One-Tie G31929Recalls & reported problems
Cook Vandergrift Inc. · One-Tie Compression Coil · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 2
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002319295), received between Jul 29, 2021 and Aug 13, 2026 — classified as 2 deaths, 1 injury.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Improper or Incorrect Procedure or Method1
Patient problems
- Great Vessel Perforation1
- No Clinical Signs, Symptoms or Conditions1
- Thrombosis/Thrombus1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
DeathAug 13, 2026Adverse Event Without Identified Device or Use Problem
The following information was reported to Cook Customer Relations by Merit Medical Systems, Inc.: Customer reports they had a lead extraction. This was a second attempt after failing a lead extraction on the previous Saturday. Patient was very sick and had an infection and lots of other co-morbidities. They began the extraction with a few different tools, Byrd Dilator Sheaths, Philips Laser, 9Fr Evolution. They were able to remove the left ventricle (LV) lead but still needed to remove the right ventricle (RV) lead. They tried with the same tools and also upsized to a 13Fr Evolution, and larger laser tool with no success of removal of the RV lead. The patient's vitals were not at a safe point to keep working on them, so they decided it was best to stop where we were and possibly bring him back and approach it from a femoral removal and use medical therapy in the short term to manage infection. As they were closing the groin and pocket area the patient began decompensating and they looked under echo and saw that the heart was slowing down and beginning to clot off and "[the patient]" ultimately expired. They confirmed that this had nothing to do with the devices themselves but just
Patient problems: Thrombosis/Thrombus
MAUDE report 50238316 ↗DeathApr 15, 2026Adverse Event Without Identified Device or Use Problem
A MERIT MEDICAL EMPLOYEE CONDUCTED A POST-MARKET CLINICAL FOLLOW UP (PMCF) REVIEW FOR THE MERIT LEAD SUPPORT DEVICES. MERIT MEDICAL THEN FORWARDED THE STUDY INFORMATION TO COOK MEDICAL. THE REFERENCED JOURNAL ARTICLE/STUDY ("TRANSVENOUS LEAD EXTRACTION: EXPERIENCE OF THE TANDEM APPROACH", AKHTAR, ET AL.) DESCRIBES THE USE OF A TANDEM APPROACH DURING TRANSVENOUS LEAD EXTRACTION PROCEDURES IN PATIENTS WITH CHRONICALLY IMPLANTED CARDIAC DEVICES WITH LEADS. THE ONE-TIE COMPRESSION COIL WAS USED TO STABILIZE THE LEAD AND IMPROVE TRACTION DURING EXTRACTION. COMPLICATIONS REPORTED INCLUDE KNOWN EXTRACTION RISKS SUCH AS VASCULAR INJURY, PERICARDIAL EFFUSION, AND PROCEDURAL COMPLICATIONS RELATED TO LEAD EXTRACTION. NO DEVICE MALFUNCTION OR PERFORMANCE ISSUES ASSOCIATED WITH THE ONE-TIE COMPRESSION COIL WERE REPORTED. HOWEVER, ONE PROCEDURE INVOLVING A 61-YEAR-OLD FEMALE LEAD TO A TEAR IN THE SUPERIOR VENA CAVA TO RIGHT ATRIUM (SVC-RA) JUNCTION RESULTING IN PERICARDIAL TAMPONADE. THIS EVENT REQUIRED STERNOTOMY AND REPAIR OF THE INJURY. THE PATIENT DIED TWO DAYS POST PROCEDURE.
Patient problems: Great Vessel Perforation
MAUDE report 24887545 ↗InjuryJul 29, 2021Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE PHYSICIAN MADE LEAD CONTROL IN THE RIGHT ATRIAL LEAD WITH LIBERATOR, ONE-TIE, AND EVOLUTION SHORTIE 9 FRENCH. ACCESS FOR PROCEDURE WAS GAINED FROM THE LEFT SUBCLAVIAN VEIN. AFTER THE TIP OF THE LEAD WAS REMOVED, LEAD EXTRACTION WAS ATTEMPTED USING A SNARE (OTHER MANUFACTURER) WITH A FEMORAL APPROACH. AFTER GRASPING THE TIP OF THE LEAD WITH THE SNARE, THE PHYSICIAN CONFIRMED THAT THE LEAD COMING OUT OF THE UPPER LIMB AND THE PORTION HELD BY THE SNARE WERE PULL-THROUGH, WITHOUT ADHESIONS IN THE MIDDLE. HOWEVER, THE PATIENT'S SIDE (DISTAL END OF THE ONE-TIE) SHOULD HAVE BEEN CUT OFF, BUT THE SURGEON'S SIDE (PROXIMAL END OF THE ONE-TIE) WAS CUT OFF INSTEAD. AS A RESULT, THE LEAD WITH THE ONE-TIE ATTACHED WAS PULLED INTO THE FEMORAL REGION BY THE SNARE. ANOTHER PHYSICIAN WHO WAS AT THE SIDE OF THE LOWER EXTREMITY INSERTED A 16FR SHEATH FROM THE INGUINAL REGION. THE 16FR SHEATH WAS REMOVED AND THE THICKENED LEAD WITH ATTACHED ONE-TIE WAS UNABLE TO PASS THROUGH THE PUNCTURE SITE AND BECAME STUCK. THE PHYSICIAN TRIED UNSUCCESSFULLY TO INTRODUCE THE ONE-TIE INTO COOK'S BYRD DILATOR SHEATH VIA THE INGUINAL LEAD, THEREFORE THE TRANSVENOUS APPROACH WAS ABANDONED. A VAS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12247021 ↗
FDA clearance and classification
The FDA regulates product code DWS (instruments, surgical, cardiovascular) under 21 CFR 870.4500 as a Class I device. 21 CFR 870.4500 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the One-Tie G31929
Has the One-Tie G31929 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002319295), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the One-Tie G31929?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jul 29, 2021 and Aug 13, 2026. In those reports, the most common specific device problem is “Improper or Incorrect Procedure or Method” (1); the most common patient problems are “Great Vessel Perforation” (1) and “Thrombosis/Thrombus” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the One-Tie G31929?
2 reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the One-Tie G31929 FDA approved?
Devices of this type (product code DWS) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Vandergrift Inc.
Who makes the One-Tie G31929, and is it still sold?
The FDA's device database lists Cook Vandergrift Inc. as the company responsible for it (brand One-Tie, model G31929). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other INSTRUMENTS, SURGICAL, CARDIOVASCULAR devices
- ACCESSRAIL Platform (Standard Blade) SB-1000Maquet Cardiovascular LLC · Instruments, Surgical, Cardiovascular1 recall6 reports
- ADVANTA SLIDER GDS 22092Atrium Medical Corporation · INSTRUMENTS, SURGICAL, CARDIOVASCULAR1 recall0 reports
- ADVANTA SLIDER GDS 22125Atrium Medical Corporation · INSTRUMENTS, SURGICAL, CARDIOVASCULAR1 recall0 reports
- ADVANTA SLIDER GDS 22025Atrium Medical Corporation · INSTRUMENTS, SURGICAL, CARDIOVASCULAR1 recall0 reports
- Lead Clippers G20003Cook Vandergrift Inc. · INSTRUMENTS, SURGICAL, CARDIOVASCULAR1 recall0 reports
- Needle NDL-107-04Galt Medical Corp. · Instruments, Surgical, Cardiovascular1 recall0 reports
- PILLING IPN005932Teleflex Incorporated · INSTRUMENTS, SURGICAL, CARDIOVASCULAR18 reports
- PILLING IPN915270Teleflex Incorporated · INSTRUMENTS, SURGICAL, CARDIOVASCULAR16 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page