Skip to content

Sof-Flex G15696Recalls & reported problems

Cook Incorporated · Sof-Flex Multi-Length Ureteral Stent Set · Class 2 · Implant

All models of Sof-Flex (14)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002156968), received between Apr 12, 2018 and Oct 21, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Device Damaged Prior to Use1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 21, 2024Break; Break

    IT WAS REPORTED THAT THE 'STRINGS' DETACHED FROM THE 'SOF-FLEX MULTI-LENGTH URETERAL STENT' DURING ATTEMPTED PATIENT SELF-REMOVAL. THE STENT WAS IMPLANTED ON (B)(6) 2024, DURING A URETEROSCOPY WITH LASER LITHOTRIPSY PROCEDURE, AND THE PATIENT ATTEMPTED TO REMOVE THE STENT, AS INSTRUCTED, ON (B)(6) 2024. THE PATIENT RETURNED TO THE CLINICAL SETTING ON (B)(6) 2024 AND RETAINED STENT WAS CONFIRMED VIA IMAGING. THE STENT WAS THEN REMOVED WITH A FLEXIBLE CYSTOSCOPE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL INTERVENTION DUE TO THIS OCCURRENCE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20494800 ↗
  • InjuryOct 21, 2024Break; Break

    IT WAS REPORTED THAT THE 'STRINGS' DETACHED FROM THE 'SOF-FLEX MULTI-LENGTH URETERAL STENT' DURING ATTEMPTED PATIENT SELF-REMOVAL. THE STENT WAS IMPLANTED ON (B)(6) 2024, DURING A RIGHT FLEXIBLE URETEROSCOPY WITH LASER LITHOTRIPSY PROCEDURE, AND THE PATIENT ATTEMPTED TO REMOVE THE STENT, AS INSTRUCTED, ON 02JUL2024 . DURING ATTEMPTED REMOVAL, THE 'STRINGS' SEPARATED FROM THE DEVICE AND THE STENT REMAINED IN THE PATIENT. THE PATIENT RETURNED TO THE CLINICAL SETTING ON (B)(6) 2024 AND RETAINED STENT WAS CONFIRMED VIA IMAGING. THE USER ATTEMPTED TO REMOVE THE STENT WITH A FLEXIBLE CYSTOSCOPE BUT WAS UNSUCCESSFUL. THE STENT IS SCHEDULED TO BE RETRIEVED UNDER A GENERAL ANESTHESIA (GA) ON (B)(6) 2024; NO INFORMATION HAS SINCE BEEN MADE AVAILABLE REGARDING THE SCHEDULED STENT REMOVAL. ADDITIONAL INFORMATION REGARDING THE EVENT HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: Foreign Body In Patient; Pain

    MAUDE report 20494664 ↗
  • MalfunctionApr 12, 2018Break; Device Damaged Prior to Use

    IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, WHEN AN USER REMOVED THE URETERAL STENT FROM THE STERILE PACKAGING AND PULLED OFF THE SUTURE, THE TOP OF THE STENT PIGTAIL PULLED OFF. IT WAS FURTHER NOTED THAT PROXIMAL STENT COIL WAS RIPPED AND SEVERED AT THE SIDEPORT. THERE WAS NO HARM TO THE PATIENT REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7426923 ↗

Other STENT, URETERAL devices

All STENT, URETERAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page