YES."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21587353 ↗
  • MalfunctionMar 12, 2025Difficult to Advance; Difficult to Advance

    ADDITIONAL INFORMATION WAS PROVIDED ON 10MAR2025. IMAGING AND PROCEDURAL NOTES FOR THE EVENT ARE NOT AVAILABLE. THE COOK CODA BALLOON WAS THE ONLY DEVICE THAT EXPERIENCED DIFFICULT ADVANCEMENT OR DIFFICULT REMOVAL IN THE PROCEDURE. ALTHOUGH THERE WAS RESISTANCE, THE DELIVERY WAS ACHIEVED, AND THE PROCEDURE WAS COMPLETED WITH A TOUCH-UP USING A CODA BALLOON. WHEN ASKED "WAS THE DEVICE DIFFICULT TO ADVANCE THROUGH THE SHEATH OR THROUGH THE PATIENT'S VESSELS, THE RESPONSE WAS "YES."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21587290 ↗
  • MalfunctionMar 12, 2025Difficult to Advance; Difficult to Advance

    ADDITIONAL INFORMATION WAS PROVIDED ON 10MAR2025. IMAGING FOR THE EVENT IS NOT AVAILABLE. PROCEDURAL NOTES FOR THE EVENT ARE NOT AVAILABLE. NONE OF THE DEVICES EXPERIENCED DIFFICULT ADVANCEMENT OR REMOVAL DURING THE PROCEDURE. WHEN ASKED "WAS THE DEVICE DIFFICULT TO ADVANCE THROUGH THE SHEATH OR THROUGH THE PATIENT'S VESSELS, THE RESPONSE WAS "YES." THE DEVICE WAS NOT DIFFICULT TO REMOVE FROM THE PATIENT. THE PHYSICIAN ALSO ATTEMPTED TO USE THE SAME DEVICE ON THE OTHER SIDE BUT CONFIRMED THAT DELIVERY WAS DIFFICULT. A BALLOON FROM ANOTHER MANUFACTURER WAS USED AND DELIVERY WAS COMPLETED ON BOTH SIDES WITHOUT RESISTANCE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21587255 ↗
  • MalfunctionJul 25, 2023Inflation Problem; Inflation Problem

    ADDITIONAL INFORMATION WAS PROVIDED ON (B)(6) 2023. PRIOR TO INSERTION OF THE DEVICE INTO THE PATIENT, THE DEVICE WAS PREPPED BY FLUSHING THE CATHETER AND ATTEMPTING TO INFLATE THE BALLOON. THE BALLOON "DID NOT INFLATE ENOUGH." NO OBVIOUS HOLES IN THE BALLOON WERE DETECTED. THE DEVICE WAS NOT USED. THE PROCEDURE WAS COMPLETED BY USING A COMPETITOR'S BALLOON CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17388723 ↗
  • MalfunctionJul 21, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT TWO CODA LP BALLOONS RUPTURED DURING AN ENDOVASCULAR ANEURYSM REPAIR PROCEDURE. THE PATIENT WAS CONSIDERED SUITABLE FOR THIS PROCEDURE, ALTHOUGH HAD VESSEL CALCIFICATION AND SLIGHT ANGULATION AT OR NEAR THE INFLATION SITE. AFTER A STENT GRAFT WAS PLACED BY RIGHT APPROACH, BALLOONING WAS PERFORMED IN THE PROXIMAL SITE OF THE STENT GRAFT. THE FIRST BALLOON RUPTURED "ON THE DISTAL SIDE OF THE BALLOON" AFTER THE SECOND OR THIRD BALLOONING. THE SECOND BALLOON ALSO RUPTURED AT THE PROXIMAL SITE OF THE STENT GRAFT, THIS TIME ON FIRST BALLOONING. THE CONTRAST TO SALINE RATIO FOR INFLATION OF THE BALLOONS WAS 10:20 AND THE VOLUME USED FOR BOTH DEVICES WAS APPROXIMATELY 20/25 ML. NO INFLATION DEVICE WAS UTILIZED FOR INFLATION OF EITHER BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER MANUFACTURER'S BALLOON DEVICE. THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED. THE FIRST BALLOON THAT RUPTURED IS CAPTURED IN THE REPORT WITH PATIENT IDENTIFIER (B)(6) (THIS REPORT). THE SECOND BALLOON THAT RUPTURED IS CAPTURED IN THE REPORT WITH PATIENT IDENTIFIER (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15075124 ↗
  • MalfunctionJul 21, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT TWO CODA LP BALLOONS RUPTURED DURING AN ENDOVASCULAR ANEURYSM REPAIR PROCEDURE. THE PATIENT WAS CONSIDERED SUITABLE FOR THIS PROCEDURE, ALTHOUGH HAD VESSEL CALCIFICATION AND SLIGHT ANGULATION AT OR NEAR THE INFLATION SITE. AFTER A STENT GRAFT WAS PLACED BY RIGHT APPROACH, BALLOONING WAS PERFORMED IN THE PROXIMAL SITE OF THE STENT GRAFT. THE FIRST BALLOON RUPTURED "ON THE DISTAL SIDE OF THE BALLOON" AFTER THE SECOND OR THIRD BALLOONING. THE SECOND BALLOON ALSO RUPTURED AT THE PROXIMAL SITE OF THE STENT GRAFT, THIS TIME ON FIRST BALLOONING. THE CONTRAST TO SALINE RATIO FOR INFLATION OF THE BALLOONS WAS 10:20 AND THE VOLUME USED FOR BOTH DEVICES WAS APPROXIMATELY 20/25 ML. NO INFLATION DEVICE WAS UTILIZED FOR INFLATION OF EITHER BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER MANUFACTURER'S BALLOON DEVICE. THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED. THE FIRST BALLOON THAT RUPTURED IS CAPTURED IN THE REPORT WITH PATIENT IDENTIFIER 368686 THE SECOND BALLOON THAT RUPTURED IS CAPTURED IN THE REPORT WITH PATIENT IDENTIFIER 368678 (THIS REPORT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15074812 ↗
  • MalfunctionAug 24, 2018Material Rupture; Material Rupture

    THERE HAS BEEN NO NEW INFORMATION RECEIVED SINCE THE LAST REPORT WAS SUBMITTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7814413 ↗
  • MalfunctionJul 16, 2018Material Rupture; Material Rupture

    IT WAS REPORTED AN ENDOVASCULAR ANEURYSM REPAIR (EVAR) AGAINST AN ABDOMINAL AORTIC ANEURYSM (AAA) WAS PERFORMED. THERE WERE NO SPECIFIC ANATOMICAL PROBLEMS IN THE PATIENT. AFTER ANOTHER MANUFACTURER'S GRAFT WAS PLACED IN THE TARGET POSITION, TOUCH-UP BALLOONING WAS CONDUCTED WITH THE CODA LP BALLOON CATHETER. THE BALLOON RUPTURED WHEN THE PHYSICIAN ATTEMPTED TO DILATE THE JUNCTION OF THE MAIN BODY AND THE LEG WITH THE CODA. ANOTHER CODA LP BALLOON CATHETER OF THE SAME LOT NUMBER WAS USED AND THE BALLOONING WAS PERFORMED WITHOUT PROBLEMS AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PHYSICIAN COMMENTED THAT THE EXACT AMOUNT OF INFLATION FLUID WAS NOT KNOWN. IT IS POSSIBLE THAT THE BALLOON GOT STUCK ON THE PROXIMAL EDGE OF THE OTHER MANUFACTURER'S GRAFT CAUSING IT TO RUPTURE. AS REPORTED, THERE WERE NO ADVERSE EFFECTS TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7691264 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →