Entuit Thrive EntuitTM Thrive Balloon Retention Gastrostomy Feeding Tube - 18 FRRecalls & reported problems
Xeridiem Medical Devices, Inc. · Entuit Gastrostomy BR Balloon Retention Feeding Tube, 18 FR Ref. Number: SBRD-18 · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 2
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002356177), received between Apr 27, 2018 and Dec 6, 2024 — classified as 2 deaths, 4 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated5
- Deflation Problem3
- Improper or Incorrect Procedure or Method3
- Migration or Expulsion of Device2
- Device Handling Problem1
- Material Rupture1
Patient problems
- No Known Impact Or Consequence To Patient3
- Death2
- Abdominal Pain1
- Erosion1
- Feeding Problem1
- Gastrointestinal Hemorrhage1
- Unspecified Gastrointestinal Problem1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryDec 6, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
(B)(6) EMAILED, CUSTOMER REPORTED PEG PROBE PLACEMENT ON (B)(6) .2024 ON 8.11 RE-PRESENTATION BECAUSE FOOD PROTRUDES NEXT TO THE TUBE, ON EXAMINATION IT WAS SHOWN THAT THE PEG TUBE WAS NO LONGER IN SITU; IN THE CASE OF THE REMOVED PEG PROBE, AN EXTERNAL PLATE WAS VISIBLE INSIDE T-TUBE WITHOUT AN ANTIMICROBIAL PLATE, REPORTED LOT NOT ON EOSKS, LOGGED AGAINST LAST SHIPPED ORDER. BASED OFF OF THIS STATEMENT IN THE REPORT "WHEN THE PEG TUBE WAS REMOVED, THERE WAS AN EXTERNAL PLATE INSIDE THE T-TUBE WITHOUT A COUNTER PLATE" CAN YOU POINT OUT THE AFFECTED AREA? BECAUSE FOOD HAS ESCAPED FROM THE CREATED TRACT, IT WAS SLIGHTLY RED. THAT'S WHY THE PAT WENT BACK TO THE HOSPITAL.
Patient problems: Unspecified Gastrointestinal Problem
MAUDE report 20871921 ↗InjuryDec 6, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE AUSTRIAN FEDERAL OFFICE FOR SAFETY IN HEALTH CARE EMAILED, CUSTOMER REPORTED: AFTER INSERTING THE PEG TUBE ON (B)(6) 2024, THE PATIENT HAD PAIN AND COAGULATED BLOOD AROUND THE PEG TUBE. IN THE CT ABDOMEN, THE PEG PROBE WAS DISLOCATED OUTSIDE THE STOMACH, THE CATHETER TIP WAS AGAINST THE OUTER STOMACH WALL, THE BALLOON INTACT (5ML). PAT. HAD TO BE ACUTELY ADMITTED TO THE OPERATING ROOM FOR (B)(6). EXPORTATION WITH NEW INSTALLATION OF A GASTRIC-TUBE, LOT CONFIRMED.
Patient problems: Abdominal Pain; Gastrointestinal Hemorrhage
MAUDE report 20866718 ↗InjuryFeb 1, 2024Material Rupture; Material Rupture
G-TUBE WAS PLACED ON (B)(6)2023; WITHIN 30 HOURS IT SHOWED ON CT THAT THE DEVICE BALLOON HAD APPARENTLY RUPTURED. STOMA TRACT CLOSED AS A RESULT AND PATIENT HAD TO GO TO GI LAB TO HAVE THE TUBE REPLACED ON (B)(6)2023.
Patient problems: Erosion
MAUDE report 18628127 ↗DeathNov 13, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
ENTUIT TUBE PLACED (B)(6). RETENTION BALLOON INFLATED. FEEDS PER TUBE INITIATED (B)(6). ON (B)(6), PT WITH SEVERE ABD PAIN, CT SHOWED MIGRATION OF TUBE OUTSIDE OF STOMACH AND BALLOON DEFLATED. PT TO OR FOR PERITONEAL LAVAGE. RESULTED IN PERITONITIS/SEPSIS. PT MADE DNR AND EXPIRED (B)(6) 2018.
Patient problems: Death
MAUDE report 8064310 ↗DeathOct 30, 2018Migration or Expulsion of Device
ENTUIT TUBE PLACED (B)(6) 2018. RETENTION BALLOON INFLATED. FEEDS PER TUBE INITIATED (B)(6). (B)(6), PT WITH SEVERE ABD PAIN, CT SHOWED MIGRATION OF TUBE OUTSIDE OF STOMACH AND BALLOON DEFLATED. PT TO OPERATING ROOM FOR PERITONEAL LAVAGE. RESULTED IN PERITONITIS/SEPSIS. PT MADE DNR AND EXPIRED (B)(6) 2018.
Patient problems: Death
MAUDE report 8028950 ↗InjuryJun 1, 2018Device Handling Problem; Device Handling Problem
THE INTERVENTIONAL RADIOLOGY DEPARTMENT HAS PLACED THREE ENTUIT THRIVE BALLOON RETENTION GASTROSTOMY TUBES IN THE LAST 6 MONTHS THAT HAVE BECOME DISLODGED OR FALLEN OUT BECAUSE SOMETHING WAS ATTACHED TO THE BALLOON PORT, WHICH IS NOT LABELED, CAUSING THE BALLOON TO DEFLATE. THE BALLOON IS A RETENTION BALLOON TO KEEP THE TUBE IN THE STOMACH. THE DISLODGED TUBE RESULTED IN A PATIENT HAVING TO HAVE A SURGICAL PROCEDURE; THE OTHER 2 HAD TO HAVE THE TUBES REPLACED. THE PORT IS NOT LABELED AS A BALLOON, WHILE THE OTHER 2 PORTS ARE LABELED AS MEDICINE PORT AND FEEDING PORT. THE PATIENT INVOLVED IN THIS EVENT REQUIRED SURGICAL INTERVENTION.
Patient problems: Feeding Problem
MAUDE report 7563503 ↗MalfunctionApr 27, 2018Deflation Problem; Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated
THE INTERVENTIONAL RADIOLOGY DEPARTMENT HAS PLACED THREE ENTUIT THRIVE BALLOON RETENTION GASTROSTOMY TUBES IN THE LAST 6 MONTHS THAT HAVE BECOME DISLODGED OR FALLEN OUT BECAUSE SOMETHING WAS ATTACHED TO THE BALLOON PORT, WHICH IS NOT LABELED, CAUSING THE BALLOON TO DEFLATE. THE BALLOON IS A RETENTION BALLOON TO KEEP THE TUBE IN THE STOMACH. ONE DISLODGED TUBE RESULTED IN A PATIENT HAVING TO HAVE A SURGICAL PROCEDURE; THE OTHER 2 HAD TO HAVE THE TUBES REPLACED. THIS PORT IS NOT LABELED AS A BALLOON, WHILE THE OTHER 2 PORTS ARE LABELED AS MEDICINE PORT AND FEEDING PORT. THE PATIENT INVOLVED IN THIS EVENT REQUIRED REPLACEMENT OF THEIR G-TUBE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7468244 ↗MalfunctionApr 27, 2018Deflation Problem; Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated
THE INTERVENTIONAL RADIOLOGY DEPARTMENT HAS PLACED THREE ENTUIT THRIVE BALLOON RETENTION GASTROSTOMY TUBES IN THE LAST 6 MONTHS THAT HAVE BECOME DISLODGED OR FALLEN OUT BECAUSE SOMETHING WAS ATTACHED TO THE BALLOON PORT, WHICH IS NOT LABELED, CAUSING THE BALLOON TO DEFLATE. THE BALLOON IS A RETENTION BALLOON TO KEEP THE TUBE IN THE STOMACH. ONE DISLODGED TUBE RESULTED IN A PATIENT HAVING TO HAVE A SURGICAL PROCEDURE; THE OTHER 2 HAD TO HAVE THE TUBES REPLACED. THIS PORT IS NOT LABELED AS A BALLOON, WHILE THE OTHER 2 PORTS ARE LABELED AS MEDICINE PORT AND FEEDING PORT. THE PATIENT INVOLVED IN THIS EVENT REQUIRED SURGICAL INTERVENTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7468206 ↗MalfunctionApr 27, 2018Deflation Problem; Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated
THE INTERVENTIONAL RADIOLOGY DEPARTMENT HAS PLACED THREE ENTUIT THRIVE BALLOON RETENTION GASTROSTOMY TUBES IN THE LAST 6 MONTHS THAT HAVE BECOME DISLODGED OR FALLEN OUT BECAUSE SOMETHING WAS ATTACHED TO THE BALLOON PORT, WHICH IS NOT LABELED, CAUSING THE BALLOON TO DEFLATE. THE BALLOON IS A RETENTION BALLOON TO KEEP THE TUBE IN THE STOMACH. ONE DISLODGED TUBE RESULTED IN A PATIENT HAVING TO HAVE A SURGICAL PROCEDURE; THE OTHER 2 HAD TO HAVE THE TUBES REPLACED. THIS PORT IS NOT LABELED AS A BALLOON, WHILE THE OTHER 2 PORTS ARE LABELED AS MEDICINE PORT AND FEEDING PORT. THE PATIENT INVOLVED IN THIS EVENT REQUIRED REPLACEMENT OF THEIR G-TUBE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7468187 ↗
Other Tubes, Gastrointestinal (And Accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-05090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page