Cook G26945Recalls & reported problems

Cook Incorporated · Central Venous Tray Triple Lumen Polyurethane Catheter · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002269453), received between Nov 9, 2017 and Nov 7, 2018 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Crack” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Crack2
  • Partial Blockage1

Patient problems

  • No Code Available2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 7, 2018Crack; Crack

    IT WAS REPORTED THE BLUE HUB ON A TRIPLE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER WAS FOUND TO BE CRACKED CAUSING LEAKAGE. THE DEVICE WAS PLACED IN THE RIGHT FEMORAL VEIN ON (B)(6) 2018 AND FOUND CRACKED ON (B)(6) 2018 A FEW HOURS AFTER THE SMARTSITE NEEDLE FREE VALVE WAS CHANGED. THE PATIENT WAS RECEIVING CONTINUOUS AND SINGLE MEDICATION INFUSIONS. THE DEVICE WAS REMOVED AND REPLACED WITH A LIKE DEVICE. MEDWATCH MW5080167 WAS RECEIVED ON 26OCT2018 DOCUMENTING THIS EVENT. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Code Available

    MAUDE report 8045605 ↗
  • InjuryAug 1, 2018Crack; Crack

    SEE H10.

    Patient problems: No Code Available

    MAUDE report 7741037 ↗
  • MalfunctionNov 9, 2017Partial Blockage; Partial Blockage

    ADDITIONAL INFORMATION: PEDIATRIC PHYSICIAN WAS PLACING A RIGHT TRIPLE LUMEN INTRAJUGULAR (IJ) LINE. THIS MDR IS BEING REPORTED BASED ON INVESTIGATION OF THE WIRE GUIDE, IT WAS DISCOVERED THE DISTAL WELD CONNECTION LOCATED AT THE ¿J¿ TIP END WAS BROKEN, RESULTING IN COIL ELONGATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7014711 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page