Skip to content

Fusion Quattro G31921Recalls & reported problems

Cook Incorporated · Fusion Quattro Extraction Balloon · Class 2

All models of Fusion Quattro (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002319219), received between Dec 7, 2017 and Nov 19, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Burst Container or Vessel1
  • Deflation Problem1
  • Detachment of Device or Device Component1
  • Material Protrusion/Extrusion1

Patient problems

  • No Consequences Or Impact To Patient2
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • Perforation1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 19, 2025Deflation Problem; Deflation Problem

    DURING AN UNKNOWN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED A COOK FUSION QUATTRO EXTRACTION BALLOON FOR STONE REMOVAL. IT WAS REPORTED THAT THE BALLOON WAS CONFIRMED TO BE NORMAL AND FUNCTIONAL BEFORE USE. THE BALLOON WAS INFLATED INSIDE THE BILIARY DUCT. WHILE ATTEMPTING TO DEFLATE THE BALLOON NEAR THE AMPULLA DURING STONE REMOVAL, THE NURSE TRIED TO MANIPULATE THE SYRINGE THAT WAS INCLUDED IN THE PRODUCT PACKAGING, BUT IT DID NOT MOVE, AND THE BALLOON DID NOT DEFLATE. AS SUCH, THE BALLOON WAS FORCEFULLY BURST AND THEN REMOVED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23596903 ↗
  • MalfunctionDec 6, 2019Burst Container or Vessel; Detachment of Device or Device Component; Burst Container or Vessel; Detachment of Device or Device Component

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK FUSION QUATTRO EXTRACTION BALLOON. A STONE WAS TAKEN OUT WITH THE FIRST USE OF THE DEVICE, BUT WHEN THEY ATTEMPTED TO TAKE OUT THE REST OF THE STONES, THEY FOUND OUT THE BALLOON WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER COOK FUSION QUATTRO EXTRACTION BALLOON. THIS EVENT WAS NOT REPORTABLE AT THAT TIME. ON 12-NOV-2019, THE DEVICE RETURNED. THE SECTIONS OF THE BALLOON DO NOT MATCH UP [PIECES OF THE BALLOON MATERIAL ARE MISSING]. IT WAS REPORTED THAT A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY; HOWEVER THE LOCATION OF THE MISSING PORTIONS IS UNKNOWN. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: Foreign Body In Patient; No Consequences Or Impact To Patient

    MAUDE report 9432763 ↗
  • MalfunctionDec 12, 2017Material Protrusion/Extrusion; Material Protrusion/Extrusion

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK FUSION QUATTRO EXTRACTION BALLOON. ALL THE INFORMATION I HAVE IS ON A NOTE THAT IS STUCK TO THE PACKAGE WHICH SAYS, "SPLIT AT END BY ANOTHER MANUFACTURER'S WIRE GUIDE." THE CUSTOMER DOES NOT RECALL HOW OR WHEN THIS OCCURRED. I DO NOT KNOW WHEN THIS OCCURRED. THE DEVICE WAS EVALUATED ON 11/17/2017 AND IT WAS OBSERVED THAT THE INTRADUCTAL EXCHANGE (IDE) PORT WAS DAMAGED AND PROTRUDING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7108435 ↗
  • InjuryDec 7, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK FUSION QUATTRO EXTRACTION BALLOON (FS-QEB-XL-A). PER THE INITIAL REPORTER: ¿DR. (B)(6) USED [FOR THE] FIRST TIME A FS-QEB-XL-A EXTRACTION BALLOON; BEFORE, HE HAD ALWAYS USED ANOTHER MANUFACTURER'S EXTRACTION BALLOONS. FIRST, A SPHINCTEROTOMY WAS PERFORMED AND ANOTHER MANUFACTURER'S WIRE GUIDE WAS INTRODUCED IN BILE DUCT OVER WHICH THE FS-QEB-XL-A WAS ADVANCED INTO THE DUCT. ACCORDING TO THE NURSE, THE BALLOON WAS INFLATED UP TO 15 MM. WHEN THE BALLOON WAS RETRACTED, IT SUDDENLY BURST. THE PHYSICIAN STATES THAT THIS CAUSED A PERFORATION OF THE PAPILLARY ROOF (WITH VISIBILITY OF THE INTRAMURAL PART OF THE BILE DUCT). A CLIP WAS PLACED TO CLOSE THE PERFORATION, AND A COOK 10 FRENCH, 7 CM COTTON LEUNG STENT WAS PLACED INTO THE BILE DUCT. A LITTLE OOZING BLEEDING WAS VISIBLE, BUT WITHOUT NEED OF FURTHER INTERVENTION. LATER, THE BLEEDING GOT MORE INTENSIVE, SO THE PATIENT RECEIVED TWO BLOOD TRANSFUSIONS, PLUS ANTIBIOTICS AND WAS ADMITTED TO THE INTENSIVE CARE UNIT. THE BALLOON WAS DISCHARGED [DISCARDED] DURING THE STRESSFUL SITUATION AND IS NOT AVAILABLE FOR FURTHER EXAMINATIONS [INVESTIGATION].¿ T

    Patient problems: Perforation

    MAUDE report 7097249 ↗

Other CATHETER, BILIARY, SURGICAL devices

All CATHETER, BILIARY, SURGICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page