Coda G23773Recalls & reported problems
Cook Incorporated · Coda, Balloon Catheter · Class 2
All models of Coda (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00827002237735), received between Oct 23, 2017 and Jul 14, 2025 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture3
- Device Contaminated During Manufacture or Shipping2
- Device Contamination with Chemical or Other Material1
- Use of Device Problem1
Patient problems
- No Patient Involvement3
- Great Vessel Perforation1
- Hemorrhage/Blood Loss/Bleeding1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- Paraplegia1
- Shock1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryJul 14, 2025Use of Device Problem; Use of Device Problem
COOK WAS NOTIFIED THAT A RUPTURE OF THE ABDOMINAL AORTA OCCURRED AFTER BALLOONING WITH A COOK CODA BALLOON. THE INDICATION FOR THE ELECTIVE PROCEDURE WAS A POST-DISSECTING THORACIC ABDOMINAL AORTIC ANEURYSM (TAAA) TYPE I EXTENT, CRAWFORD CLASSIFICATION. A RESCUE REPAIR HAD BEEN PERFORMED AFTER PRIOR OPEN REPAIR WITH A THORACIC SURGICAL GRAFT FOR A TYPE A DISSECTION. PRE-PROCEDURAL COMPUTED TOMOGRAPHY (CT) WITH CONTRAST IMAGING WAS COMPLETED ON (B)(6) 2024 (151 DAYS PRIOR TO THE PROCEDURE). THE BILATERAL COMMON ILIAC AND INTERNAL ILIAC ARTERIES WERE PATENT. THE BILATERAL VERTEBRAL ARTERIES WERE PATENT. THE AORTIC VALVE WAS NATIVE. THE MAXIMUM DIAMETER OF THE DISEASED AORTA ON CENTER LINE WAS 68 MM. THE INTENDED PROXIMAL LANDING ZONE WAS ZONE 2: LEFT COMMON CAROTID TO LEFT SUBCLAVIAN ARTERY. THE INTENDED DISTAL LANDING ZONE WAS ZONE 5: MID DESCENDING AORTA TO CELIAC ARTERY. THE PATIENT UNDERWENT THE FIRST PROCEDURE OF THE STAGED PROCEDURES ON (B)(6) 2025 UNDER GENERAL ANESTHESIA. THE PATIENT WAS PRESCRIBED ACETYLSALICYLIC ACID (ASA) AT THE TIME OF THE PROCEDURE. PERCUTANEOUS ACCESS WAS ACHIEVED IN THE RIGHT AND LEFT FEMORAL ARTERIES. THE LEFT FEMORAL VEIN WAS ACCESSED FOR CARDIAC RAP
Patient problems: Great Vessel Perforation; Hemorrhage/Blood Loss/Bleeding; Paraplegia; Shock; Spinal Cord Injury
MAUDE report 22500445 ↗MalfunctionApr 22, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A CODA BALLOON CATHETER RUPTURED. DURING AN OPERATION ON (B)(6) 2025 AT 14:23 HOURS, THE BALLOON OF THE CODA DEVICE BURST DURING INFLATION IN THE PATIENT'S AORTA. A SECOND CODA FROM THE SAME LOT WAS OBTAINED AND EXPERIENCED THE SAME FAILURE. FINALLY, A COMPETITOR'S BALLOON WAS USED TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING EVENT DETAILS HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21884358 ↗MalfunctionOct 24, 2019Material Rupture; Material Rupture
NO ADDITIONAL PATIENT/EVENT DETAILS HAVE BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9235304 ↗MalfunctionSep 10, 2019Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED A "STRANGE PARTICLE" WAS FOUND INSIDE THE PRIMARY PACKAGING OF A CODA BALLOON CATHETER. NO PATIENT CONTACT WAS MADE AS THE ISSUE WAS NOTICED BEFORE THE PRODUCT WAS OPENED. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT.
Patient problems: No Patient Involvement
MAUDE report 8983198 ↗MalfunctionJul 16, 2019Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
NO ADDITIONAL INFORMATION REGARDING THE EVENT AND/OR PATIENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Patient Involvement
MAUDE report 8795886 ↗MalfunctionJun 13, 2018Material Rupture
IT WAS REPORTED THAT DURING THE TEVAR PROCEDURE, WHEN THE CODA BALLOON CATHETER WAS USED, THE BALLOON RUPTURED AFTER INFLATION. THERE WAS A DELAY REPORTED AND ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7597105 ↗MalfunctionOct 23, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
THE DISTRIBUTOR REPORTED THAT DURING INSPECTION OF THE CODA BALLOON CATHETER IT WAS OBSERVED THAT THERE WAS A HAIR INSIDE THE PRIMARY PACKAGE. THIS PRODUCT WAS NOT USED ON A PATIENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 6969829 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page