CXI G18367Recalls & reported problems
Cook Incorporated · CXI Support Catheter · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002183674), received between Apr 24, 2018 and Aug 30, 2018 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Kinked” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Kinked1
- Material Perforation1
Patient problems
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 30, 2018Material Perforation; Material Perforation
IT WAS REPORTED, ANOTHER MANUFACTURER'S GUIDEWIRE PERFORATED THROUGH THE SIDE OF THE CXI SUPPORT CATHETER DURING ANGIOPLASTY OF A CALCIFIED LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA). THE COMPLAINT DEVICE WAS BEING USED FROM THE DISTAL SFA IN A RETROGRADE MANNER TO CROSS THE OCCLUSION. IT IS NOT KNOWN AT WHAT POINT IN THE PROCEDURE THE COMPLAINT DEVICE WAS PERFORATED BECAUSE IT WASN'T DISCOVERED UNTIL IT'S REMOVAL FROM THE PATIENT'S ANATOMY. ACCORDING TO THE INITIAL REPORTER, THERE WAS NO HARM TO THE PATIENT OR ADDITIONAL PROCEDURE REQUIRED AS A RESULT OF THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7834508 ↗InjuryApr 24, 2018Kinked; Kinked
IT WAS REPORTED THAT THE 0.018 HYDROPHILIC WIRE UTILIZED THROUGH THE CXI SUPPORT CATHETER KINKED WHILE INSIDE THE PATIENT'S VASCULATURE AND REQUIRED A SURGICAL INTERVENTION TO REMOVE IT. ACCESS WAS GAINED VIA THE PEDAL ARTERY. AS THE PHYSICIAN WAS TRYING TO ADVANCE THE WIRE THROUGH THE HEAVILY CALCIFIED, NARROWED, POPLITEAL ARTERY LESION, IT KINKED. ADDITIONAL ATTEMPTS USING OTHER CXI SUPPORT CATHETER SIZES WERE UNSUCCESSFUL DUE TO THE HEAVY CALCIFICATION OF THE LESION; THEREFORE, THE PROCEDURE WAS ABORTED. NO OTHER DEVICE MALFUNCTIONS WERE REPORTED. ACCORDING TO THE INITIAL REPORTER, A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. NO ADDITIONAL ADVERSE EFFECT WAS REPORTED.
Patient problems: No Code Available
MAUDE report 7454684 ↗
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.