Cook G04035Recalls & reported problems

Cook Incorporated · Central Venous Catheter Set Double Lumen Polyurethane · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002040359), received between Mar 9, 2018 and Jul 11, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Crack” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack2
  • Unraveled Material2
  • Difficult to Remove1
  • Fracture1
  • Leak/Splash1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • Foreign Body In Patient1
  • Perforation of Vessels1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 11, 2024Fracture; Unraveled Material; Fracture; Unraveled Material

    IT WAS REPORTED THAT THE WIRE GUIDE INCLUDED IN THE DOUBLE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER SET WAS DIFFICULT TO REMOVE DURING A CENTRAL VENOUS CATHETER INSTALLATION PROCEDURE ON A 3-MONTH-OLD MALE PATIENT. CENTRAL VENOUS CATHETER INSTALLATION INTO THE RIGHT FEMORAL VEIN WAS SCHEDULED ON (B)(6) 2024 DUE TO POOR PERIPHERAL ACCESS, THE NEED FOR SEDATION AND ANTIBIOTIC TREATMENT. UNDER ASEPTIC AND ULTRASOUND-GUIDED TECHNIQUE, A DOUBLE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER WAS INSERTED INTO THE TARGETED VESSEL. THE VEIN WAS CANALIZED, AND THE WIRE GUIDE WAS ADVANCED UNTIL A STOP WAS FELT. A DECISION WAS THEN MADE TO WITHDRAW THE WIRE. DIFFICULTY WAS EXPERIENCED DURING ATTEMPTED REMOVAL AND AN ENTRAPMENT BECAME EVIDENT. ONCE AGAIN, AN ATTEMPT WAS MADE TO REMOVE THE WIRE GUIDE. IT WAS THEN DISCOVERED THAT THE WIRE GUIDE HAD FRAYED, WITH ONLY A PORTION BEING REMOVED FROM THE PATIENT. UPON ULTRASOUND EVALUATION OF FEMORAL VESSELS AND EXTRALUMINAL SPACE BY A VASCULAR SURGEON, THE METAL GUIDE WAS FOUND TO BE UNRAVELED DISTALLY, WITH A KNOT IN THE TIP. THE GUIDE WAS THEN PULLED UNDER ULTRASOUND, BUT THE GUIDE WAS UNABLE TO BE MOVED. FINALLY, TOGETHER WITH A PEDIATRIC SURGEON, T

    Patient problems: Foreign Body In Patient

    MAUDE report 19724135 ↗
  • MalfunctionMar 13, 2019Crack; Leak/Splash

    NOTED CVC CRACKED AT HUB, MEDICATIONS LEAKING. LINE WAS IN PLACE FOR 6 DAYS. REQUIRED REPLACEMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8416033 ↗
  • MalfunctionFeb 1, 2019Crack; Crack

    NO NEW EVENT DESCRIPTION INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8296946 ↗
  • InjuryMar 9, 2018Difficult to Remove; Unraveled Material; Difficult to Remove; Unraveled Material

    THE CUSTOMER REPORTED THAT, DURING THE USE OF THE LEFT INTERNAL JUGULAR VEIN DOUBLE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER SET, THE WIRE GUIDE BECAME TRAPPED IN THE INTRODUCER NEEDLE, MAKING IT IMPOSSIBLE TO PASS OR WITHDRAW THE WIRE GUIDE. UPON ATTEMPTED REMOVAL, THE WIRE GUIDE FRAYS. THE CUSTOMER INDICATED THAT, AS A RESULT OF THE PRODUCT PROBLEM, THE PATIENT'S BLOOD VESSEL RUPTURED. THE CUSTOMER DID NOT REPORT WHAT WAS DONE TO CORRECT THE CONDITION, AND IT IS UNKNOWN IF ADDITIONAL PROCEDURES WERE REQUIRED. THE COMPLAINT PRODUCT IS AVAILABLE FOR RETURN; HOWEVER, AS OF THE DATE OF THIS REPORT, NO DEVICE HAS YET BEEN RECEIVED FOR EVALUATION. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Perforation of Vessels

    MAUDE report 7327288 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →