Entuit Thrive EntuitTM Thrive Balloon Retention Gastrostomy Feeding Tube - 18 FRRecalls & reported problems

Xeridiem Medical Devices, Inc. · Entuit Gastrostomy BR Balloon Retention Feeding Tube, 18 FR Ref. Number: SBRD-18 · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
2
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002356177), received between Apr 27, 2018 and Dec 6, 2024 — classified as 2 deaths, 4 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated5
  • Deflation Problem3
  • Improper or Incorrect Procedure or Method3
  • Migration or Expulsion of Device2
  • Device Handling Problem1
  • Material Rupture1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • Death2
  • Abdominal Pain1
  • Erosion1
  • Feeding Problem1
  • Gastrointestinal Hemorrhage1
  • Unspecified Gastrointestinal Problem1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 6, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    (B)(6) EMAILED, CUSTOMER REPORTED PEG PROBE PLACEMENT ON (B)(6) .2024 ON 8.11 RE-PRESENTATION BECAUSE FOOD PROTRUDES NEXT TO THE TUBE, ON EXAMINATION IT WAS SHOWN THAT THE PEG TUBE WAS NO LONGER IN SITU; IN THE CASE OF THE REMOVED PEG PROBE, AN EXTERNAL PLATE WAS VISIBLE INSIDE T-TUBE WITHOUT AN ANTIMICROBIAL PLATE, REPORTED LOT NOT ON EOSKS, LOGGED AGAINST LAST SHIPPED ORDER. BASED OFF OF THIS STATEMENT IN THE REPORT "WHEN THE PEG TUBE WAS REMOVED, THERE WAS AN EXTERNAL PLATE INSIDE THE T-TUBE WITHOUT A COUNTER PLATE" CAN YOU POINT OUT THE AFFECTED AREA? BECAUSE FOOD HAS ESCAPED FROM THE CREATED TRACT, IT WAS SLIGHTLY RED. THAT'S WHY THE PAT WENT BACK TO THE HOSPITAL.

    Patient problems: Unspecified Gastrointestinal Problem

    MAUDE report 20871921 ↗
  • InjuryDec 6, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE AUSTRIAN FEDERAL OFFICE FOR SAFETY IN HEALTH CARE EMAILED, CUSTOMER REPORTED: AFTER INSERTING THE PEG TUBE ON (B)(6) 2024, THE PATIENT HAD PAIN AND COAGULATED BLOOD AROUND THE PEG TUBE. IN THE CT ABDOMEN, THE PEG PROBE WAS DISLOCATED OUTSIDE THE STOMACH, THE CATHETER TIP WAS AGAINST THE OUTER STOMACH WALL, THE BALLOON INTACT (5ML). PAT. HAD TO BE ACUTELY ADMITTED TO THE OPERATING ROOM FOR (B)(6). EXPORTATION WITH NEW INSTALLATION OF A GASTRIC-TUBE, LOT CONFIRMED.

    Patient problems: Abdominal Pain; Gastrointestinal Hemorrhage

    MAUDE report 20866718 ↗
  • InjuryFeb 1, 2024Material Rupture; Material Rupture

    G-TUBE WAS PLACED ON (B)(6)2023; WITHIN 30 HOURS IT SHOWED ON CT THAT THE DEVICE BALLOON HAD APPARENTLY RUPTURED. STOMA TRACT CLOSED AS A RESULT AND PATIENT HAD TO GO TO GI LAB TO HAVE THE TUBE REPLACED ON (B)(6)2023.

    Patient problems: Erosion

    MAUDE report 18628127 ↗
  • DeathNov 13, 2018Migration or Expulsion of Device; Migration or Expulsion of Device

    ENTUIT TUBE PLACED (B)(6). RETENTION BALLOON INFLATED. FEEDS PER TUBE INITIATED (B)(6). ON (B)(6), PT WITH SEVERE ABD PAIN, CT SHOWED MIGRATION OF TUBE OUTSIDE OF STOMACH AND BALLOON DEFLATED. PT TO OR FOR PERITONEAL LAVAGE. RESULTED IN PERITONITIS/SEPSIS. PT MADE DNR AND EXPIRED (B)(6) 2018.

    Patient problems: Death

    MAUDE report 8064310 ↗
  • DeathOct 30, 2018Migration or Expulsion of Device

    ENTUIT TUBE PLACED (B)(6) 2018. RETENTION BALLOON INFLATED. FEEDS PER TUBE INITIATED (B)(6). (B)(6), PT WITH SEVERE ABD PAIN, CT SHOWED MIGRATION OF TUBE OUTSIDE OF STOMACH AND BALLOON DEFLATED. PT TO OPERATING ROOM FOR PERITONEAL LAVAGE. RESULTED IN PERITONITIS/SEPSIS. PT MADE DNR AND EXPIRED (B)(6) 2018.

    Patient problems: Death

    MAUDE report 8028950 ↗
  • InjuryJun 1, 2018Device Handling Problem; Device Handling Problem

    THE INTERVENTIONAL RADIOLOGY DEPARTMENT HAS PLACED THREE ENTUIT THRIVE BALLOON RETENTION GASTROSTOMY TUBES IN THE LAST 6 MONTHS THAT HAVE BECOME DISLODGED OR FALLEN OUT BECAUSE SOMETHING WAS ATTACHED TO THE BALLOON PORT, WHICH IS NOT LABELED, CAUSING THE BALLOON TO DEFLATE. THE BALLOON IS A RETENTION BALLOON TO KEEP THE TUBE IN THE STOMACH. THE DISLODGED TUBE RESULTED IN A PATIENT HAVING TO HAVE A SURGICAL PROCEDURE; THE OTHER 2 HAD TO HAVE THE TUBES REPLACED. THE PORT IS NOT LABELED AS A BALLOON, WHILE THE OTHER 2 PORTS ARE LABELED AS MEDICINE PORT AND FEEDING PORT. THE PATIENT INVOLVED IN THIS EVENT REQUIRED SURGICAL INTERVENTION.

    Patient problems: Feeding Problem

    MAUDE report 7563503 ↗
  • MalfunctionApr 27, 2018Deflation Problem; Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated

    THE INTERVENTIONAL RADIOLOGY DEPARTMENT HAS PLACED THREE ENTUIT THRIVE BALLOON RETENTION GASTROSTOMY TUBES IN THE LAST 6 MONTHS THAT HAVE BECOME DISLODGED OR FALLEN OUT BECAUSE SOMETHING WAS ATTACHED TO THE BALLOON PORT, WHICH IS NOT LABELED, CAUSING THE BALLOON TO DEFLATE. THE BALLOON IS A RETENTION BALLOON TO KEEP THE TUBE IN THE STOMACH. ONE DISLODGED TUBE RESULTED IN A PATIENT HAVING TO HAVE A SURGICAL PROCEDURE; THE OTHER 2 HAD TO HAVE THE TUBES REPLACED. THIS PORT IS NOT LABELED AS A BALLOON, WHILE THE OTHER 2 PORTS ARE LABELED AS MEDICINE PORT AND FEEDING PORT. THE PATIENT INVOLVED IN THIS EVENT REQUIRED REPLACEMENT OF THEIR G-TUBE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7468244 ↗
  • MalfunctionApr 27, 2018Deflation Problem; Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated

    THE INTERVENTIONAL RADIOLOGY DEPARTMENT HAS PLACED THREE ENTUIT THRIVE BALLOON RETENTION GASTROSTOMY TUBES IN THE LAST 6 MONTHS THAT HAVE BECOME DISLODGED OR FALLEN OUT BECAUSE SOMETHING WAS ATTACHED TO THE BALLOON PORT, WHICH IS NOT LABELED, CAUSING THE BALLOON TO DEFLATE. THE BALLOON IS A RETENTION BALLOON TO KEEP THE TUBE IN THE STOMACH. ONE DISLODGED TUBE RESULTED IN A PATIENT HAVING TO HAVE A SURGICAL PROCEDURE; THE OTHER 2 HAD TO HAVE THE TUBES REPLACED. THIS PORT IS NOT LABELED AS A BALLOON, WHILE THE OTHER 2 PORTS ARE LABELED AS MEDICINE PORT AND FEEDING PORT. THE PATIENT INVOLVED IN THIS EVENT REQUIRED SURGICAL INTERVENTION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7468206 ↗
  • MalfunctionApr 27, 2018Deflation Problem; Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated

    THE INTERVENTIONAL RADIOLOGY DEPARTMENT HAS PLACED THREE ENTUIT THRIVE BALLOON RETENTION GASTROSTOMY TUBES IN THE LAST 6 MONTHS THAT HAVE BECOME DISLODGED OR FALLEN OUT BECAUSE SOMETHING WAS ATTACHED TO THE BALLOON PORT, WHICH IS NOT LABELED, CAUSING THE BALLOON TO DEFLATE. THE BALLOON IS A RETENTION BALLOON TO KEEP THE TUBE IN THE STOMACH. ONE DISLODGED TUBE RESULTED IN A PATIENT HAVING TO HAVE A SURGICAL PROCEDURE; THE OTHER 2 HAD TO HAVE THE TUBES REPLACED. THIS PORT IS NOT LABELED AS A BALLOON, WHILE THE OTHER 2 PORTS ARE LABELED AS MEDICINE PORT AND FEEDING PORT. THE PATIENT INVOLVED IN THIS EVENT REQUIRED REPLACEMENT OF THEIR G-TUBE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7468187 ↗

Other Tubes, Gastrointestinal (And Accessories) devices

All Tubes, Gastrointestinal (And Accessories) devices →