Cook Incorporated · CXI Support Catheter · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002183728), received between Dec 7, 2017 and Jun 5, 2024 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem2
  • Break1
  • Fitting Problem1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Code Available1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 5, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS ORIGINALLY REPORTED, DURING A RUN-OFF ANGIOGRAM OF THE RIGHT LEG, A WIRE WAS DIFFICULT TO ADVANCE THROUGH A CXI SUPPORT CATHETER, WHICH WAS ALSO DIFFICULT TO FLUSH. ACCESS WAS OBTAINED IN THE SUPERFICIAL FEMORAL ARTERY TO TARGET THE PERONEAL ARTERY. THE ANATOMY WAS CALCIFIED. THE HYDROPHILIC COATING WAS ACTIVATED, AND THE PHYSICIAN USED THE CATHETER ONE TIME. WHEN THE USER ATTEMPTED TO RE-USE THE CATHETER FOR A SECOND INSERTION, THE CATHETER WAS UNABLE TO BE FLUSHED, AND AN UNSPECIFIED WIRE WOULD NOT ADVANCE THROUGH THE CATHETER. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. UPON RETURN AND INITIAL EVALUATION OF THE DEVICE, AN UNKNOWN PIECE OF FOREIGN MATTER WAS INSIDE THE DISTAL TIP OF THE CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19467068 ↗
  • MalfunctionDec 16, 2022Break; Break

    AS REPORTED, PRIOR TO USE FOR AN UNKNOWN PROCEDURE, THE TIP OF A CXI SUPPORT CATHETER WAS NOTED TO BE SEPARATED WHEN THE USER PULLED THE CATHETER FROM THE SPIRAL HOLDER. ANOTHER MANUFACTURER'S DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE DEVICE HAD BEEN STORED VERTICALLY AT THE USER FACILITY AND THE PACKAGE WAS NOT DAMAGED. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15997330 ↗
  • MalfunctionNov 14, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem

    NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9322725 ↗
  • InjuryDec 7, 2017Material Separation; Fitting Problem; Material Separation; Fitting Problem

    IT WAS REPORTED THAT DURING A PROCEDURE, THE CXI SUPPORT CATHETER HUB BROKE OFF DURING REMOVAL OF THE DEVICE AND CATHETER TUBING SEPARATED. THE PHYSICIAN UNSUCCESSFULLY ATTEMPTED TO CROSS THE PATIENT'S CHRONIC TOTAL OCCLUSION (CTO) FROM AN UP AND OVER APPROACH, THEN PROCEEDED TO GAIN ACCESS FROM THE DORSALIS PEDIS USING A PEDAL MICROPUNCTURE AND TRIED TO CROSS THE CTO RETROGRADE. A 25 GAUGE APPROACH CTO MICROGUIDE WIRE WITHOUT THE PEDAL SHEATH, USING THE CXI SUPPORT CATHETER OVER THE MIRCROGUIDE WIRE, WAS USED TO SUCCESSFULLY CROSS THE CTO WITH NOTED DIFFICULTLY. THE MICROGUIDE WIRE WAS ADVANCED THROUGH THE CTO, AND AS THE PHYSICIAN ATTEMPTED TO REMOVE THE CXI SUPPORT CATHETER IT FELT STUCK TO THE MICROGUIDE WIRE, PROMPTING HIM TO FORCIBLY PULL ON THE CATHETER HUB. THE CATHETER HUB COMPLETELY BROKE OFF OF THE CATHETER, AND THE PHYSICIAN ATTEMPTED REMOVAL OF THE CATHETER WITHOUT THE HUB BY PULLING ON IT FOR 20- 30 MINUTES. THE CATHETER WAS ABLE TO BE COMPLETELY REMOVED; HOWEVER IT BROKE INTO PIECES DURING REMOVAL OVER THE MICROGUIDE WIRE AS THE PHYSICIAN HAD TO USE FORCE IN ORDER TO MAINTAIN WIRE ACCESS. THE MICROGUIDE WIRE WAS SNARED FROM THE UP AND OVER APPROACH, AS INITIALLY INTE