Coda G36042Recalls & reported problems
Cook Incorporated · Coda, Balloon Catheter · Class 2
All models of Coda (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002360426), received between Nov 7, 2017 and Jun 13, 2018 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Rupture3
- Leak/Splash2
- Inflation Problem1
- Split1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 13, 2018Material Rupture
IT WAS REPORTED THAT DURING THE EVAR PROCEDURE, WHEN THE CODA BALLOON CATHETER WAS USED, THE BALLOON RUPTURED. THERE WERE NO PATIENT INJURIES OR ADDITIONAL PROCEDURES REPORTED AND ANOTHER DEVICE WAS USED TO CONTINUE THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7598285 ↗MalfunctionJun 13, 2018Leak/Splash; Split
THE COMPLAINANT REPORTED THAT A LEAKAGE WAS FOUND IN THE CODA BALLOON CATHETER DURING THE PROCEDURE. IT WAS FURTHER NOTED THAT THERE WAS A SPLIT IN THE ENTIRE LENGTH OF THE BALLOON. THERE WERE NO REPORTED SERIOUS INJURIES TO THE PATIENT AS A RESULT OF THE PRODUCT PROBLEM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7596871 ↗MalfunctionDec 15, 2017Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash
THE USER FACILITY REPORTED THE PATIENT UNDERWENT AN ENDOVASCULAR ANEURYSM REPAIR (EVAR) PROCEDURE TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). A MAINBODY STENT GRAFT AND TWO LEGS WERE PLACED. THE PROXIMAL SIDE OF THE CONTRALATERAL LEG WAS SLIGHTLY PROTRUDING INTO THE MAINBODY. A COMPETITOR'S BALLOON WAS USED TO PERFORM A TOUCH UP OF THE STENT GRAFT, BUT THE BALLOON RUPTURED. THE REPORTER ADVISED THE BALLOON MAY HAVE BEEN RUPTURED DUE TO THE PROTRUDING CONTRALATERAL LEG. ANOTHER OF THE SAME BALLOON WAS USED TO PERFORM THE TOUCH UP. THEN A COOK CODA BALLOON CATHETER WAS USED TO TREAT A SMALL ENDOLEAK WHICH APPEARED TO BE A PROXIMAL TYPE I ENDOLEAK AT THE PROXIMAL SIDE OF THE MAIN BODY. THERE WAS NO INFORMATION REGARDING THE CODA BALLOON HAVING A PROBLEM DURING PREPARATION. WHEN THE USER ATTEMPTED TO INFLATE THE CODA BALLOON AROUND THE PROXIMAL SIDE OF THE MAINBODY, THE BALLOON WOULD NOT INFLATE AND CONTRAST LEAKED. A SECOND SAME SIZE CODA BALLOON WAS USED, HOWEVER THE SAME THING OCCURRED. AS NOTED BY THE REPORTER, THEY ATTEMPTED TO INFLATE BOTH OF THESE BALLOONS AWAY FROM THE PROXIMAL SIDE OF THE LEG INSIDE OF THE MAINBODY. THE USER THEN USED THE ALTERNATE MANUFACTURERS BALLOON TO CO
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7123805 ↗MalfunctionDec 4, 2017Material Rupture; Material Rupture
THE USER FACILITY REPORTED THAT AN EMERGENT ENDOVASCULAR ANEURYSM REPAIR (EVAR) WAS PERFORMED. THE CODA BALLOON CATHETER WAS USED FOR TOUCH-UP BALLOONING. THE BALLOON RUPTURED DURING USE. ANOTHER BALLOON WAS USED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7081691 ↗MalfunctionNov 7, 2017Material Rupture; Material Rupture
THE USER FACILITY REPORTED THAT THE (B)(6) PATIENT UNDERWENT AN ENDOVASCULAR ANEURYSM REPAIR (EVAR). THE PATIENT ANATOMICAL FORM WAS SUITABLE FOR THE PROCEDURE AND THE PROCEDURE WAS PERFORMED AS LABELED. ANOTHER MANUFACTURER'S GUIDE WIRE WAS USED. BEFORE FINAL ANGIOGRAPHY, THE CODA BALLOON CATHETER WAS USED TO SEAL THE PROXIMAL SIDE, BUT IT RUPTURED DURING USE. ANOTHER CODA BALLOON CATHETER WAS USED INSTEAD TO CONTINUE THE PROCEDURE. THERE HAS BEEN NO ADVERSE EFFECT TO THE PATIENT REPORTED. THE PATIENT HAS BEEN REPORTED TO BE ¿RECOVERED¿.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7005602 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page