Skip to content

Zilver Flex G18282Recalls & reported problems

Cook Ireland Ltd · Zilver Flex 35 Vascular Stent · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002182820), received between May 20, 2026 and May 20, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 20, 2026Physical Resistance/Sticking

    DURING STENT DEPLOYMENT, THE PHYSICIAN OBTAINED RETROGRADE TIBI-PERONEAL ACCESS AND ATTEMPTED TO EXTEND THE STENT INTO HUNTER¿S CANAL. THE PHYSICIAN STATED ¿FELT LIKE THE STENT WASN'T FULLY OPENING" . A GUIDEWIRE COULD NOT BE ADVANCED THROUGH THE NEWLY PLACED STENT, ALTHOUGH IMAGING SUGGESTED IT WAS OPEN. MULTIPLE ATTEMPTS WERE MADE TO CROSS THE STENT, INCLUDING THE USE OF A BALLOON, WITHOUT SUCCESS. PUSHING THE STENT RESULTED IN DEFORMATION, CAUSING THE STENT TO COLLAPSE/CRUSH INTO ITSELF. THE PROCEDURE WAS DISCONTINUED, AND A SECOND PROCEDURE WAS SCHEDULED FOR TWO WEEKS LATER [(B)(6) 2026]. DETAILS OF ACCESS SHEATH USED (NAME, FR SIZE, LENGTH)? - 6FR TERUMO SLENDER WHAT WAS THE TARGET LOCATION FOR THE STENT? - SFA (SUPERFICIAL FEMORAL ARTERY) WAS THE DEVICE USED PERCUTANEOUSLY? - YES WAS PRE-DILATION PERFORMED AHEAD OF PLACEMENT OF THE STENT? - YES WHAT WAS THE ACCESS SITE CHOSEN? - COMMON FEMORAL ARTERY WITH RETROGRADE WAS A STENT PREVIOUSLY PLACED DURING PREVIOUS PROCEDURES? - NO DID THE PATIENT EXHIBIT DIFFICULT ANATOMY? - HEAVILY CALCIFIED WAS THE APPROACH IPSILATERAL OR CONTRALATERAL? - IPSILATERAL RETROGRADE ACCESS WHAT INTERVENTION (IF ANY) WAS REQUIRED? - ADDITIONAL PROCE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25232294 ↗

Other STENT, SUPERFICIAL FEMORAL ARTERY devices

All STENT, SUPERFICIAL FEMORAL ARTERY devices →

Sources and updates