Skip to content

Cook Spectrum G35136Recalls & reported problems

Cook Incorporated · Spectrum, Central Venous Tray Minocycline/Rifampin Impregnated Triple Lumen Polyurethane Catheter · Class 2

All models of Cook Spectrum (67)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002351363), received between Jun 21, 2024 and Jun 26, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Crack” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack3

Patient problems

  • Insufficient Information2
  • Low Blood Pressure/ Hypotension1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 26, 2024Crack; Crack

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Insufficient Information

    MAUDE report 19616761 ↗
  • InjuryJun 25, 2024Crack; Crack

    IT WAS REPORTED THAT THE BLUE HUB OF A COOK SPECTRUM MINOCYCLINE/RIFAMPIN IMPREGNATED TRIPLE LUMEN CENTRAL VENOUS CATHETER TRAY CRACKED. THE DEVICE WAS PLACED IN THE PATIENT'S LEFT INTERNAL JUGULAR ON (B)(6) 2024. ALCOHOL SWABS WERE USED TO CLEAN THE HUBS DURING NEEDLELESS CAP PLACEMENT AND FOR HUB ACCESS. ON (B)(6) 2024, THE PATIENT HAD BLOOD PRESSURE MEDICATION INFUSING THROUGH THE BLUE LUMEN WHEN A SUDDEN DROP IN BLOOD PRESSURE WAS OBSERVED. ADDITIONALLY, IT WAS NOTED THE BED LINEN WAS WET. AT THIS TIME, THE CRACK IN THE BLUE HUB WAS DISCOVERED. THE INFUSION WAS MOVED TO A DIFFERENT LUMEN, AND THE BLOOD PRESSURE STABILIZED. THEN, THE DEVICE WAS REMOVED AND REPLACED. THE CUSTOMER NOTED THEY DO NOT BELIEVE THE NEEDLELESS ENDCAPS ARE BEING OVERTIGHTENED. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT.

    Patient problems: Low Blood Pressure/ Hypotension

    MAUDE report 19609760 ↗
  • InjuryJun 21, 2024Crack; Crack

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Insufficient Information

    MAUDE report 19588789 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page