Skip to content

Jeffery G12016Recalls & reported problems

Cook Incorporated · Jeffery Wire Guide Exchange Set With Nitinol Wire Guide · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002120167), received between Jun 28, 2021 and Dec 5, 2022 — classified as 2 injuries.

The most frequently reported specific device problem is “Material Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Separation2

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 5, 2022Material Separation; Material Separation

    IT WAS REPORTED THE WIRE GUIDE IN A JEFFREY WIRE GUIDE EXCHANGE SET SEPARATED DURING A PERCUTANEOUS TRANSHEPATIC BILIARY DRAINAGE PROCEDURE. THE WIRE WAS ADVANCED INTO THE PATIENT AND WAS NAVIGATED AROUND A BEND. A COMPETITOR'S DILATOR WAS ADVANCED OVER THE WIRE AND THE WIRE WAS REMOVED. AT THIS TIME, A PORTION OF THE WIRE BROKE AND WAS RETAINED IN THE BILE DUCT. A SAFETY WIRE WAS THEN ADVANCED TO MAINTAIN ACCESS TO THE BILE DUCT AND THE WIRE FRAGMENT WAS RETRIEVED USING A SNARE. THE PROCEDURE WAS COMPLETED WITH A NEW WIRE GUIDE. IT WAS NOTED THAT THE PROCEDURE WAS PROLONGED DUE TO THIS INCIDENT. NO OTHER ADVERSE EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15920900 ↗
  • InjuryJun 28, 2021Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 12076031 ↗

Other ACCESSORIES, CATHETER devices

All ACCESSORIES, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page