Cook G17080Recalls & reported problems
Cook Incorporated · Balloon Uterine Stent · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002170803), received between Sep 14, 2016 and Jan 3, 2019 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Contamination /Decontamination Problem1
- Deflation Problem1
Patient problems
- Anxiety1
- No Code Available1
- No Consequences Or Impact To Patient1
- Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 3, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT IS REPORTED THAT WHILE PREPARING FOR AN UNSPECIFIED PROCEDURE USING A BALLOON UTERINE STENT, THE END USER OPENED THE PACKAGE TO INSPECT THE DEVICE PRIOR TO USE AND FOUND AN UNKNOWN BLACK STAIN ON THE BALLOON. THIS DEVICE WAS NOT USED. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE ALLEGED MALFUNCTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8214403 ↗InjurySep 14, 2016Deflation Problem; Deflation Problem
IT WAS REPORTED ON (B)(6) 2016 A PHYSICIAN PERFORMED AN OPEN MYOMECTOMY USING AN INTRAUTERINE BALLOON STENT. THE DEVICE WAS REPORTED TO HAVE BEEN TESTED BY INFLATING AND DEFLATING THE BALLOON PRIOR TO INSERTION AND THE DOCTOR EXPERIENCED DIFFICULTY DEFLATING THE BALLOON, HAVING TO USE MORE FORCE AND TWISTING WITH THE SYRINGE TO OPEN THE VALVE AND RELEASE THE FLUID. ON (B)(6), 24 HOURS AFTER INSERTING THE DEVICE, THE PHYSICIANS MADE THE FIRST UNSUCCESSFUL ATTEMPT TO REMOVE THE DEVICE USING THE SYRINGE. THE DOCTOR ATTEMPTED TO DEFLATE THE BALLOON FOR APPROXIMATELY 30 MINUTES USING EXCESSIVE FORCE, TWISTING, CUTTING THE VALVE OFF AND TRYING A DIFFERENT SYRINGE. SUPPORT WAS CALLED AND THE HOSPITAL STAFF WAS ADVISED BY THE MANUFACTURER TO USE ULTRASOUND TO DETERMINE THE BALLOON VOLUME AND APPLY LUBRICATION. THREE HOURS LATER THE ATTENDING PHYSICIAN'S PARTNER CALLED SUPPORT AND REPORTED THE BALLOON WAS APPARENTLY FULLY INFLATED AND SHE COULD SEE THE BALLOON WHEN TRYING TO MANIPULATE IT. AT THIS TIME THE PARTNER SUGGESTED PUNCTURING THE BALLOON WITH A NEEDLE. THE PHYSICIAN ADMINISTERED A PUDENDAL BLOCK AND CERVICAL BLOCK ON THE PATIENT FOR PAIN AND USING A PUDENDAL NEEDLE WAS ABLE TO PUNC
Patient problems: Anxiety; No Code Available; Pain
MAUDE report 5948799 ↗
Other INSTRUMENT, MANUAL, SPECIALIZED OBSTETRIC-GYNECOLOGIC devices
- Centurion HUD56STCenturion Medical Products Corporation · INSTRUMENT, MANUAL, SPECIALIZED OBSTETRIC-GYNECOLOGIC1 recall0 reports
- Centurion HUD112STCenturion Medical Products Corporation · INSTRUMENT, MANUAL, SPECIALIZED OBSTETRIC-GYNECOLOGIC1 recall0 reports
- Centurion HUD12STCenturion Medical Products Corporation · INSTRUMENT, MANUAL, SPECIALIZED OBSTETRIC-GYNECOLOGIC1 recall0 reports
- Centurion HUD910STCenturion Medical Products Corporation · INSTRUMENT, MANUAL, SPECIALIZED OBSTETRIC-GYNECOLOGIC1 recall0 reports
- Alexis ® O C-Section Protector - Retractor G6313Applied Medical Resources Corporation · Instrument, Manual, Specialized Obstetric-Gynecologic14 reports
- AESCULAP PL430RAesculap AG · INSTRUMENT, MANUAL, SPECIALIZED OBSTETRIC-GYNECOLOGIC3 reports
- Alexis ® O C-Section Protector - Retractor G6314Applied Medical Resources Corporation · Instrument, Manual, Specialized Obstetric-Gynecologic2 reports
- Word G55442Cook Incorporated · INSTRUMENT, MANUAL, SPECIALIZED OBSTETRIC-GYNECOLOGIC2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.