Skip to content

Cook G17870Recalls & reported problems

Cook Incorporated · Cook, Silicone Balloon HSG Catheter · Class 2

Quick summary

This Cook Silicone Balloon HSG catheter is used to inflate a balloon inside the uterus to check the fallopian tubes. There are 44 adverse event reports, with no deaths and one injury described. Most reports describe the balloon failing to deflate or bursting, sometimes found before use or separating, and in nearly all cases the patient was not affected. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
44
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 44 adverse event reports that list this exact device (DI 00827002178700), received between Feb 9, 2018 and Jan 9, 2025 — classified as 1 injury, 43 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (25 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem25
  • Material Rupture17
  • Device Contaminated During Manufacture or Shipping1
  • Material Separation1
  • Suction Problem1

Patient problems

  • No Consequences Or Impact To Patient22
  • No Clinical Signs, Symptoms or Conditions14
  • No Patient Involvement4
  • No Known Impact Or Consequence To Patient3
  • Device Embedded In Tissue or Plaque1
  • Discomfort1

Recent adverse event reports

The 25 most recent of 44 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 9, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21111817 ↗
  • MalfunctionDec 16, 2021Material Rupture; Material Rupture

    NO NEW PATIENT OR EVENT INFORMATION TO REPORT SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13017439 ↗
  • MalfunctionDec 12, 2021Material Rupture; Material Rupture

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12981901 ↗
  • MalfunctionDec 8, 2021Material Rupture; Material Rupture

    THERE IS NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12962801 ↗
  • MalfunctionDec 8, 2021Material Rupture; Material Rupture

    THERE IS NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12962737 ↗
  • MalfunctionDec 8, 2021Material Rupture; Material Rupture

    THERE IS NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12961381 ↗
  • MalfunctionDec 2, 2021Material Rupture; Material Rupture

    AS REPORTED, THE OPERATOR CHECKED THE COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER PRIOR TO USE AND DISCOVERED THE BALLOON WAS RUPTURED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER NEW DEVICE. NO ADVERSE EFFECTS HAVE BEEN REPORTED DUE TO THE ALLEGED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12920865 ↗
  • MalfunctionDec 2, 2021Material Rupture; Material Rupture

    AS REPORTED, THE OPERATOR CHECKED THE COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER PRIOR TO USE AND DISCOVERED THE BALLOON WAS RUPTURED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER NEW DEVICE. NO ADVERSE EFFECTS HAVE BEEN REPORTED DUE TO THE ALLEGED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12920725 ↗
  • MalfunctionAug 13, 2021Material Rupture; Material Rupture

    NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12320460 ↗
  • MalfunctionJun 11, 2021Material Rupture; Material Rupture

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11982859 ↗
  • MalfunctionMay 6, 2021Material Rupture; Material Rupture

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11782635 ↗
  • MalfunctionApr 30, 2021Material Rupture; Material Rupture

    AS REPORTED, A COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER RUPTURED IN THE PATIENT WHEN INFLATED WITH 1.3ML SALINE. THE DEVICE WAS NOT TESTED PRIOR TO USE. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED AS A RESULT OF THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11755620 ↗
  • MalfunctionApr 20, 2021Material Rupture; Material Rupture

    AS REPORTED, A COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER LEAKED DURING AN UNSPECIFIED PROCEDURE. THE LEAKAGE OCCURRED AT THE JOINT OF THE INFLATION PORT. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED AS A RESULT OF THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11692706 ↗
  • MalfunctionFeb 2, 2021Material Rupture; Material Rupture

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11269713 ↗
  • MalfunctionJan 6, 2021Material Rupture; Material Rupture

    NO NEW PATIENT OR EVENT INFORMATION SINCE THE LAST REPORT WAS SUBMITTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11122958 ↗
  • MalfunctionJan 29, 2020Deflation Problem; Deflation Problem

    NO ADDITIONAL PATIENT OR EVENT INFORMATION SINCE THE LAST REPORT WAS SUBMITTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9641546 ↗
  • MalfunctionJan 27, 2020Deflation Problem; Deflation Problem

    NO NEW PATIENT OR EVENT INFORMATION SINCE THE LAST REPORT WAS SUBMITTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9631381 ↗
  • MalfunctionJan 27, 2020Deflation Problem; Deflation Problem

    NO ADDITIONAL PATIENT OR EVENT INFORMATION HAS BEEN RECEIVED SINCE THE LAST REPORT WAS SUBMITTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9631285 ↗
  • MalfunctionDec 19, 2019Material Rupture; Material Rupture

    NO NEW PATIENT OR EVENT INFORMATION HAS BEEN PROVIDED SINCE THE LAST REPORT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9497014 ↗
  • MalfunctionNov 1, 2019Material Rupture; Material Rupture

    ADDITIONAL INFORMATION WAS RECEIVED ON 14NOV2019: A HYSTEROSONOGRAPHY PROCEDURE WAS BEING PERFORMED. ANOTHER SAME TYPE DEVICE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9267928 ↗
  • MalfunctionAug 12, 2019Deflation Problem; Deflation Problem

    UPDATE 21AUG2019: THE PROCEDURE BEING PERFORMED WAS A SONOISTEROSALPINGOGRAPHY IN VAGINA, CERVIX, AND UTERUS. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8884043 ↗
  • MalfunctionAug 12, 2019Deflation Problem; Deflation Problem

    UPDATE 21AUG2019: THE PROCEDURE BEING PERFORMED WAS A SONOISTEROSALPINGOGRAPHY IN VAGINA, CERVIX, AND UTERUS. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8883918 ↗
  • MalfunctionAug 12, 2019Deflation Problem; Deflation Problem

    UPDATE (B)(6) 2019: THE PROCEDURE BEING PERFORMED WAS A SONOISTEROSALPINGOGRAPHY IN VAGINA, CERVIX, AND UTERUS. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8883830 ↗
  • MalfunctionAug 12, 2019Deflation Problem; Deflation Problem

    NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8883733 ↗
  • MalfunctionAug 12, 2019Deflation Problem; Deflation Problem

    IT WAS REPORTED, DURING AN UNKNOWN PROCEDURE USING AN COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER, THERE WAS DIFFICULTY DEFLATING THE BALLOON. EVENTUALLY THE USER WAS ABLE TO DEFLATE THE BALLOON. THERE WERE MORE THAN ONE OCCURRENCE OF THIS TYPE OF EVENT. (PATIENT REFERENCE IDENTIFIER (B)(6)) ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. ADDITIONAL DETAILS HAVE BEEN REQUESTED REGARDING THE PATIENT AND EVENT. AT THIS TIME NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8883634 ↗

Other Cannula, manipulator/injector, uterine devices

All Cannula, manipulator/injector, uterine devices →