CXI G18372Recalls & reported problems
Cook Incorporated · CXI Support Catheter · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002183728), received between Dec 7, 2017 and Jun 5, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Contamination /Decontamination Problem2
- Break1
- Fitting Problem1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 5, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
AS ORIGINALLY REPORTED, DURING A RUN-OFF ANGIOGRAM OF THE RIGHT LEG, A WIRE WAS DIFFICULT TO ADVANCE THROUGH A CXI SUPPORT CATHETER, WHICH WAS ALSO DIFFICULT TO FLUSH. ACCESS WAS OBTAINED IN THE SUPERFICIAL FEMORAL ARTERY TO TARGET THE PERONEAL ARTERY. THE ANATOMY WAS CALCIFIED. THE HYDROPHILIC COATING WAS ACTIVATED, AND THE PHYSICIAN USED THE CATHETER ONE TIME. WHEN THE USER ATTEMPTED TO RE-USE THE CATHETER FOR A SECOND INSERTION, THE CATHETER WAS UNABLE TO BE FLUSHED, AND AN UNSPECIFIED WIRE WOULD NOT ADVANCE THROUGH THE CATHETER. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. UPON RETURN AND INITIAL EVALUATION OF THE DEVICE, AN UNKNOWN PIECE OF FOREIGN MATTER WAS INSIDE THE DISTAL TIP OF THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19467068 ↗MalfunctionDec 16, 2022Break; Break
AS REPORTED, PRIOR TO USE FOR AN UNKNOWN PROCEDURE, THE TIP OF A CXI SUPPORT CATHETER WAS NOTED TO BE SEPARATED WHEN THE USER PULLED THE CATHETER FROM THE SPIRAL HOLDER. ANOTHER MANUFACTURER'S DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE DEVICE HAD BEEN STORED VERTICALLY AT THE USER FACILITY AND THE PACKAGE WAS NOT DAMAGED. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15997330 ↗MalfunctionNov 14, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem
NO NEW PATIENT OR EVENT INFORMATION TO REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9322725 ↗InjuryDec 7, 2017Material Separation; Fitting Problem; Material Separation; Fitting Problem
IT WAS REPORTED THAT DURING A PROCEDURE, THE CXI SUPPORT CATHETER HUB BROKE OFF DURING REMOVAL OF THE DEVICE AND CATHETER TUBING SEPARATED. THE PHYSICIAN UNSUCCESSFULLY ATTEMPTED TO CROSS THE PATIENT'S CHRONIC TOTAL OCCLUSION (CTO) FROM AN UP AND OVER APPROACH, THEN PROCEEDED TO GAIN ACCESS FROM THE DORSALIS PEDIS USING A PEDAL MICROPUNCTURE AND TRIED TO CROSS THE CTO RETROGRADE. A 25 GAUGE APPROACH CTO MICROGUIDE WIRE WITHOUT THE PEDAL SHEATH, USING THE CXI SUPPORT CATHETER OVER THE MIRCROGUIDE WIRE, WAS USED TO SUCCESSFULLY CROSS THE CTO WITH NOTED DIFFICULTLY. THE MICROGUIDE WIRE WAS ADVANCED THROUGH THE CTO, AND AS THE PHYSICIAN ATTEMPTED TO REMOVE THE CXI SUPPORT CATHETER IT FELT STUCK TO THE MICROGUIDE WIRE, PROMPTING HIM TO FORCIBLY PULL ON THE CATHETER HUB. THE CATHETER HUB COMPLETELY BROKE OFF OF THE CATHETER, AND THE PHYSICIAN ATTEMPTED REMOVAL OF THE CATHETER WITHOUT THE HUB BY PULLING ON IT FOR 20- 30 MINUTES. THE CATHETER WAS ABLE TO BE COMPLETELY REMOVED; HOWEVER IT BROKE INTO PIECES DURING REMOVAL OVER THE MICROGUIDE WIRE AS THE PHYSICIAN HAD TO USE FORCE IN ORDER TO MAINTAIN WIRE ACCESS. THE MICROGUIDE WIRE WAS SNARED FROM THE UP AND OVER APPROACH, AS INITIALLY INTE
Patient problems: No Code Available
MAUDE report 7095348 ↗
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.