Skip to content

Cook G14090Recalls & reported problems

Cook Incorporated · Brush Biopsy Set · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002140905), received between Nov 28, 2018 and Nov 15, 2023 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent2
  • Break1
  • Contamination /Decontamination Problem1
  • Device Contaminated During Manufacture or Shipping1
  • Unraveled Material1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 15, 2023Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    AS REPORTED, CONTAMINATION WAS FOUND INSIDE THE PACKAGE OF THE URETERAL BRUSH BIOPSY SET PRIOR TO A URETERAL BIOPSY PROCEDURE. THE PACKAGE WAS NOT OPENED, AND THE DEVICE DID NOT MAKE PATIENT CONTACT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER DEVICE OF THE SAME TYPE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18141233 ↗
  • MalfunctionJan 3, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED, A RED, RUST LIKE MATERIAL WAS DISCOVERED INSIDE THE UNOPENED PACKAGE OF A URETERAL BRUSH BIOPSY SET. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE ALLEGED MALFUNCTION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9544565 ↗
  • MalfunctionNov 29, 2018Break; Material Twisted/Bent

    IT WAS REPORTED THAT WHEN THE USER OPENED THE PACKAGE, IT WAS NOTED A KINK IN THE URETERAL BRUSH BIOPSY SET. HOWEVER, IT WAS FURTHER NOTED THAT THE COIL WAS BROKEN NEAR THE BRUSH AREA. THERE WERE NO ADVERSE EFFECTS REPORTED DUE TO THE EVENT AND NO REPORT OF AN ADDITIONAL PROCEDURE. A NEW PRODUCT WAS USED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8115308 ↗
  • MalfunctionNov 28, 2018Unraveled Material; Material Twisted/Bent

    IT WAS REPORTED THAT WHEN THE USER OPENED THE PACKAGE, IT WAS NOTED THAT THE TIP OF URETERAL BRUSH WAS BENT. IT WAS FURTHER NOTED THAT THE COIL HAD BEEN UNWOVEN AND WAS IN A DAMAGED STATE. ANOTHER DEVICE WAS USED INSTEAD, AND THE PROCEDURE WAS COMPLETED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8111133 ↗

Other CATHETER, UROLOGICAL devices

All CATHETER, UROLOGICAL devices →

Sources and updates