Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002148406), received between Apr 10, 2018 and Oct 7, 2024 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn2
  • Material Separation1
  • Tear, Rip or Hole in Device Packaging1
  • Unraveled Material1

Patient problems

  • No Consequences Or Impact To Patient3
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 7, 2024Material Separation; Material Separation

    AS REPORTED, A 'SOF-FLEX PEDIATRIC DOUBLE PIGTAIL URETERAL STENT SET' WAS PLACED ON (B)(6) 2024 FOR A STENT RE-IMPLANT PROCEDURE AND FRACTURED AT THE PATIENTS HOME ON (B)(6) 2024. THE PATIENT REPORTED TO HAVE PAIN AND PASSED PART OF THE STENT AT HOME; THE PATIENT THEN RETURNED TO THE USER FACILITY FOR REMOVAL OF THE REMAINING STENT FRAGMENTS REMOVED FROM THE URETHRA AND BLADDER. ANOTHER STENT WAS NOT PLACED. THE STENT WAS NOT SET TO BE REMOVED FOR ANOTHER 2 WEE KS. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY OTHER ADDITIONAL INTERVENTION OR EXPERIENCE ANY OTHER ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: Foreign Body In Patient; Pain

    MAUDE report 20385678 ↗
  • MalfunctionOct 13, 2021Unraveled Material; Unraveled Material

    NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12625388 ↗
  • MalfunctionMar 4, 2020Material Split, Cut or Torn; Material Split, Cut or Torn

    NO NEW PATIENT OR EVENT INFORMATION HAS BEEN RECEIVED SINCE THE LAST REPORT WAS SUBMITTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9787731 ↗
  • MalfunctionAug 23, 2019Material Split, Cut or Torn; Material Split, Cut or Torn

    PRIOR TO A STENT PLACEMENT, THE DISTAL TIP OF A SOF-FLEX PEDIATRIC DOUBLE PIGTAIL URETERAL STENT SET WAS FOUND TO BE SPLIT OPEN. ANOTHER SOF-FLEX PEDIATRIC DOUBLE PIGTAIL URETERAL STENT SET WAS OPENED TO CARRY OUT THE PROCEDURE. NO ADVERSE EFFECTS TO THE PATIENT WERE REPORTED DUE TO THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8925202 ↗
  • MalfunctionApr 10, 2018Tear, Rip or Hole in Device Packaging

    THE COMPLAINANT REPORTED THAT THERE WAS A TEAR IN ONE FENESTRATED PIGTAIL END OF THE SOF-FLEX PEDIATRIC DOUBLE PIGTAIL URETERAL STENT SET. THERE WAS NO REPORTED HARM TO THE PATIENT AND NO DELAY IN THE CASE. IT IS NOT KNOWN IF THE DEVICE MADE PATIENT CONTACT, BUT NO SERIOUS INJURIES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7419669 ↗